BSL HARDWARE GUIDE
The BSL Hardware Guide describes how to connect and setup various signal electrodes, transducers and other devices
for use with the Biopac Student Lab System using an MP acquisition unit and includes sections that detail different
applications and uses for the Biopac Student Lab PRO System.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the MP36 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for BIOPAC Systems MP36

  • Page 1 01.06.2014 BSL HARDWARE GUIDE The BSL Hardware Guide describes how to connect and setup various signal electrodes, transducers and other devices for use with the Biopac Student Lab System using an MP acquisition unit and includes sections that detail different applications and uses for the Biopac Student Lab PRO System.
  • Page 2 There are analog input channels (four on MP36/35 units, two on MP45), one of which can be used as a trigger input. The MP Unit must be connected to the computer and electrodes, transducers, and/or I/O devices must be connected to the MP Unit.
  • Page 3 Simple Sensor Connectors Regardless of the type of device connected, every sensor or I/O device connects to the MP36/35 using a “Simple Sensor” connector. Simple Sensor connectors are designed to plug only one way into the MP unit—no need to worry about plugging things in upside down or into the wrong socket! ...
  • Page 4 Back Panel Back Panel, MP36/35 The back panel of the MP36/35 has an analog output port, a USB port, an I/O Port, a Trigger Port, a DC input, a fuse holder, a power switch, and the unit’s serial number. The back panel of the MP45 has a USB cable and headphone port.
  • Page 5 Cleaning Procedures Before cleaning, be sure to unplug the power supply from the MP36/35 or unplug the MP45 USB cable from the computer. To clean the MP36/35, use a damp, soft cloth. Abrasive cleaners are not recommended as they might damage the housing.
  • Page 6 Complies with IEC60601-1 EMC complies with IEC60601-1-2 CE Marked Dimensions/Weight: MP36: 7 cm x 29 cm x 25 cm / 1.4 kg MP45: 3 cm x 18 cm x 10 cm / 0.3 kg Additional Specs MP36 Only Analog Output: Back panel DUSB 9m labeled “Analog Out”...
  • Page 7 CH 1, CH 2, CH 3, CH 4 Shield drive Shield drive 9 PIN FEMALE DSUB Vin+ Vin+ (1 of 4 for MP36/35 or 1 of 2 for MP45) Vin Vin Shield drive Shield drive +5 V (100 mA max aggregate) +5 V (50 mA max) ID resistor lead 1;...
  • Page 8 NOTE: This “Core Packages” table refers to page numbering from the Biopac Student Lab Catalog. See the catalog for information contained in any of these page references. BIOPAC Hardware | | Page 7 - 11 MP36/35/45 Updated: 12.23.2013...
  • Page 9 NOTE: This“Lesson Hardware Guide” table refers to page numbering from the Biopac Student Lab Catalog. See the catalog for information contained in any of these page references. BIOPAC Hardware | | Page 8 - 11 MP36/35/45 Updated: 12.23.2013...
  • Page 10 The following three pages of lesson descriptions refer to page numbering from the Biopac Student Lab Catalog. See the catalog for information contained in any of these page references. BIOPAC Hardware | | Page 9 - 11 MP36/35/45 Updated: 12.23.2013...
  • Page 11 HARDWARE GUIDE support@biopac.com www.biopac.com BIOPAC Hardware | | Page 10 - 11 MP36/35/45 Updated: 12.23.2013...
  • Page 12 HARDWARE GUIDE support@biopac.com www.biopac.com BIOPAC Hardware | | Page 11 - 11 MP36/35/45 Updated: 12.23.2013...
  • Page 13 BSL STIMULATORS Modular Stimulators (0-100 V): Low Voltage Stimulator/Adapter: BSLSTMB for MP36/36R/35 OUT3 Output Adapter for built-in Stimulator (MP36 only) BSLSTMA for MP30 SS58L Low Voltage Stimulator (MP35 only) See also: HSTM01, ELSTM1, ELSTM2, EL300S and EL400 electrodes. Lab set up note...
  • Page 14: Front Panel Terminology

    info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com Pulse rate (Hz) = 1000 / Repetition period (milliseconds) Also called — Pulse frequency Repetition rate Events per second Pulse Repetition Use when referring to either Pulse rate or Pulse period. Pulse level The amplitude of the pulse, expressed in Volts. The output of the BSLSTM is 0 Volts when the pulse is not active.
  • Page 15 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com level reads 5.1 Volts (Range 0-10V) or 51 Volts (Range 0-100V). As shown in the following diagram, the indicator mark is between the two dials. Close-up of “Level” adjustment knob Stimulus output Stimulus pulse output for connection to external electrodes or other devices. This is a standard BNC style connector.
  • Page 16 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com Calibration The “Reference Output” signal from the BSLSTM must be calibrated to ensure accurate results. 1. Choose the correct Preset (via MP3X menu > Setup Channels).  For example, if using the BSLSTMA/B, don’t choose a “BSLSTM...” Preset. Also, make sure the Preset matches the Voltage Range that will be used (0-10V, or 0-100V).
  • Page 17 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com CONNECTING THE BSLSTM TO THE MP3X 1) Turn the MP3X unit OFF. 2) Confirm that Power switch on the back of the BSLSTM is in the OFF position. 3) Set the Range on the front of the BSLSTM to 0-10V. 4) Set the Level to 1 Volt.
  • Page 18 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com BSLSTMA/B SPECIFICATIONS (This new unit uses digital display and a keyed range switch) Pulse width Controlled by: Computer, with lockable width limit Range: .049 – 100 milliseconds Resolution: 2 microseconds Accuracy: 5% (Can be improved to better than 2% using the “Correction factor” in the “Stimulator Preferences’...
  • Page 19 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com BSLSTM SPECIFICATIONS (This older unit uses dial reading and a flip range switch) Pulse width Controlled by: Computer, with lockable width limit Range: .2 – 100 milliseconds Resolution: 2 microseconds Accuracy: 5% (Can be improved to better than 2% using the “Correction factor” in the “Stimulator Preferences’...
  • Page 20 HARDWARE GUIDE support@biopac.com www.biopac.com LOW VOLTAGE STIMULATOR OUT3 The MP36 includes a built-in low voltage stimulator—just use the Analog Out port.  For connection to BIOPAC electrodes, add the OUT3 BNC Adapter. SS58L The MP35 uses the SS58L Low Voltage Stimulator to the Analog Out port.
  • Page 21 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com TIMULATOR ELECTRODE UIDELINES — PLEASE READ — It is very important to follow the electrode placement guidelines when connecting stimulator electrodes from the BSLSTM to a subject. The BSLSTM can output lethal levels of energy! ...
