4. Operation
4.7. Programing
The instrument is supplied with the default configuration specified in Start-up, page 15.
To modify the default configuration and adapt it to specific applications follow the diagram below.
Limit.
Select the relay values.
Input "ON" (closed contact).
Output "OFF" (open contact).
Conductivity: between 0.0 µS &
200 mS.
Salinity: between 0.0 mg/l &
278 g/l
* LIMIT 2 / CLEANING
The option LIMIT 2 will only appear
if relay 2 is configured as a limit
relay.
Alarm
Selection of the time lag before
activating the alarm. Select between
1 and 99 seconds, minutes or hours.
0=deactivated.
Cleaning
T on: cleaning time. Min. value: 3
sec. (relay activated or cleaning).
To deactivate the cleaning function,
T on=0 must be programmed.
Toff: Cleaning deactivation time.
Min. value: 3 s (relay deactivated ).
Example: Ton: 3 sec., Toff: 1 h; it
will clean during 3 sec. each hour.
Post-cleaning time: 0 to 99 sec.
Output current.
Select de conductivity values
corresponding to the current output
(4 mA and 20 mA).
Conductivity: between 0.0 µS
& 200 mS.
Salinity: between 0.0 mg/l &
278 g/l t
Hold. Selection of the current output
when the instrument is in Hold
state: fixed at 21 mA or at the mA
corresponding to the last measured
value.
(*) LIMIT and 4 - 20 mA OUTPUT: Allow the introduction of 5 digits. Place the decimal on the desired point.
20
si629 C_HACH LANGE_Ed 0510.indd ANG:20
si629 C_HACH LANGE_Ed 0510.indd ANG:20
MEASURING 22.3°C
545 µS/cm
@ 25°
CODE 000
LLIMIT (*)
ALARM
CLEANING
4 to 20 mA OUTPUT(*)
HOLD
Continue at
the next page.
Enter code 100
Press
until the desired option is reached.
ON - - - - - - OFF
1000 µS
800 µS
Vary the underligned digit with
and accept with
ALARM
01 min
CLEANING
Ton= 03 s Toff=01 h
4 mA
20 mA
200 µS
2.0 mS
> 21 mA
last value
Vary with
20/5/10 12:06:56
20/5/10 12:06:56