XIOS XG WiFi System and Sensors
Operating Instructions and Installation
English
XIOS XG Select
XIOS XG Supreme

Advertisement

Table of Contents
loading

Summary of Contents for Sirona XIOS XG Select

  • Page 1 11.2017 New as of: XIOS XG WiFi System and Sensors Operating Instructions and Installation English XIOS XG Select XIOS XG Supreme...
  • Page 2: Table Of Contents

    Table of contents Sirona Dental Systems GmbH Gebrauchsanweisung und Installation Table of contents General information ....................Dear Customer,.................... Contact information..................General information about this operating manual ........Other valid documents ................. Intended use ....................Indications and contraindications..............Structure of the document................
  • Page 3 Sirona Dental Systems GmbH Table of contents Gebrauchsanweisung und Installation 3.10 PC system....................3.11 RFID Reader....................3.12 Technical data....................3.13 Certification, registration and standards ............3.14 Symbols ....................... 3.15 Position of the Labels................... Installation....................... Position the PC .................... Install the PC software .................
  • Page 4 Table of contents Sirona Dental Systems GmbH Gebrauchsanweisung und Installation Enabling exposure readiness............... Sliding the hygienic protective sleeve over the sensor ........ Position the sensor ..................5.6.1 Position sensor with single-use sensor holder system ....5.6.1.1 Notes on the sensor holder tabs........
  • Page 5 Sirona Dental Systems GmbH Table of contents Gebrauchsanweisung und Installation Electromagnetic compatibility ................. 113 Accessories....................113 Electromagnetic emission ................114 Interference immunity .................. 115 Working clearances ..................116 Disposal ........................118 64 08 137 D3610 D3610.201.02.09.02    11.2017...
  • Page 6: General Information

    X-ray image in order to amplify the relevant structures. XIOS XG Supreme sensors offer higher resolution compared with XIOS XG Select sensors. In addition to SIDEXIS XG / SIDEXIS 4 (SIDEXIS XG version 2.5.6 and above) the SIDEXIS plug-in must be installed for XIOS XG.
  • Page 7: Contact Information

    Customer service center In the event of technical queries, please use our online contact form at www.sirona.com. In the navigation bar, go to the menu commands "CONTACT" / "Customer Service Center" and then click the "CONTACT FORM FOR TECHNICAL QUESTIONS" button.
  • Page 8: General Information About This Operating Manual

    1 General information Sirona Dental Systems GmbH Gebrauchsanweisung und Installation General information about this operating manual Observe the Operating Instructions Please familiarize yourself with the unit by reading through these Operating Instructions before putting it into operation. It is essential that you comply with the specified warning and safety information.
  • Page 9: Intended Use

    Sirona Dental Systems GmbH 1 General information Gebrauchsanweisung und Installation Intended use The XIOS XG X-ray system is designed for digital acquisition of intraoral X-ray images for diagnostic purposes. The system is used on patients by dental professionals. The digital images produced are transferred to a PC and displayed on a monitor.
  • Page 10: Structure Of The Document

    1 General information Sirona Dental Systems GmbH Gebrauchsanweisung und Installation Structure of the document 1.7.1 Identification of the danger levels To prevent personal injury and material damage, please observe the warning and safety information provided in these operating instructions. Such information is highlighted as follows: DANGER An imminent danger that could result in serious bodily injury or death.
  • Page 11: Safety Instructions

    Sirona Dental Systems GmbH 2 Safety Instructions Gebrauchsanweisung und Installation Safety Instructions Qualifications of operating personnel The system may only be operated by skilled or properly trained personnel. Personnel undergoing education or training, or who are using the unit as part of general training may only operate the unit under the constant supervision of an experienced person.
  • Page 12: Allocation Of Acquisition System To Patient