  • Page 22: Hardware Guide

    info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com SS1LA SHIELDED ELECTRODE ADAPTER The fully-shielded electrode interface cable permits high resolution recording of biopotential signals. The 3-meter adapter cable accepts standard Touchproof connectors. Use this lead adapter with:  LEAD120 for EL120 contact post electrodes or ...
  • Page 23 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com SS2L ELECTRODE LEAD SET  “SS2L” is used to reference SS2L, SS2LA, or SS2LB lead sets;  SS2LB is recognized by current release BSL Lessons. This fully shielded cable assembly permits high-resolution recording of biopotentials. Each lead set has three pinch leads designed to snap directly onto standard disposable electrodes (such as the EL500 series electrodes).
  • Page 24: Usage Recommendations

    info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com SS3LA EDA (ELECTRODERMAL ACTIVITY) TRANSDUCER The SS3LA transducer connects to a single MP3X input channel to record electrodermal activity (skin conductance or, with proper setup, skin resistance). Two Ag-AgCl electrodes are mounted in individual, ergonomically designed, polyurethane housings for improved contact. They attach to the fingers by a Velcro strap or can be taped to any other part of the body.
  • Page 25 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com SS3LA SPECIFICATIONS Electrode Type: Ag/AgCl, shielded Weight: 4.5 grams Range: .1 – 100 siemens (normal human range is 1 – 20 siemens) Cable Length: 2 meters Surface Area: 6mm contact area Connector Type: 9 Pin DIN Gel Cavity Area 1.66 mm Sterilizable: Yes (contact BIOPAC)
  • Page 26 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com PULSE PHOTOPLETHYSMOGRAM TRANSDUCERS  TSD200 for MP150/MP100 System  SS4LA for MP3X and MP45 System The TSD200/SS4LA consist of a matched infrared emitter and photo diode, which transmits changes in blood density (caused by varying blood pressure) in specific body locations. When the TSD200 is attached to the skin, the infrared light is modulated by blood pulsing through the tissue below.
  • Page 27 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com TSD200/200C/SS4LA SPECIFICATIONS 860 nm  60 nm Emitter/Detector Wavelength: Optical Low Pass Filter Cutoff Wavelength: 800 nm Note The operational range of the emitter and detector fall within the wavelength range of 800 nm to 920 nm. The filter is placed over the receiver, the filter of 800 nm is an optical lowpass, so wavelengths longer than 800 nm will pass thru.
  • Page 28 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com SS5LB RESPIRATORY EFFORT TRANSDUCER The SS5LB transducer is used to record respiration via chest or abdomen expansion and contraction. This transducer is useful for determining how deeply someone is breathing and for calculating the person’s breathing rate or respiration rate.
  • Page 29 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com TEMPERATURE TRANSDUCERS SS6L: Fast Response SS7L: Waterproof Probe SS8L: Liquid Immersion Probe SS18L Digit Surface SS6L TEMPERATURE TRANSDUCER The SS6L is a small fast-response thermistor used to measure small variations in temperature, either on the skin surface or in exhaled airflow. The recorded temperature changes during breathing can be used to indicate respiration rate.
  • Page 30 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com SS8L LIQUID IMMERSION PROBE Use this stainless steel probe for dry or wet bath temperature measurements. SS8L SPECIFICATIONS Response time: 3.6 sec Max operating temp: 60°C Accuracy & Interchangeability: ±0.2°C Compatibility: YSI(r) series 400 Dimensions: 4 mm X 115 mm Cable: 3 meters...
  • Page 31 0 Volts, and when the button is pressed it reads +5 mV. SS10L SPECIFICATIONS Cable Length: 2 meters Connector Type: 9 Pin DIN to MP36/35 front panel input BIOPAC Hardware | SS10L | Page 1 - 1 Updated: 7.1.2013...
  • Page 32 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com MEDIUM-FLOW PNEUMOTACH TRANSDUCER  SS11LA for MP3X and MP45 System  TSD117 & TSD117-MRI for MP150/MP100 System  RX117 Replacement Airflow Head  See also: AFT series of accessories for airflow and gas analysis These medium-flow airflow transducers are designed to measure human subject respiratory, bi-directional airflow (liters/sec) and can be used to measure respiratory flow in a wide range of tests and conditions relating to airflow and lung volume.
  • Page 33 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com Use the following measurement procedure for determining lung volume: 1. Breathe normally for 3 cycles (start on inspire) 2. Inspire as deeply as possible 3. Return to normal breathing for 3 cycles 4. Expire as deeply as possible 5.
  • Page 34 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com To remove residual offset after the flow data has been collected, select a portion of the baseline (zero flow reading) and calculate the mean value using the popup measurements. Subtract this mean value from the raw data to obtain a mean corrected flow signal.
  • Page 35 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com 4. Pull down the Presets pop-up menu and select Airflow. 5. Click on the View/Change Parameters button. 6. Click on the Scaling button. 7. Click on Cal1: Leave the Input value reading and enter 0 for the Scale value. 8.
  • Page 36 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com Proper handling of the Calibration Syringe Assembly 3. Pump the plunger several times before the recording. Always pull and push the plunger all the way until it stops when using the syringe. This assures that the full volume of air (0.6 liter) flows in and out of the airflow transducer.
  • Page 37 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com 2) Breathe through the airflow Hints for obtaining optimal data: transducer, following the proper a) Keep the Airflow Transducer upright at all times. procedure defined to the right. b) Always insert on and breathe through the side of the SS11LA airflow transducer labeled “Inlet.”...
  • Page 38 SS11LA & TSD117 Technical Specifications TRANSDUCER: TSD117 TSD117-MRI SS11LA Interface: DA100C MECMRI-DA to DA100C MP36/35/30/45 Flow Rate: ±300 Liters/min (±5 liters/sec) 10 liters/sec (highest linearity (±5 liters/sec) Nominal Output: 60 µV/[liters/sec] (normalized to 1 V 60 µV/[liters/sec] excitation) ¼” 25 TPI mounting nut:...
  • Page 39 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com See also: Force Transducer Tension Adjuster (HDW100A) SS12LA VARIABLE RANGE FORCE TRANSDUCER SS12LA Variable Range Force Transducer SS12LA Sample Setup Force transducers are devices capable of transforming a force into a proportional electrical signal. The SS12LA variable range force transducer element is a cantilever beam load cell incorporating a thin-film strain gauge.
  • Page 40 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com SS12LA SPECIFICATIONS* Lever Arm Position Full Scale Range 10Hz Noise 1Hz Noise (hook ring) (FSR) 50 grams 50 grams 2.5 mg 1 mg 100 grams 100 grams 5 mg 2 mg 200 grams 200 grams 10 mg 4 mg 500 grams...
  • Page 41 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com Gain set too high — Clipped Force data e) If the signal is clipped, decrease the Gain setting by one step (e.g., from x5000 to x2500) and collect new data at the lower gain setting. ...
