User manual for micromotor attachment handpieces CHIRANA
Straight handpiece 110 D
Contra angle handpiece with light 120 L
Contra angle handpiece 120
Contra angle handpiece 120LB
Contra angle handpiece 120LR
1.
TECHNICAL DETAILS :
Input RPM
Max.40 000
Gear Ratio
Color Marking
Blue point
Bur Cooling
side spray
Light
Ø shank / bur
Ø 2.35 / 44.5
length (mm)
2.
IMPORTANT INSTRUCTIONS:
- Straight handpiece and contra angle handpieces are intended
for use in dentistry by trained professionals only.
- Use under safe working conditions only. Follow the proper use
instruction in accordance with this manual.
- Protect the eyes of patient and assisting staff from direct
contact with light source.
- Protect yourself, patients and others from possible accidents.
- Avoid contamination from the product.
- Repairs and maintenance can be done by trained
professionals only.
3.
STANDARD ACCESSORIES :
Jet cleaning needle
SMIOIL spray with nozzle
KN lubricating tip
KT lubricating tip (only 120LR)
4.
FIRST USE:
- Check the content of the package to make sure you received
the attachment with accessories, user manual and a warranty
card.
- Micromotor handpiece attachment connection dimensions
conform to ISO 3964
- Use only burs and cutters that correspond to ISO 1797-1 with
shank diameter of 2.334 ÷ 2.350 mm, maximal length of 22 mm
for contra angle heads and maximal length of 44.5 mm for
straight attachments.
- Pay attention to the warnings and instructions provided by
manufacturers of burs and cutters.
- DO NOT USE burs and cutters that deviate from the indicated
data.
- DO NOT USE burs with worn shank. They could fall out during
treatment and injure the patient. Once in a while check the
sharpness of burs under a magnifying glass.
4.1. Coupling with micromotor
Attaching:
With a slight twisting motion slip the attachment
handpiece onto the micromotor coupling until the
driving and chuck mechanisms snaps into place.
Removal:
Remove the attachment handpiece off of the
micromotor by a slight pull along the micromotor axis
while simultaneously twisting the attachment around.
Micromotor attachment handpieces CHIRANA are manufactured according to directive MDD 93/42 / EEC under
ISO 13485: 2012 quality management system.
No. 39 021 2567 Pic.1
No. 39 033 2827 Pic.2
No. 39 033 2837 Pic.2
No. 39 033 2853
No. 39 033 2854
110D
120
Max.40 000
-1
-1
min
min
1:1
1:1
Blue point
side spray
-
-
-
1 pc
No. T 070 2603 4
1 pc
No. 39 031 2690,1
1 pc
No. 070 2634 4
1 pc
No. 3152 7608
pic. 3
Adjust the flow of water exiting the nozzle of the handpiece cooling
spray to min. 50 ml/min. at the inlet pressure of 200kPa. Air flow of the
cooling spray 1.5NL/min at the inlet pressure 200kPa.
120L
120LB
Max.40 000
Max.40 000
-1
-1
min
min
1:1
1:1
Blue point
Blue point
side spray
4-spray
yes
yes
Ø 2.35 / 22
-
4.2. Attaching head to the contra angle
168 D-T
168 D
(Does not apply for 110D, 120LB, 120LR) Rotate the clamping ring of
contra angle in the direction of the arrow until it stops and hold it in
position. Insert the head into the contra angle so the grooves on the head
match the slots inside the contra angle. Release the clamping ring.
4.3. Straight handpiece bur insertion and replacement
pic. 5
The bur gripping mechanism is released by rotating the clamping ring in
the indicated direction. Reverse rotation will tighten and lock the inserted
bur.
4.4 Bur insertion and replacement on 120LB, 120LR contra angles
Inserting the bur - press the push button firmly with your thumb and
simultaneously insert the bur all the way into the hole. The correct
clamping position of latch burs is found by rotating the bur around its axis
(applies for 120LB). To remove the bur - press the push button with your
thumb and pull the bur out. After each change of the drill it is necessary
to check its clamping by pulling it outwards from the head.
Minimum requirement for the cooling air supplied by
micromotor is 6 NL/min.
Insufficient cooling can increase the temperature of
external parts.
DO NOT attach or remove any handpiece
attachments from the micromotor while it is running.
120LR
Max.40 000
-1
min
1:5
Red point
4-spray
yes
Ø 1.6 / 22
- Never insert the head during rotation.
- Check the head's proper connection with contra
angle handpiece before each use.
Fig.6
pic. 4

Advertisement

loading
Need help?

Need help?

Do you have a question about the 110D and is the answer not in the manual?

Questions and answers

Summary of Contents for Chirana Medical 110D

  • Page 1 - Repairs and maintenance can be done by trained professionals only. (Does not apply for 110D, 120LB, 120LR) Rotate the clamping ring of contra angle in the direction of the arrow until it stops and hold it in STANDARD ACCESSORIES : position.
  • Page 2 Contact an authorized service organization or the manufacturer CHIRANA Medical, a. s. When repairing, use the original spare parts from the device producer CHIRANA Medical, a. s. 7. DEVICE DISPOSAL Disposal must be done in strict compliance with all applicable regulations.

This manual is also suitable for:

120120l120lb120lr