EPX-SERISE USER MANUAL Rev. G
Ecotron Co., Ltd
Made by Korea
#504~505Ho, Hanshin IT Tower II, 47, Digital-ro 9-gil, Geumcheon-Gu, Seoul, Korea
Obelis S.A
Avenue de Tervuren, 34, bte 44 1040 Brussels, Belgium / E-mail:mail@obelis.net
Representative: Mr.Gideon ELKAYAM / TEL:32.2.732.59.54, FAX:32.2.732.60.03
D/N: EPX-SERISE, Rev. G
EPX-Series
Diagnostic X-ray System
User Manual
1

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the EPX Series and is the answer not in the manual?

Questions and answers

Summary of Contents for Ecotron EPX Series

  • Page 1 EPX-SERISE USER MANUAL Rev. G EPX-Series Diagnostic X-ray System User Manual Ecotron Co., Ltd Made by Korea #504~505Ho, Hanshin IT Tower II, 47, Digital-ro 9-gil, Geumcheon-Gu, Seoul, Korea Obelis S.A Avenue de Tervuren, 34, bte 44 1040 Brussels, Belgium / E-mail:mail@obelis.net Representative: Mr.Gideon ELKAYAM / TEL:32.2.732.59.54, FAX:32.2.732.60.03...
  • Page 2 EPX-SERISE USER MANUAL Rev. G (This page intentionally left blank)
  • Page 3: Table Of Contents

    EPX-SERISE USER MANUAL Rev. G CONTENTS 1. INTRODUCTION----------------------------------------------------------------------------------------- 6 2. NOTICE OF THE SAFE OPERATION ------------------------------------------------------------- 8 3. RADIATION SAFETY ----------------------------------------------------------------------------------- 9 4. SAFETY AND SPECIFICATIONS ------------------------------------------------------------------ 10 5. SYSTEM CONFIGURATION------------------------------------------------------------------------- 47 6. OPERATION MODE ----------------------------------------------------------------------------------- 50 7. MAINTENANCE PROCEDURE -------------------------------------------------------------------- 58 8.
  • Page 4 EPX-SERISE USER MANUAL Rev. G REVISON HISTORY Revision Number Date Description SEP 11, 2013 First Edition SEP 30, 2014 European Authorized Representative Change JAN 12, 2015 Manual Updates Mar 23, 2015 Manual Updates JUN 11, 2015 Manual Updates SEP 01, 2015 Manual Updates JUN 21, 2016 Manual Updates...
  • Page 5 EPX-SERISE USER MANUAL Rev. G (This page intentionally left blank)
  • Page 6: Introduction

    EPX-SERISE USER MANUAL Rev. G 1.INTRODUCTION The EPX Series Innovation diagnostic X-ray system features excellent x-ray Image, quality, accuracy, reproducibility and long-term stability. Fast kV rising time minimize unnecessary patient dose and supply superior Image quality, fast Image response. The operator control console is designed to be simple and user-friendly and one can select or change x- ray parameters easily using a large soft membrane switch.
  • Page 7 EPX-SERISE USER MANUAL Rev. G 1.1 COMPOSITION This product is composed of the X-ray component part, accessories part. Detail information of each component follows below. (1) X-ray component part  Main Power board  SMPS  Mono block X-ray tank ...
  • Page 8: Notice Of The Safe Operation

    EPX-SERISE USER MANUAL Rev. G 2. NOTICE OF THE SAFE OPERATION This user’s guide is designed to ensure correct use and operation of EPX series. Please read all the lines thoroughly before you use this equipment. Incorrect use and operation exceeding described conditions in this manual may occur damage of the machine and shorten its life.
  • Page 9: Radiation Safety

    EPX-SERISE USER MANUAL Rev. G 3. RADIATION SAFETY 1. Users and operators should wear appropriate protecting devices and clothes. Stay distance from the radiant sources and all the possible secondary radiation zones. Eliminate all unnecessary objects near the exposure zones. The distance from the focus to skin should be kept at least 8 inch (20cm).
  • Page 10: Safety And Specifications