  • Page 42: Pressure Transducer Calibration

    info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com SS13L PRESSURE TRANSDUCER The SS13L pressure transducer is used to measure direct arterial or venous blood pressure in animals or to record pressure changes within a closed system such as an organ or tissue bath system. Connect to the tubing via the standard rotating Luer-lok fittings.
  • Page 43 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com Note: In this sample dialog, the transducer was brought to a pressure of 50 mmHg and the Cal 1 button was pressed. The transducer was then brought to a pressure of 180 mmHg, and Cal 2 was pressed. b) Bring the transducer to the lowest expected pressure.
  • Page 44 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com RX104A REPLACEMENT ELEMENT The RX104A is a replacement element for the SS13L Pressure Transducer. It does not include the Smart Sensor connector and cable. BIOPAC Hardware | SS13L – RX104A | Page 3 - 3 Updated: 1.31.2013...
  • Page 45 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com See also: Tension Adjuster (HDW100A) SS14L DISPLACEMENT TRANSDUCER For use in recording very slight movements in a range of physiological preparations, the SS14L incorporates a semi-isotonic strain gauge and a stainless steel lever that can be mounted in any position. SS14L SPECIFICATIONS Sensitivity Range: 1mm to 100mm...
  • Page 46 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com TSD108 AND SS17L PHYSIOLOGICAL SOUNDS TRANSDUCER (CONTACT MICROPHONE) The physiological sounds transducer connects to the DA100C amplifier (TSD108) or the MP3x/4x hardware (SS17L). The transducer can be used with the Noninvasive Blood Pressure Cuff or as a stand-alone device. If used with the cuff, Korotkoff sounds can be recorded for easy determination of systolic and diastolic blood pressure.
  • Page 47: Blood Pressure Measurement

    info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com BLOOD PRESSURE CUFF AND TRANSDUCER  TSD120 for MP150/MP100 System  RX120 Series Cuff for TSD120  SS19L/LA for MP3x & MP45 System TSD120 RX120A and RX120F cuff options BLOOD PRESSURE MEASUREMENT The most common form of indirect blood pressure measurement employs a pressure cuff, pump and pressure transducer.
  • Page 48 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com CALIBRATION A) TSD120 The TSD120’s built-in pressure transducer will require an initial calibration prior to use. To calibrate the transducer, wrap the cuff into a roll and begin to inflate the cuff slowly with the pump bulb. The pressure change will be noticeable on the mechanical indicator.
  • Page 49 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com Example in BSL 4 – initial scaling dialog: Clicking CAL 2 results in an Input value of 0.071 mV. Adding 0.071 to the initial value of 61.44 results in an adjusted CAL 1 value of 61.51 mV. (Your result may vary slightly from the example).
  • Page 50 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com RX120 SERIES BLOOD PRESSURE CUFFS The RX120 series are optional blood pressure cuffs, of varying sizes, which can be quickly and easily swapped in and out of the noninvasive blood pressure cuff transducer. Use a single transducer and substitute one cuff for another to accommodate a wide range in limb circumferences.
  • Page 51 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com Goniometers & Torsiometers   TSD130 Series BN-GON-XDCR Series   SS20L-SS24L BN-TOR-XDCR Series   SS20-SS24 BN-GON-F-XDCR In the example above, the TSD130A was connected directly to a DA100C amplifier, the DA100C gain was set to 1,000, and AcqKnowledge was used to calibrate the signal to provide angular measurements from approximately +90°...
  • Page 52 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com ATTACHMENT TO THE SUBJECT Various combinations of display and recording instrumentation have been carefully developed fulfilling the requirements of specific research applications. Due to the wide range of applications, one method of attachment cannot be recommended.
  • Page 53: Maintenance And Service

    No periodic maintenance is required to ensure the correct functioning of the sensors. The sensors contain no user serviceable components. If the sensor fails, it should be returned to BIOPAC Systems, Inc.  Please request a Return Merchandise Authorization (RMA) number before returning the sensor and include a description of what has been observed and what instrumentation was in use at the time of sensor failure in the return package.
  • Page 54 HARDWARE GUIDE support@biopac.com www.biopac.com Specifications Part # MP1XX via DA100C TSD130A TSD130B TSD130C TSD130D TSD130E Telemetry TEL100C SS20 SS21 SS22 SS23 SS24 MP36/36R/35/30/45 SS20L SS21L SS22L SS23L SS24L BioNomadix via BN-GONIO BN-GON-110- BN-GON-150- BN-TOR-110- BN-TOR-150- BN-GON-F- XDCR XDCR XDCR XDCR XDCR...
  • Page 55 OVERVIEW OF THE BIOPAC GONIOMETER SERIES As with all measuring equipment, to correctly interpret the data, understanding the working principles (i.e., what the sensor measures) before use is helpful. BIOPAC Systems, Inc. manufactures three types of sensors: The single axis finger goniometer permits the measurement of angles in one plane.
  • Page 56 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com The single axis torsiometers permit the measurement of rotation in one plane, e.g. forearm pronation/supination. Axial rotation of one endblock relative to the other along axis Z-Z is measured from the gray plug. If the torsiometer is bent in planes X-X or Y-Y, the output remains constant.
  • Page 57 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com SIGN CONVENTIONS The sign convention for certain joints will differ, depending which side of the body the sensor is attached to. The following figures show sign conventions for the most common joints. BIOPAC Hardware | Gonio-Torsio | Page 7 - 9 Updated: 11.21.2013...
  • Page 58 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com WRIST – Goniometer TSD130A/SS20L/SS20/BN-GON-110-XDCR Attach the telescopic endblock to the back of the hand, with the center axis of the hand and endblock coincident (top of figure — viewed in the frontal plane). While fully flexing the wrist (middle and bottom of figure), extend the goniometer to Position 2 (as shown on page 2) and attach the fixed endblock to the forearm so that when viewed from the dorsal plane, the axes of the forearm and endblock are coincident.
  • Page 59 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com KNEE – Goniometer TSD130B/SS21L/SS21/BN-GON-150-XDCR Mount the telescopic endblock laterally on the leg so the axes of the leg and endblock coincide, when viewed in the sagittal plane. With the leg fully extended in the position of reference, extend the goniometer to Position 2 (maximum length, as shown on page 2) and attach the fixed endblock to the thigh so the axes of the thigh and endblock coincide.
  • Page 60 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com SS25LA HAND DYNAMOMETER Use the hand dynamometer to measure grip force—use in isolation or combine with EMG recordings for in-depth studies of muscular activity. The lightweight, ergonomically designed transducer provides direct readings in kilograms or pounds. The simple calibration procedure makes this device easy to use for precise force measurements, and the isometric design improves experiment repeatability and accuracy.
  • Page 61 0-90 Kg Dimensions: 17.78 cm (long) x 5.59 cm (wide) x 2.59 cm (thick) Nominal Output: 13.2 µV/kg for 1 V excitation Max excitation for MP36/36R is 5 V, for nominal output 66 µV/kg Rated Output: 100 kg Linearity: Sensitivity: 0.75 kg...
  • Page 62 Condition: Trace metallic parts do not make contact to the subject; must be used with MRI cables provided. Equipment • The SS26LB/SS27L accelerometers connect to the MP36/35 Data Acquisition Unit. • The TSD109 series accelerometers connect to the HLT100C High Level Transducer module.
  • Page 63 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com IMPORTANT Make sure the selected channel is not already assigned to any other BIOPAC module; up to 5 Accelerometers can be used with a single MP System. If contention exists, the channel data will be corrupted. See also: Setup notes for external devices and channel contention issues.
  • Page 64 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com BIONOMADIX WIRELESS ACCELEROMETER The BioNomadix wireless Tri-axial Accelerometer (BN-ACCL3) is a broad spectrum acceleration measurement system. The transmitter can be attached to any part of the subject's body to measure three-axis acceleration associated with movement in that particular location. The system comes factory preset to support an operational range of ±16 G, with a maximum system bandwidth of 400 Hz.
  • Page 65 18.3 mm (dia) x 0.36 mm (thick) and 30 cm pigtail lead FSR Active Area: 12.7 mm (dia) Cable Length: 7.6 m Cable Length – BN-STRIKE-XDCR: 30 cm Interface: SS28LA MP36/35 System TSD111A HLT100C SS28A TEL100C BN-STRIKE-XDCR BN-STRIKE RX111 REPLACEMENT HEEL-TOE STRIKE SENSOR Replacement strike sensor for Heel/Toe Strike transducers.
  • Page 66 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com SS29L MULTI-LEAD ECG CABLE The SS29L Multi-Lead ECG Cable permits high- resolution ECG recordings. This multi-lead set can simultaneously record Leads I, II, III, aVR, aVL, aVF, plus one precordial chest lead V(1-6). A 12- Lead ECG recording can be obtained by alternating the chest lead electrode from position V1 through V6.
  • Page 67 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com SS30L ELECTRONIC STETHOSCOPE TRANSDUCER The SS30L stethoscope was developed to teach the standard procedure for listening to heart sounds and Korotkoff sounds with a “normal” stethoscope, and record simultaneous sound data. A microphone in the SS30L records sound as it is heard and the BSL software displays the sound wave during and after recording (a variety of acoustical signals can be recorded).
  • Page 68 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com SS31L NONINVASIVE CARDIAC OUTPUT SENSOR The SS31L records the parameters associated with Cardiac Output measurements. The SS31L incorporates a precision high-frequency current source, which injects a very small (400µA rms) current through the measurement tissue volume defined by the placement of a set of current source electrodes.
  • Page 69 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com SS36L REFLEX HAMMER This is a classic reflex hammer with a transducer attached to perform reflex measurements. It uses a Taylor Hammer—the most common type of reflex hammer used by doctors and nurses—and incorporates electronics to record the time and the relative strength of the impact.
  • Page 70 (with schematics and design notes) that demonstrate a very important subset of circuit design for recording and processing physiological signals. Students will use the MP36/35 and BSL PRO software to evaluate their designs. See Lessons H25, H26. Circuitry Hardware ...
  • Page 71 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com SS40L – 42L DIFFERENTIAL PRESSURE TRANSDUCER SS40L ±2.5cm H SS41L ±12.5cm H SS42L ±25cm H The SS40L-SS42L series differential pressure transducers are designed for low range pressure monitoring. The transducers plug directly into the MP3X general- purpose differential amplifier.
  • Page 72 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com SS43L VARIABLE ASSESSMENT (PSYCH) TRANSDUCER Use this handheld, slide control transducer to record subjective responses to a variety of different stimuli. Use multiple transducers to allow several people to simultaneously answer the same question or otherwise respond to stimuli.
  • Page 73 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com PNEUMOTACH AIRFLOW TRANSDUCERS  TSD137 SERIES FOR MP150/MP100 SYSTEM  SS46L-SS52L SERIES FOR MP3X AND MP45 SYSTEM  RX137 SERIES REPLACEMENT FLOW HEADS The TSD137/SS46L-SS52L series pneumotachs RX137A can be used to perform a variety of small animal TSD160 and human pulmonary measurements relating to SS40L...
  • Page 74 TSD/RX137 & SS46L-SS52L SERIES SPECIFICATIONS Part# DA100C TSD137B TSD137C TSD137D TSD137E TSD137F TSD137G TSD137H TSD160/SS40L RX137B1 RX137C1 RX137D1 RX137E1 RX137F1 RX137G1 RX137H1 MP36/35/30/45 SS46L SS47L SS48L SS49L SS50L SS51L SS52L Range (ml/sec): ±50 ±83 ±166 ±583 ±1565 ±2666 ±13333 Dead Space (cc): 18.15 13.87...
  • Page 75 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com PNEUMOTACH 200 SERIES AIRFLOW TRANSDUCERS These flow transducers are designed for humans and animals ranging in size from mice to medium-sized dogs. They include a detachable flow head (RX237B through H) and a differential pressure transducer (TSD160A or SS40L).
  • Page 76 Monitor switch data as a digital input channel. Independently mark events, or provide responses, on Connects to the digital input on the MP36/35 only. up to eight channels simultaneously. Assign separate digital channels as event markers for individual SS53L Hand switch analog input channels.
  • Page 77: Specifications

    info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com SS56L HAND CLENCH FORCE BULB SS56L measures proportionality of bulb pressure to clench force in “kgf/m^2” units (a pressure unit). This measure is accurate for the relative measures recorded in BSL Lesson 2 Electromyography (EMG) II. SS56L is recognized by current release BSL Lessons.
  • Page 78 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com SS57L EDA LEAD FOR DISPOSABLE SETUPS Snaps to two EL507 disposable EDA (isotonic gel) electrodes. This disposable setup is an alternative to the reusable SS3LA EDA (GSR) Transducer. Range: 0.1-100 µMho (normal human range is 1-20 µMho) Excitation: 0.5 V DC Pinch Leads:...
  • Page 79 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com SS60L SIGNAL CABLE FOR SS39L BREADBOARD Use this signal cable to add signal inputs to the SS39L Signal Processing Breadboard, which ships with one combination power/signal cable. BIOPAC Hardware | SS60L | Page 1 - 1 Updated: 7.17.2012...