    EPX-SERISE USER MANUAL Rev. G 4. SAFETY AND SPECIFICATIONS 4.1 SAFETY AND WARNING SYMBOLS Never allow unqualified personnel to operate the X-ray X-RAY EQUIPMENT IS DANGEROUS TO BOTH PATIENT AND OPERATOR UNLESS MEASURES OF PROTECTION ARE STRICTLY OBSERVED system. High voltage symbol used to indicate the presence of high voltage.
  • Page 11 EPX-SERISE USER MANUAL Rev. G DANGER HIGH VOLTAGE LABEL This label attached to the main fuse cover, main noise filter cover and FET driver board. Main voltage is present inside the x- ray system whenever the main disconnect is switched on. Additionally, the DC link capacitors will remain charged for up to 3 minutes after the AC mains is disconnected or the AC switch is switched off.
  • Page 12 Service and operating personnel only authorized and properly trained by ECOTRON should be allowed to work with this X-ray system equipment. The appropriate personnel must be made aware of the inherent dangers associated with the servicing of high voltage equipment and the danger of excessive exposure to X-ray radiation during system operation.
  • Page 13 EPX-SERISE USER MANUAL Rev. G ● Never operate this X-ray system in areas where there is a risk of explosion. Detergents and disinfectants, including those used on patients, may create explosive mixtures of gases. Please observe the relevant regulations. ● The operator console or anything electrically connected to it, must never be used within 6ft(1.8m) of the patient environment.
  • Page 14 Inherent Filtration: At least 0.8mmAl Certification Statement ▪ This product complies with 21 CFR 1020.30 – 1020.31 ▪ This product complies with 21 CFR Subchapter J. Ecotron Co., Ltd Made by Korea #504~505Ho, Hanshin IT Tower II, 47, Digital-ro 9-gil, Geumcheon-Gu, Seoul, Korea Obelis S.A...
  • Page 15 Inherent Filtration: At least 0.8mmAl Certification Statement ▪ This product complies with 21 CFR 1020.30 – 1020.31 ▪ This product complies with 21 CFR Subchapter J. Ecotron Co., Ltd Made by Korea #504~505Ho, Hanshin IT Tower II, 47, Digital-ro 9-gil, Geumcheon-Gu, Seoul, Korea Obelis S.A...
  • Page 16 Inherent Filtration: At least 0.8mmAl Certification Statement ▪ This product complies with 21 CFR 1020.30 – 1020.31 ▪ This product complies with 21 CFR Subchapter J. Ecotron Co., Ltd Made by Korea #504~505Ho, Hanshin IT Tower II, 47, Digital-ro 9-gil, Geumcheon-Gu, Seoul, Korea Obelis S.A...
  • Page 17 Inherent Filtration: At least 0.8mmAl Certification Statement ▪ This product complies with 21 CFR 1020.30 – 1020.31 ▪ This product complies with 21 CFR Subchapter J. Ecotron Co., Ltd Made by Korea #504~505Ho, Hanshin IT Tower II, 47, Digital-ro 9-gil, Geumcheon-Gu, Seoul, Korea Obelis S.A...
  • Page 18 Inherent Filtration: At least 0.5mmAl Certification Statement ▪ This product complies with 21 CFR 1020.30 – 1020.31 ▪ This product complies with 21 CFR Subchapter J. Ecotron Co., Ltd Made by Korea #504~505Ho, Hanshin IT Tower II, 47, Digital-ro 9-gil, Geumcheon-Gu, Seoul, Korea Obelis S.A...
  • Page 19 Inherent Filtration: At least 0.5mmAl Certification Statement ▪ This product complies with 21 CFR 1020.30 – 1020.31 ▪ This product complies with 21 CFR Subchapter J. Ecotron Co., Ltd Made by Korea #504~505Ho, Hanshin IT Tower II, 47, Digital-ro 9-gil, Geumcheon-Gu, Seoul, Korea Obelis S.A...
  • Page 20 ▪ Inherent Filtration: 0.5mmAl Certification Statement ▪ This product complies with 21 CFR 1020.30 – 1020.31 ▪ This product complies with 21 CFR Subchapter J. Ecotron Co., Ltd Made by Korea #504~505Ho, Hanshin IT Tower II, 47, Digital-ro 9-gil, Geumcheon-Gu, Seoul, Korea...