  • Page 80 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com SS61L FINGER TWITCH TRANSDUCER Palmar attachment recommended: “UP” label facing out “UP” label toward skin for posterior (dorsal) attachment Use this transducer to record finger twitch responses from human subjects receiving electrical stimulation (using the HSTM01). The transducer conforms to the shape of the finger and attaches via a Velcro® strap and tape. SPECIFICATIONS Transducer Dimensions: 14.6 cm (long), 0.50 cm (wide...
  • Page 81 HARDWARE GUIDE support@biopac.com www.biopac.com SS62L SPEECH FREQUENCY MICROPHONE Use this precision microphone for speech frequency analysis and other acoustic studies. For use with the MP36/35 only, requires continuous high-speed sample rate. BIOPAC Hardware | SS62L | Page 1 - 1...
  • Page 82 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com SS63L – SS66L FORCE TRANSDUCER SERIES SS63L Force Transducer - 50 g SS64L Force Transducer - 100 g SS65L Force Transducer – 200 g SS66L Force Transducer - 500 g SS63L – SS66L SPECIFICATIONS Noise: with 10 Hz LP filter: 2.5 mg with 1 Hz LP Filter: 1.0 mg...
  • Page 83 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com SS67L PRESSURE PAD/RESPIRATION TRANSDUCER The SS67L consists of an SS41L differential pressure transducer, RX110 pressure pad, and tubing. The multipurpose pressure pad/respiration transducer can be used to: 1. Noninvasively measure respiration—from a small mouse to a human. 2.
  • Page 84 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com SS68L PH PROBE TRANSDUCER The SS68L probe transducer can measure pH within the range of 0-14. The electrode provides approximately a single digit pH value change for every 5 mV change in the electrode reading, either positive or negative depending on whether the pH is above 7 or below it.
  • Page 85 Order interface only as BSL-TCI16 SS69L Components The SS69L transducer measures dissolved oxygen. The SS69L includes a dissolved oxygen probe and an interface to the BSL MP36/MP35 or MP30 unit.  See BSL PRO Lesson #A07 Fish Respiration and Q10.
  • Page 86 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com Calibration — optional  Calibration is optional. To measure relative change, probe calibration is not essential. To improve accuracy for discrete measurements, probe calibration is recommended. a. Zero-Oxygen (CAL 1) i) Launch the BIOPAC software and generate the scaling dialog for the probe channel. (Select MP menu >...
  • Page 87 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com TABLE 1 Dissolved O2 (mg/L) in air-saturated distilled water (at various temp. & pressure) 14.76 14.59 14.38 14.19 13.00 13.80 13.61 13.42 13.23 13.04 12.84 12.65 12.46 0C 14.38 14.19 14.00 13.82 13.63 13.44 13.26 13.07 12.88 12.70...
  • Page 88 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com CONVERSION FORMULA FOR % SATURATION Set CAL 1 Input to 0% and CAL 2 Input to 100%, and then use the following formula to enter the Values: % Saturation = (actual DO result / Saturated DO value from Table 1) x 100 For example, if the probe result is 6.1 mg/L at a temperature of 20ºC and a pressure of 740 mmHg, the corresponding Table 1 value is 8.93 mg/L, so % Saturation = ( 6.1 / 8.93) x 100 = 68%...
  • Page 89 SS9LA has a built-in divide by 10 attenuation which provides a ±20 V input range on MP36 or MP36R and a ±10 V input range on MP35 or MP45. The 2-meter cable terminates in a male BNC for easy connections.
  • Page 90 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com See also: Student Accessory Pack BSL-ACCPACK AFT SERIES AIRFLOW & GAS ANALYSIS ACCESSORIES Includes the following airflow accessories: AFT1 AFT9 AFT11E AFT15 AFT26 AFT2 AFT10 AFT11F AFT20 AFT30 AFT3 AFT10S AFT11G AFT21 AFT31 AFT4 AFT11A AFT11H AFT22 AFT35...
  • Page 91 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com MOUTHPIECES MRI Use: MR Safe AFT Mouthpiece Components: Polyethylene EVA Copolymer, Thermoplastic Rubber, Polycarbonate Plastic AFT2 Disposable Mouthpiece Available in Packs of 10 or 250 22 mm OD; connects to the SS11LA or TSD117 via the AFT1. AFT8 Autoclavable Mouthpiece Available in Packs of 1 or 10...
  • Page 92 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com FACEMASKS, FACEMASK ACCESSORIES AFT10 Disposable Adult Facemask These mouthpieces connect to 22 mm breathing circuits. Connects directly to the AFT1, AFT22 non- rebreathing T-valve, or SS11LA/TSD117 airflow transducer (via AFT11B coupler). Includes hook-ring to secure AFT10S adjustable head strap. (22 mm ID/25 mm OD) MRI Use: MR Safe AFT10 Facemask Components: Thermoplastic Elastomer, Polyvinyl Chloride (PVC) Plastic...
  • Page 93 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com AFT11 Series Coupler Guide Item 1 Item 2 Coupler Item 1 Item 2 Coupler 15 mm OD 22 mm ID AFT11B 25 mm ID 25 mm ID AFT11C 20 mm OD 22 mm ID AFT11B 25-30 mm OD 25-30 mm OD AFT11A...
  • Page 94 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com The AFT20 connects the CO2100C or O2100C directly to the sampling port of a mixing chamber. The AFT20 also permits sampling connections to the Non-rebreathing “T” Valves (AFT21 or AFT22). MRI Use: MR Safe AFT20 Gas Sampling Kit Components: Tubing: 1.8 m of 1.5 mm diameter polyethylene tubing with ...
  • Page 95 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com AFT T-VALVES AFT21 Non-Rebreathing “T” Valve: Female, 35 mm High performance, very low dead space, low airflow resistance valve, suitable for high airflow applications (e.g. exercise physiology). The non-rebreathing “T” valve incorporates a Female Luer connector gas sampling port for interfacing with the AFT20.
  • Page 96 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com PART SUMMARY FOR TYPICAL AIRFLOW / GAS ANALYSIS APPLICATIONS PULMONARY FUNCTION High Flow Med. Flow Low Flow Very Low Flow Exercising Resting human Child, Pig, Dog Small Animals Part # human AFT2 Mouthpiece AFT3 Noseclip AFT6A Calibration Syringe AFT7/7L Tubing X (2)
  • Page 97 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com TRANSDUCER CONNECTOR INTERFACES (TCI) BSL-TCI13 Interface MP30 to Piezo BSL-TCI0 Interface Grass BSL-TCI14 Interface 1/4 phono BSL-TCI1 Interface Beckman 5 pin BSL-TCI15 Interface 5-pin DIN Vernier BSL-TCI2 Interface WPI BSL-TCI16 Interface BT connector, Vernier BSL-TCI3 Interface Lafayette BSL-TCI17 Interface 5-pin Intellitool BSL-TCI4 Interface Honeywell BSL-TCI18 Interface 2 mm Hg Strain...
  • Page 98 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com BSL-TCI PIN-OUTS BSL-TCI3 BSL-TCI7 BSL-TCI8 BSL-TCI14 BSL-TCI9 BSL-TCI15 BSL-TCI10 Gould 12-pin: discontinued BSL-TCI11 BSL-TCI16 BSL-TCI12 BSL-TCI17 BIOPAC Hardware | BSL-TCI | Page 2 - 3 Updated: 1.31.2013...
  • Page 99 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com BSL-TCI PIN-OUTS BSL-TCI13 BSL-TCI18 BSL-TCI19 BSL-TCI21 BSL-TCI20 BSL-TCI22 BIOPAC Hardware | BSL-TCI | Page 3 - 3 Updated: 1.31.2013...
  • Page 100 BSLCBL11: Stimulator to Electrode – BNC to 2x Electronic Test Clip (spring-loaded) BSLCBL12: Stimulator to Electrode – BNC to 2x Toothless Alligator BSLCBL8/9: High Impedance – 1.5 mm Touchproof BSLCBL14: MP36/35 Input Adapter for Research Amplifiers INTERFACE CABLES BSLCBL1A BSLCBL2A...
  • Page 101 Each 1-meter cable has two clips And terminates with one BNC connector to interface with the BSLSTM, SS58L Stimulator, or OUT3 for MP36 low volt stimulator and silver or platinum wire electrodes. High-impedance cables These fully-shielded, high-impedance electrode interface...
  • Page 102 Biopotential or Transducer amplifier modules. Also used to connect a ground electrode lead (e.g. LEAD110A) to the UIM100C module (required when using the TSD150 active electrodes). One CBL201 is required for each Touchproof socket. For MP36/35/45 Systems CBL201 is used to update older model SS1L Shielded Lead Adapters. CBL202 2 mm Male pins Adapts transducers with a 6.3 mm (¼) mono phone plug to the DA100C.
  • Page 103 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com CBL206 Lead junction TPF to 4X TPM Reference four electrodes from one. Connect via the MEC110C to the NICO100C and EBI100C cardiac output amplifier modules CBL207 1 m, BNC (m) to 2 x 1.5 mm TP (m). Use with ...
  • Page 104: Usb Cable

    info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com CBLUSB USB CABLE 2.5 meter replacement USB cable connects the MP3X to a USB port. Includes and provides EMI protection to maintain BSL Systems certified safety rating (CE, EMC). BIOPAC Hardware | CBLUSB | Page 1 - 1 Updated: 1.30.2013...
  • Page 105 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com HSTM01 HANDHELD HUMAN-SAFE STIMULATING PROBE The HSTM01 handheld human-safe stimulating probe provides a superior degree of safety and comfort when using the Biopac Student Lab Stimulator for human stimulation. The ergonomic design allows the subject to focus on electrode placement instead of worrying about holding the electrode.
  • Page 106 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com ELECTRODES In selecting the application site for any style of electrode, care should be taken that: 1) Electrode site is dry and free of excessive hair. 2) Electrode is not placed over scar tissue or on an area of established erythema or with a lesion of any kind. 3) Skin is properly prepared.
  • Page 107 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com ELECTRODES In selecting the application site for any style of electrode, care should be taken that: 1) Electrode site is dry and free of excessive hair. 2) Electrode is not placed over scar tissue or on an area of established erythema or with a lesion of any kind. 3) Skin is properly prepared.
  • Page 108 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com ELECTRODES In selecting the application site for any style of electrode, care should be taken that: 1) Electrode site is dry and free of excessive hair. 2) Electrode is not placed over scar tissue or on an area of established erythema or with a lesion of any kind. 3) Skin is properly prepared.
  • Page 109 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com  All EL250 Series electrodes require adhesive disks (ADD200 series) and recording gel (GEL1 or the preferred recording gel). See the Electrode Accessories section for further description. Instructions for EL250 Series Electrodes 1) Store electrodes in clean, dry area. 2) After use, clean electrode with cold to tepid water a) DO NOT use hot water.
  • Page 110 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com ELECTRODES In selecting the application site for any style of electrode, care should be taken that: 1) Electrode site is dry and free of excessive hair. 2) Electrode is not placed over scar tissue or on an area of established erythema or with a lesion of any kind. 3) Skin is properly prepared.
  • Page 111 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com ELECTRODES In selecting the application site for any style of electrode, care should be taken that: 1) Electrode site is dry and free of excessive hair. 2) Electrode is not placed over scar tissue or on an area of established erythema or with a lesion of any kind. 3) Skin is properly prepared.
  • Page 112: Disposable Electrodes

    info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com EL500 SERIES DISPOSABLE ELECTRODES The EL500 Series snap electrodes provide the same signal transmission as BIOPAC’s reusable EL504 electrodes, with added EL510 convenience and hygiene. Each peel-and-stick EL508 electrode is pre-gelled and designed for one use only. Use disposable snap electrodes with BIOPAC’s SS2L electrode lead set.
  • Page 113 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com Part Ag-AgCl Adhesive/Disposable Electrode Type EL506 This unique disposable strip electrode is designed for bioimpedance applications. The electrode is silver laminated on medical grade cloth, with industry-standard medical grade adhesive, medium Alternative tackiness. The silver-silver chloride (Ag-AgCl) electrode provides accurate and clear transmission band of surface biopotentials and is latex free.
  • Page 114 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com Part Ag-AgCl Adhesive/Disposable Electrode Type EL510 is a disposable RT electrode set of three electrodes with hydrogel centers and hydrocolloid EL510 ends that terminate in Touchproof leads. Each box includes 20 sets of 3 electrodes. ...
  • Page 115 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com ELECTRODES In selecting the application site for any style of electrode, care should be taken that: 1) Electrode site is dry and free of excessive hair. 2) Electrode is not placed over scar tissue or on an area of established erythema or with a lesion of any kind. 3) Skin is properly prepared.
  • Page 116 ELSTM2 SPECIFICATIONS Needle Length: 2.5 cm Needle Diameter: 0.3 mm Cable length: 2.5 m Connector type: Interface: BSLSTM Stimulator or SS58L for MP35 or OUT3 for MP36 BIOPAC Hardware | ELSTM Series | Page 1 - 1 Updated: 5.14.2013...
  • Page 117 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com ELECTRODE ACCESSORIES & GELS Abrasive Pads Before applying electrodes, abrade the skin lightly with an ELPAD (above left) to remove non-conductive skin cells and sensitize skin for optimal adhesion. Each ELPAD package contains 10 abrasive pads. Adhesive Use adhesive tape for attaching Active Electrodes and other devices.