  • Page 21 EPX-SERISE USER MANUAL Rev. G 4.2.8 Symbol on the product label of X-ray system SYMBOL Description Catalogue number Serial number Caution Consult instructions for use WEEE Symbol Manufacturer Date of manufactured Authorized representative in the European Community...
  • Page 22 EPX-SERISE USER MANUAL Rev. G 4.3 Symbol on the Packing of X-ray system SYMBOL Description KEEP DRY: This symbol means keep away from rain Fragile: This symbol means do not drop of product UP: This symbol means “This way up” CAUTION INCORRECT CONNECTIONS OR USE OF UNAPPROVED EQUIPMENT MAY RESULT IN INJURY OR EQUIPMENT DAMAGE.
  • Page 23 ASSOCIATED HAZARDS, BUT ASSUMES NO RESPONSIBILITIES FOR AFTER- SALE POERATING AND SAFETY PRACTICES. ECOTRON.,Ltd ACCEPTS NO RESPONSIBILITY FOR ANY GENERATOR NOT MAINTAINED OR SERVICED ACCORDING TO THE SERVICE MANUAL OR ANY X-RAY SYSTEM THAT HAS BEEN MODIFIED IN ANY WAY.
  • Page 24 EPX-SERISE USER MANUAL Rev. G 4.4. EPX-Series DUTY CYCLE LIMIT NOTE THE FOLLOWING SECTION CONTAINS IMPORTANT INFORMATION. PLEASE READ AND UNDERSTAND THIS MATERIAL BEFORE CONTINUING. Internal X-ray system components will heat up during normal use of the x-ray system. This is similar to X-ray tube heating during normal x-ray system operation.
  • Page 25 EPX-SERISE USER MANUAL Rev. G 4.5 DAILY X-RAY TUBE WARM-UP PROCEDURE NOTE THE FOLLOWING PROCEDURE PRODUCES X-RAYS. OBSERVE ALL SAFETY PRECAUTIONS TO PROTECT PERSONNEL. Use the procedure when the generator has not been used for several days. This procedure provides for exposures at medium power before the tube is used at maximum mA or kV values.
  • Page 26 EPX-SERISE USER MANUAL Rev. G 4.6 APPLICABLE STANDARDS The EPX series comply with the regulatory requirements and design standards in this section as follows: This product complies with DHHS requirements of 21 CFR Subchapter J. 1) SAFETY ● EN 60601-1:2006/A1:2013 ●...
  • Page 27 ● In according with Annex IX: CLASS II b VERIFICATION – IEC 60580 ● The EPX series measures data in accordance with the reference tube voltage and mAs values using DAP certified according to the requirements of IEC60580, and displays output doses by calculating...
  • Page 28 EPX-SERISE USER MANUAL Rev. G 4.7.2 SPECIFICATION TABLE 4.7.2.1 EPX-F1600 Max Output Power 1.6 kW Voltage 200V-240V Input Power Phase & Frequency Single / 50/60 Hz Tube voltage current 40KV ~ 50KV 30mA 0.4 ~ 20 40KV ~ 50KV 25mA 25~32 51KV ~ 60KV 25mA...
  • Page 29 EPX-SERISE USER MANUAL Rev. G 4.7.2.2 EPX-F2400 Max Output Power 2.4 KW Input Power Voltage 200V-240V Phase & Frequency Single, 50/60 Hz Tube voltage current 40KV ~ 60KV 40mA 0.4 ~ 64 40KV ~ 60KV 35mA 80 ~ 100 61KV ~ 70KV 30mA 0.4 ~ 100 Radiography KV Range in 1KV steps...
  • Page 30 EPX-SERISE USER MANUAL Rev. G 4.7.2.3 EPX-F2800 Max Output Power 2.8 kW Voltage 200V-240V Input Power Phase & Frequency Single / 50/60 Hz Tube voltage current 40KV ~ 60KV 40mA 0.4 ~ 100 61KV ~ 70KV 30mA 0.4 ~ 64 61KV ~ 70KV 25mA 80 ~ 100...
  • Page 31 EPX-SERISE USER MANUAL Rev. G Timer Push button illuminator with 30 sec timer Lamp 24V, 50W Halogen Weight 13 Kg 4.7.2.4 EPX-F3200 Max Output Power 3.2 KW Voltage 200V-240V Input Power Phase & Frequency Single / 50/60 Hz Tube voltage current 40KV ~ 50KV 60mA...
  • Page 32 EPX-SERISE USER MANUAL Rev. G 4.7.2.5 EPX-F4000 Max Output Power 4.0 KW Voltage 200V-240V Input Power Phase & Frequency Single / 50/60 Hz Tube voltage current 50KV 80mA 0.4~16mAs 50KV 70mA 20~32mAs 50KV 60mA 40~100mAs 51KV ~ 60KV 60mA 0.4~32mAs 51KV ~ 60KV 50mA 40~100mAs...