  • Page 118 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com GEL104 Electrode Gel - salt free - 250 g (8.5 oz). SPECTRA 360® electrode gel.The only salt-free and chloride-free electrically conductive gel, recommended for all electromedical procedures. Salt-free characteristics make it particularly suitable for long-term applications. Spectra 360 differs significantly from all other electrically conductive media...it works by wetting the skin, thereby reducing skin resistance.
  • Page 119 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com BSL-ACCPACK The BSL Accessory Pack includes the consumable items to run 16 BSL Lessons. Make students accountable for their own lab equipment and reduce the burden on department budgets. School bookstores can purchase the BSL Accessory Packs and sell them to students. Includes: 60 x EL503 Disposable Electrodes 10 x EL507 Disposable EDA (GSR) Electrodes (ten electrodes total) AFT1 Disposable Bacterial Filter...
  • Page 120 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com ELECTRODE LEADS LEAD108 SERIES — MRI-CONDITIONAL/RADIOTRANSLUCENT LEADS FOR EL508/EL509 Use the LEAD108 series with EL508 MRI-conditional, radiotranslucent electrodes and EL509 disposable radiotranslucent dry electrodes. All LEAD108 Series terminate in 1.5 mm female Touchproof sockets. MRI Usage: MR Conditional Condition: Tested 3T-9T...
  • Page 121 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com Important Usage Notes:  If using multiple biopotential modules, do not connect the ground (GND) for the other modules — establish one ground per subject.  If using a GSR100C Electrodermal Response Amplifier with the EBI100C or the NICO100C, please note that the black I- (GND) connection will shunt current from the GSR100C excitation source.
  • Page 122 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com GASCAL CALIBRATION GAS AND GASREG REGULATOR Use the single stage, non-corrosive, general-purpose GASREG regulator with the GASCAL Calibration Gas Cylinder. Single-stage pressure regulators reduce the cylinder pressure to the delivery or outlet pressure in one step, and are generally good for short duration applications.
  • Page 123 MP System with a GASSYS2 module and some airflow accessories. The GASSYS2 connects directly to the MP36 System (GASSYS EA/EB) or via the UIM100C (GASSYS RA/RB) and requires two channels. The non-rebreathing T-valve directs only expired air to the GASSYS2. Because only expired air is directed to the module, the system acts to average respiratory outflows.
  • Page 124 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com Dimensions/Weight 1.7-liter chamber 5-liter chamber Length: 22 cm 55 cm Height: 18 cm 18 cm Width: 14 cm 14 cm Weight: 1.38 kg 1.96 kg GASSYS2 Software Scaling Defaults AcqKnowledge Software Default %O Scaling Default %CO Scaling BSL PRO Software Default %O...
  • Page 125 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com Cleaning the BIOPAC GASSYS2 Note: Never clean the sensor base of the device. The two sensors, a screen and a copper-colored gas detector, are highly sensitive. 1) Unscrew the top knob attachment. 2) Remove the plastic lid from the flow chamber. 3) Gently pull the clear cylinder off the sensor base.
  • Page 126 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com 5) Remove the chamber stand (gently pull back the chamber stand from the electronics base). 6) Clean the flow chamber with one of two methods: a. Use a soft cloth and Cidex Plus Sterilizing and Disinfecting Solution cleanser.
  • Page 127 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com TENSION ADJUSTERS HDW100A TENSION ADJUSTER HDW200A 3RD-PARTY TENSION ADJUSTER ADAPTER HDW100A and TSD125/SS12LA The HDW100A tension adjuster operates with the TSD105A, TSD125, SS12LA force transducers and SS14L displacement transducer. The rugged design and stability of the mounting allow for fine position control. The position adjuster is located on the top for easy access and smooth operation.
  • Page 128 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com MICROMANIPULATOR This manual micromanipulator is a reliable, durable, and economical solution for high-precision experiments.  Vernier scales allow readings to 0.1 mm  X-axis fine control allows readings to 10 µm  Includes tilting base ...
  • Page 129 The system outputs a continuous blood pressure waveform that is similar to a direct arterial pressure waveform. Monitor displays values for systolic, diastolic, mean blood pressure, and heart rate. Add BSL-TCI5 to use with an MP36 System. NIBP100D System includes: ...
  • Page 130 Deflation time: < 15 sec Protection against electric shock: Type BF Interface MP150/100 to DA100C via TCI105 (sold separately) MP36 via BSL-TCI5 (sold separately) Output Sensor bridge voltage: 2 - 10 V (external monitor) Sensitivity: 5 µV/V/mmHg BP Wave Out: CNAP™ transducer cable 0.3m; connector RJ11 6P4C (e.g. Abbott IBP catheter) Maximum delay of analog out signal: 80 msec ...
  • Page 131 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com Hyperbaric/Hypobaric Chamber Setup 1. Cuff controller and CNAP monitor must be in the same chamber with the same"pressure" environment as both are equipped with pressure sensor for surrounding pressure. 2. Pressure must be increased / decreased continuously rather than abruptly. 3.
  • Page 132 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com NON-INVASIVE SMALL ANIMAL TAIL BLOOD PRESSURE SYSTEMS NIBP250 Blood Pressure Amplifier NIBP200A Blood Pressure System NIBP Amplifiers with built-in pump automatically inflate the tail cuff to occlude the vessel in the tail of a rat or similar small animal, and then slowly deflate the cuff when the inflation point is reached, providing a linear drop in pressure.
  • Page 133 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com SPECIFICATIONS Cut-off Pressure Range: 100 – 300 mmHg (adjustable by 1mmHg steps) Pressure Accuracy: 300 mmHg Full Scale 1% Pressure Sensitivity: 0.1 mmHg Pressure Signal output: 300 mmHg/3 Volt DC Pulse Gain Levels: x1, x2, x4, x5, x8, x16, x32 (adjustable) Pulse Signal Output: 0- 4 Volt DC Pulse Display:...
  • Page 134 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com ANIMAL PREPARATION Optional Heating Chamber Restrainer Animal Holders Tail Cuff/Sensor 1. Turn the Animal Heating Chamber on. 2. Set the temperature value (press and hold P.Set and then press the up or down arrow to reach the desired value). ...
  • Page 135 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com SOFTWARE SETUP (AcqKnowledge 4.1 and higher) 1. Launch AcqKnowledge 2. Select the “Create/Record a new experiment” option. 3. Select MP150 > Set Up Channels > “Add New Module...” a. From the new module list, select UIM100C-A1 (or whichever channel CBL150-PRE pressure cable is connected to) and click “Add.”...
  • Page 136 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com Input volts Scale (Map) value Cal 1 Cal 2 mmHg Units Label: The scaling must be adjusted as the cut-off pressure switch settings are changed. If the pressure switch is set to 300 mmHg, then the settings should be: Input volts Scale (Map) value...
  • Page 137 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com Source = CH 1 Pos Edge Trigger Level = 0.33 (based on 1 V ≈ 100 mmHg) Delay = 0 9. Close out of Set up Triggering. RECORDING 1. Confirm that the animal is ready and that the IRSENSOR is attached to the tail.
  • Page 138 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com 5. On the graphical display, starting from the point of first pulse, select an area to the maximum. 6. Review the results for Max (Systolic), Min (Diastolic), and Mean measurements. Calculation of BPM Heart 1. Set a measurement for BPM. 2.
  • Page 139: Save Results

    info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com SAVE RESULTS - Previously saved results can be displayed by pressing the “Load” button under “Results.” - Placing the cursor on a desired measurement and pressing OK will load the recorded pressure, pulse curves and previously calculated results.
  • Page 140 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com NERVE CHAMBERS: NERVE1 AND NERVE2 These acrylic, desktop Nerve Chambers have 15 stainless steel pins for recording and stimulating a variety of different nerve preparations. Each stainless steel pin is spaced 5mm apart to provide a variety of recording and stimulating configurations.
  • Page 141 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com NERVE CHAMBER CONNECTIONS To connect the Nerve Chamber to MP-series Biopotential amplifiers (MCE100C, ERS100C, or EMG100C), use three JUMP100 connectors and three CBL200 adapter cables. Optionally, for additional lead length, use one MEC110C extension cable. 1.
  • Page 142 Cable Length: 2 meters Connector Type: 9 Pin DIN (female OUT1A These ultra-wide frequency response headphones connect directly to the headphone port on the MP36 or MP36R data acquisition unit. Features of these multi-purpose headphones include:  High dynamic range ...
  • Page 143 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com See also: SS9LA BNC Input Adapter OUT2 BNC (M) OUTPUT ADAPTER This BNC adapter is designed to output signals from the MP3X unit to other devices (such as external amplified speakers and scopes). This 2- meter adapter cable terminates in a male BNC for easy connections.
  • Page 144 BSL System stimulator (model BSLSTM): use BSLCBL6 and Radio Shack P/N 274-047 ¼" to 1/8" phono adapter • BSL MP36 data acquisition unit Analog Out port: use OUT3 plus BSLCBL6 and Radio Shack P/N 274- 047 ¼" to 1/8" phono adapter •...
  • Page 145 Connect a BSLCBL6 (interface cable: BNC to 3.5 mm) to the OUT3. c. Connect the OUT103 3.5 mm plug to the BSLCBL6 3.5 mm socket. d. Set MP36 > Output Control 'Low Voltage Stim' option  Set Pulse width to 100 msec.
  • Page 146 MP150 Starter system. AC300A +12 volt, 1.25 amp Connects the MP or GASSYSTEM2 to mains wall outlet. One supply is included with MP36/35 system, MP100 or GASSYSTEM2. See also: IPS100C Isolated Power Supply BIOPAC Hardware | In-line Power Supplies | Page 1 - 1 Updated: 12.19.2013...
  • Page 147 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com BAT100A RECHARGEABLE BATTERY PACK BAT100A with Recharger The BAT100A is a high energy density and lightweight battery pack designed to operate MP150 or MP3X Systems. A universal input voltage 3 amp battery charger is also included. The battery pack is lightweight and comes with a supplied carrying case with integral shoulder strap.
  • Page 148 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com BAT100A BATTERY PACK SPECIFICATIONS Battery Chemistry: LiFePO (Lithium Iron Phosphate) Output Capacity: 12V @ 15 amp-hours Working Output Voltage Range: 13.2 V – 12 V Output Connector: DC Barrel Plug (5.5 mm OD, 2.1 mm ID – Center positive) Operating Time: MP3X with 4 sensors: 26 hours nominal MP150 with 4 modules: 16 hours nominal...
  • Page 149 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com See also: TCI series of available interfaces SS-KIT-IN TRANSDUCER CONNECTOR INTERFACE KIT - INPUT This kit is for users who wish to adapt their own transducers to the Biopac Student Lab PRO System. The kit comes with a Smart Sensor connector, cable and components to properly interface with the transducers.
  • Page 150 Out “-“: pin 3 The “Analog Out” port on the back panel of the MP36/MP35 or MP30 (MP3X) can output pulses (digital) or analog voltage levels, or it can pipe out analog signals from one of the input channels. The port is controlled through one of the Output Control Panels in the Biopac Student Lab (BSL) PRO software, which is described in the BSL PRO manual.
  • Page 151: Assembly Notes

    Notes Pins 1 and 2 For the MP36/MP35, pins 1 and 2 can output analog or pulses when using MP3X firmware revision 1.26.037.030 or greater. When run under previous firmware, pins 1 and 2 can only be used for analog output. To identify the firmware revision, launch the BSL PRO software and check the Help >...
  • Page 152 The synchronization signal(s) coming from the STP35W can be directed to a BSL MP36/35 System running on a PC or a Macintosh, but it's not possible to run the STP35W on the same computer as the BSL MP36/35 System. The STP35W requires that the SuperLab software and a Digital I/O card (PCI slot required) be placed on a second computer.
  • Page 153 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com TISSUE BATH 1, 2, 4, 8 TISSUE BATH STATIONS The Tissue Bath Station is completely modular, and can be purchased in multiples of one unit. The System includes all of the glassware, tubing, reservoir, tissue hooks and mounting accessories, force transducer and micrometer tension adjuster.
  • Page 154 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com TISSUE BATH ACCESSORIES / REORDER PARTS Tissue Holders Tissue Clips Warming Coil Oxygen Filter Tissue Bath Reservoir Mount Accessories Field Stimulation Electrode RXHOLDER-S Tissue Holder (stainless steel) RXCOIL Warming Coil RXHOLDER-G Tissue Holder (glass) RXO2FILTER Oxygen Filter (glass) RXHOLDER-TR Triangle Tissue Holder (stainless) RXBATH...
  • Page 155: Error Codes

    info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com CIRCULATOR A/B HEATING CIRCULATORS Heating circulators are used with Tissue Bath Stations and include a digital temperature display and the following controls: Preset Temperature Power Heater Circulation Inlet Outlet ports are on the back, along with the power cord.
  • Page 156 info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com CIRCULATOR SETUP & USAGE GUIDELINES 1. Connect a hose from the INLET on the back of the circulator to the tissue bath OUTPUT.  For more than one tissue bath, connect the tissue baths serially. 2.
  • Page 157: Troubleshooting

    info@biopac.com HARDWARE GUIDE support@biopac.com www.biopac.com TROUBLESHOOTING  There is no water circulation or very little. 1. Check the hose connections and be sure they are connected to the correct positions. 2. Check that the hoses are not bent or twisted (which might impede the flow of water). 3.