  • Page 33 EPX-SERISE USER MANUAL Rev. G Lamp 24V, 50W Halogen Weight 13 Kg 4.7.2.6 EPX-F5000 Max Output Power 5.0 KW Voltage 200V-240V Input Power Phase & Frequency Single / 50/60 Hz Tube voltage current 50KV 100mA 0.4~10mAs 50KV 80mA 13~32mAs 50KV 70mA 40~100mAs 51KV ~ 60KV...
  • Page 34 Shell Material White, Engineering plastics Useful time 1.2 million times 4.7.2.8 Mobile Stand Two step X-ray exposure hand switch Available products EPX Series only Arm angle 0° ~ 150° Max arm vertical 1.4 m Wt. Load(허용무게) 30 kg Wheel size 50Ø, 24mm...
  • Page 35 EPX-SERISE USER MANUAL Rev. G 4.7.3 kV, mAs, mA TABLE 4.7.3.1 EPX-F1600 mAs & Time TABLE 40KV ~ 50KV Tube Current[mA] 30mA 25mA 1000 1280 Current Time[ms] 51KV ~ 60KV 25mA 20mA 1250 1600 61KV ~ 70KV 20mA 15mA 1000 1667 2133 71KV ~ 80KV...
  • Page 36 EPX-SERISE USER MANUAL Rev. G 4.7.3.2 EPX-F2400 mAs & Time TABLE 40KV ~ 60KV 40mA 35mA Tube Current[mA] 1000 1250 1600 2286 2857 Current Time[ms] 61KV ~ 70KV 30mA 1067 1333 1667 2133 2667 3333 71KV ~ 80KV 30mA 25mA 1067 1600 2000 2560 3200 81KV ~ 90KV 25mA...
  • Page 37 EPX-SERISE USER MANUAL Rev. G 4.7.3.3 EPX-F2800 mAs & Time TABLE 40KV ~ 60KV 40mA Tube Current[mA] 1000 1250 1600 2000 2500 Current Time[ms] 61KV ~ 70KV 30[mA} 25mA 1067 1333 1667 2133 3200 4000 71KV ~ 80KV 35mA 30mA 25mA 1067 1600 2000 2560 3200 81KV ~ 90KV...
  • Page 38 EPX-SERISE USER MANUAL Rev. G 4.7.3.4 EPX-F3200 mAs & Time TABLE 40KV ~ 50KV 60mA Tube Current[mA] 1067 1333 1667 Current Time[ms] 51KV ~ 60KV 50mA 1000 1280 1600 61KV ~ 70KV 40mA 1000 1250 1600 2000 2500 71KV ~ 80KV 40mA 35mA 1143 1429 1829 2286...
  • Page 39 EPX-SERISE USER MANUAL Rev. G 4.7.3.5 EPX-F4000 mAs & Time TABLE 50KV 80mA 70mA 60mA Tube Current[mA] 1067 1333 1667 Current Time[ms] 51KV ~ 60KV 60mA 50mA 1000 1280 1600 2000 61KV ~ 70KV 50mA 40mA 1000 1250 1600 2000 2500 71KV ~ 80KV 45mA 40mA...
  • Page 40 EPX-SERISE USER MANUAL Rev. G 4.7.3.6 EPX-F5000 mAs & Time TABLE 50KV 100mA 80mA 70mA Tube Current[mA] 1143 1429 Current Time[ms] 51KV ~ 60KV 80mA 70mA 60mA 1067 1333 1667 61KV ~ 70KV 70mA 60mA 50mA 1000 1280 1600 2000 71KV ~ 80KV 60mA 50mA...
  • Page 41 EPX-SERISE USER MANUAL Rev. G 4.7.4 Dosimetry Data TABLE 4.7.4.1 EPX-F1600 Unit: μGy*㎡ 11.1 21.8 34.1 47.2 60.7 75.4 70.7 218.2 302.8 392.4 4.7.4.2 EPX-F2400 Unit: μGy*㎡ 11.1 21.8 34.1 47.2 60.7 75.4 90.8 70.7 218.2 302.8 392.4 590.1 1231.2 1489.9 221.3 433.8...
  • Page 42 EPX-SERISE USER MANUAL Rev. G 4.7.4.4 EPX-F3200 Unit: μGy*㎡ 12.6 25.4 40.1 56.5 74.2 112.8 81.9 165.8 259.1 363.7 479.6 601.9 731.1 650.1 1198.9 1508.3 1843.9 261.7 522.7 1138.7 4.7.4.4 EPX-F4000 Unit: μGy*㎡ 19.4 32.2 46.7 61.4 77.5 92.8 132.5 217.8 406.6 513.9...
  • Page 43 EPX-SERISE USER MANUAL Rev. G 4.7.5 MECHANICAL SPECIFICATION EPX-F1600/F2400/F2800/F3200/F4000/F5000 D imension (unit : mm) Side view Front view...
  • Page 44 EPX-SERISE USER MANUAL Rev. G Hand Switch D imension (unit : mm) Mobile Stand D imension (unit : mm)
  • Page 45 EPX-SERISE USER MANUAL Rev. G 4.8 WORKING ENVIRONMENT Avoid the following places for the normal operation and safe storage: a. where the equipment is exposed to water vapor. b. where the equipment is exposed to direct sunlight. c. where the equipment is exposed to dust. d.
  • Page 46 EPX-SERISE USER MANUAL Rev. G 4.9 CUSTOMER SUPPORT Address any questions regarding X-ray system operation to: ECOTRON Co.,Ltd #505Ho, Hanshin IT Tower II, 60-18, Gasang-Dong, Geumcheon-Gu, Seoul 153-023, Korea TEL:+82-2-2025-3760, FAX:+82-2-2025-3764 E-mail:Ecotron@ecotron.co.kr Web-site: http://www.ecotron.co.kr In USA Contact DRGEM USA Inc.
  • Page 47: System Configuration

    EPX-SERISE USER MANUAL Rev. G 5. SYSTEM CONFIGURATION EPX Series is a radiological device for professional radiologist uses. It should be applied for the radiographic diagnosis and operated by qualified practitioners. Users have to comply with safety and health regulations concerning the ionizing radiation protection and the electrical and mechanical safety of the veterinary medical devices.
  • Page 48 EPX-SERISE USER MANUAL Rev. G The operation of the machine should not be started unless all verifications and connections are fully checked. CAUTION For the first time installation of the (machine) or operation after long period of non-use, it needs to be preheated to secure long service life of the tube.
  • Page 49 EPX-SERISE USER MANUAL Rev. G 5.2 CONNECTION OF POWER CODE AND HAND SWITCH Description Function Power Connector(inlet) Connect the power cable Power Switch Power on/off switch Hand Switch Connector Connect the hand switch cable Please check whether the voltage and frequency of the input power are in accordance with the figures written in the system labels, which attached on the body of the machine.
  • Page 50: Operation Mode

    Select mAs value by up or down buttons Ready indicator Lighted push the hand switch one time for preparation PC interface connecting indicator Lighted when EPX series is connected to PC Exposure indicator Alert lamp during x-ray exposure Error lamp...
  • Page 51 EPX-SERISE USER MANUAL Rev. G 6.1.1 Normal Mode a. Connect power plug into the device. b. Turn on the power switch located on the back of the equipment. c. Select the KV values using the KV up/down switch. d. Select the mAs values using the mAs up/down switch. e.
  • Page 52 EPX-SERISE USER MANUAL Rev. G 6.1.4 Cooling Time Display Mode EPX-series have the cooling time function for protecting the device. The cooling time function is to stop the device when operator uses the device over standard. EPX-series will start automatically the cooling time procedure when operator shoots x-ray twice in a row.
  • Page 53 EPX-SERISE USER MANUAL Rev. G 6.2 Touch Main Screen of Operation Mode (Option) 6.2.1 Normal Mode of Touch Main Screen After Start-up Screen is successfully launched, the following Touch Main Screen will be displayed. Description Function kV Indicator Display kV Value (1KV increment) Select kV value by tapping up (increase) or down (decrease) kV value up/down control buttons buttons...
  • Page 54 EPX-SERISE USER MANUAL Rev. G can be accomplished by pressing down the hand-switch completely. Menu button Reserved Ready indicator Lighted push the hand switch one time for preparation X-ray exposure indicator Alert lamp during X-ray exposure Error indicator When the system has error, this Error indicator will light up. ⓐ...
  • Page 55 EPX-SERISE USER MANUAL Rev. G 6.2 Hand Switch Name Description stage: The equipment gets ready when the button is half pressed with a thumb. Hand Switch stage: The equipment exposes X-ray when the button is fully pressed (Dead-man type) with a thumb. It automatically turns off according to the value of mAs set by the program or when the thumb is off the switch.
  • Page 56 EPX-SERISE USER MANUAL Rev. G 6.3 Mobile Stand Name Description Arm of a stand to fix X-ray equipment or change its direction Height adjustment Handle or gearbox to adjust height of the equipment Gantry Gantry to fix arm or height adjustment component Rack to hold up X-ray device or stand components (wheels included for Rack carrying the equipment)
  • Page 57 EPX-SERISE USER MANUAL Rev. G The operator must keep the precautions when using mobile stand with device CAUTION Precautions for use ● Mobile stand arm should be 0° position when moving. ● Mobile stand casters should put on the brake when using. ●...
  • Page 58: Maintenance Procedure

    EPX-SERISE USER MANUAL Rev. G 7. MAINTENANCE PROCEDURE 7.1 General caution It is recommended to follow the maintenance procedure described below, for the reliable operation. The routine inspection should be committed by a trained expert There may cause serious injury by conducting unauthorized service or changing the instrument and manufacturer shall not be responsible for the resulting compensation.
  • Page 59 EPX-SERISE USER MANUAL Rev. G 7.4 Troubleshooting These are the information about indicating a system malfunction to do not appear in the message window. If the device occurs a problem to except on the below troubleshooting table, consult manufacture or the agent. The troubleshooting table does not put it for all problems. Problem Item Cause Action...
  • Page 60 EPX-SERISE USER MANUAL Rev. G 7.6 Maintenance Checklist MAINTENACE CHECKLIST MODEL: ___________________ SERIAL NUMBER: ___________________ Measuring tool: Oscilloscope (EP: 1V = 20kV, IP: 1V = 5.8mA) Dose Area Product Meter Checklist Verdict Check that mA accuracy within boundary of ±3%. Check that mAs accuracy within boundary of ±5%.
  • Page 61: Disposal Of Waste

    EPX-SERISE USER MANUAL Rev. G 8. DISPOSAL OF WASTE 8.1 Disposal of old Electrical & Electronic Equipment  This symbol indicates that this product shall not be treated as household waste. Instead, it shall be handed over to the applicable collection point for the recycling of electrical and electronic equipment. By ensuring this product is disposed of correctly, you will help prevent potential negative consequences for the environment and human health, which could otherwise be caused by inappropriate waste handling of this product.
  • Page 62: Quality Warranty

     Any product with a defect in appearance only shall not be returned to nor replaced by Ecotron Co.,Ltd. Ecotron Co.,Ltd. shall not be liable any incidental or consequential damages arising out of or relating to the use of the product.
  • Page 63: Intended Operator And Service Personal Profile

    - Local language Language understanding Maximum - Understanding of manual that is writing in English Minimum - Only authorized and properly trained by ECOTRON Experience Maximum - No maximum - Mild reading vision impairment or vision corrected to log MAR 0.2...
  • Page 64: Emc Declaration

    EPX-SERISE USER MANUAL Rev. G 11. EMC DECLARATION Guidelines and manufacturers: electromagnetic emission The Diagnostic X-ray System is used in the following electromagnetic settings. Users of the Diagnostic X-ray System should check whether their systems are used in these settings. Emission test Compliance Electromagnetic setting: guidelines Since the Diagnostic X-ray System only uses RF energy...
  • Page 65 EPX-SERISE USER MANUAL Rev. G The main power’s quality Voltage loss in the <5% UT(<95%Dip at <5% UT(<95%Dip at the UT), power supply, the UT), 0.5 cycles 0.5 cycles should be equal to the short 40% UT(60% Dip at the 40% UT(60% Dip at the UT), those of general intermittence and...
  • Page 66 EPX-SERISE USER MANUAL Rev. G Guidelines and manufacturers: electromagnetic tolerance The Diagnostic X-ray System is used in the following electromagnetic settings. Users of the Diagnostic X-ray System should check whether their systems are used in these settings. Tolerance IEC 60601 test level Suitability level Electromagnetic setting: guidelines test...
  • Page 67 EPX-SERISE USER MANUAL Rev. G Guidelines and manufacturers: electromagnetic tolerance It is very difficult to accurately predict the magnetic field strength of wireless (mobile/wireless) telephones, land mobile radio base station, amateur wireless, AM, FM wireless and TV broadcasting systems. To assess electromagnetic settings using fixed RF receivers, area walkdown is needed. If the magnetic field strength measured at the point where the Diagnostic X-ray System is used exceeds the applicable RF compliance level, you should check whether the Diagnostic X-ray System is operating normally.
  • Page 68: Appendix A: Technical Chart

    EPX-SERISE USER MANUAL Rev. G 12. APPENDIX A: TECHNICAL CHART Film Screen: Lanex Regular Screen (400 speed) 40" SID Adults Pediatric EXAMI- THICK- Dosimetry THICK- Dosimetry NATION Grid Grid NESS (μGy*㎡) NESS (μGy*㎡) 2-4cm HAND 8.828 4.498 5-7cm 2-4cm WRIST 9.792 4.498 5-7cm...
  • Page 69 EPX-SERISE USER MANUAL Rev. G L/SPINE(LAT) 32-34cm 590.1 238.72 12-14cm 102.8 PELVIS/HIP 20-22cm 228.25 40.944 26-28cm 374.48 12-16cm 11.92 CHEST(AP) 22-24cm 19.24 7.04 30-34cm 30.22 32-36cm 29.752 CHEST(LAT) 14.532 36-40cm 37.362 8-12cm 100.05 ABDOMEN 18-22cm 228.25 39.872 28-32cm 392.4...
  • Page 70 EPX-SERISE USER MANUAL Rev. G CR System: 40" SID Adults Pediatric EXAMI- THICK- Dosimetry THICK- Dosimetry NATION Grid Grid NESS (μGy*㎡) NESS (μGy*㎡) 2-4cm HAND 17.488 8.828 5-7cm 2-4cm WRIST 19.432 9.792 5-7cm 2-4cm 17.488 8.828 FOREARM 5-7cm 29.18 14.641 (AP) 8-10cm 34.1...
  • Page 71 EPX-SERISE USER MANUAL Rev. G 12-14cm 205.9 102.8 PELVIS/HIP 20-22cm 365.52 182.49 26-28cm 749.52 374.48 12-16cm 24.12 11.92 CHEST(AP) 22-24cm 38.76 19.24 30-34cm 60.7 30.22 32-36cm 60.192 29.752 CHEST(LAT) 36-40cm 75.4 37.362 8-12cm 200.35 100.05 ABDOMEN 18-22cm 365.52 182.49 28-32cm 788.24 392.4...
  • Page 72 EPX-SERISE USER MANUAL Rev. G DR System: Flap Panel with CSI Screen, 40" SID Adults Pediatric EXAMI- THICK- Dosimetry THICK- Dosimetry NATION Grid Grid NESS (μGy*㎡) NESS (μGy*㎡) 2-4cm HAND 8.828 4.498 5-7cm 2-4cm WRIST 9.792 4.498 5-7cm 2-4cm 8.828 FOREARM 5-7cm 14.641...
  • Page 73 EPX-SERISE USER MANUAL Rev. G 12-14cm 102.8 PELVIS/HIP 20-22cm 228.25 40.944 26-28cm 374.48 12-16cm 11.92 CHEST(AP) 22-24cm 19.24 7.04 30-34cm 30.22 32-36cm 29.752 CHEST(LAT) 14.532 36-40cm 37.362 8-12cm 100.05 ABDOMEN 18-22cm 228.25 39.872 28-32cm 392.4 The patient population can be divided into several subgroups which are defined according to age and approximate average height and weight measurements: ...
  • Page 74: Appendix B: Pediatrics And Small Patients

    EPX-SERISE USER MANUAL Rev. G 13. APPENDIX B: PEDIATRICS AND SMALL PATIENTS ECOTRON strongly suggests reducing radiation dose to As Low As Reasonably Achievable (ALARA) in all patients, especially pediatric and small patients, whenever it is determined that an x-ray is necessary.
  • Page 75 Protocols based on patient type; adult and pediatric patients. These protocols should be considered a baseline. ECOTRON strongly recommends that you work with your Radiologist and Physicist to determine the lowest possible dose for the desired image quality. Once patient size is selected, further adjustments to kVp, MAS, Filtration and Grid can be made to further minimize patient dose.

This manual is also suitable for:

Epx-f1600Epx-f2400Epx-f2800Epx-f3200Epx-f4000Epx-f5000