    2 Safety Instructions Sirona Dental Systems GmbH Gebrauchsanweisung und Installation Allocation of acquisition system to patient During routine practice, the clear allocation of the acquisition system to the patient to be examined is ensured. This also applies to the allocation of the X-rays to the patient data stored by SIDEXIS XG / SIDEXIS 4.
  • Page 13: Trouble-Free Operation

    The owner is responsible for making sure that all inspections are performed. In the event that the system owner fails to fulfill the obligation to perform scheduled inspection activities or ignores error messages, Sirona Dental Systems GmbH and its authorized dealers cannot assume any liability for damages.
  • Page 14: Changes And Extensions To The Device

    Sirona. The user is responsible for any damage resulting from the use of non-approved accessories. If devices that are not approved by Sirona are connected, they must comply with the applicable standards, e.g.: ● IEC 60950-1 or IEC 62368-1 for IT equipment (e.g. PC) and ●...
  • Page 15: Electrostatic Discharge

    Sirona Dental Systems GmbH 2 Safety Instructions Gebrauchsanweisung und Installation Electrostatic discharge 2.13 Protective measures Electrostatic discharge (abbreviated: ESD – ElectroStatic Discharge) Electrostatic discharge from people can damage electronic components when the components are touched. Damaged components usually have to be replaced. Repairs must be performed by qualified personnel.
  • Page 16 2 Safety Instructions Sirona Dental Systems GmbH Gebrauchsanweisung und Installation Integrated circuits (logical circuits and microprocessors) are used in order to implement a wide variety of functions in a device. The circuits must be miniaturized to a very high degree in order to include as many functions as possible on these chips.
  • Page 17: System Description

    Sirona Dental Systems GmbH 3 System description Gebrauchsanweisung und Installation System description System structure WiFi interface [→ 19] with RFID transponder Charger [→ 19] USB cable [→ 20] Sensor (size 0, 1 or 2) [→ 21] Single-use sensor holder system with localizer rings, guide rods, sensor holder tabs, [→ 21] and hygienic protective...
  • Page 18 3 System description Sirona Dental Systems GmbH Gebrauchsanweisung und Installation Optional RFID reader with USB interface [→ 26] SIDEXIS PC [→ 25] with ● Access to the practice's wireless network ● SIDEXIS XG / SIDEXIS 4 (SIDEXIS XG version 2.56 or higher) ●...
  • Page 19: Wifi Interface

    Sirona Dental Systems GmbH 3 System description Gebrauchsanweisung und Installation WiFi Interface The WiFi interface transfers the X-ray images to a SIDEXIS PC via a wireless practice network (WLAN). The WiFi interface contains a rechargeable storage battery. Further information can be found in the "Operation"...
  • Page 20: Power Supply Unit

    USB cable A USB cable is supplied by Sirona with the delivery. It meets the USB 2.0 standard and is designed especially for use on XIOS XG.
  • Page 21: Sensors

    Sirona Dental Systems GmbH 3 System description Gebrauchsanweisung und Installation Sensors XIOS XG can be operated with two types of sensor – the XIOS XG Select and XIOS XG Supreme sensors. Both types of sensor are available in three sizes (0, 1 and 2).
  • Page 22: Aimright Reusable Sensor Holder System

    3 System description Sirona Dental Systems GmbH Gebrauchsanweisung und Installation 3.7.2 Aimright reusable sensor holder system There are different sensor holders available depending on the type of exposure. The sensor holders are color-coded. The same localizer ring is used for all sensor holders.
  • Page 23: Intraoral X-Ray Tube Assembly

    Sirona Dental Systems GmbH 3 System description Gebrauchsanweisung und Installation Intraoral X-ray tube assembly The following requirements apply to the intraoral X-ray tube assembly: Intraoral X-ray tube assembly with multipulse technology (direct current) 0.14 – 1.4 mAs at 60 – 70 kV and 8" cone This information must be modified accordingly for other tube lengths or single-pulse X-ray tube assemblies.
  • Page 24: Wireless Access Point

    3 System description Sirona Dental Systems GmbH Gebrauchsanweisung und Installation Wireless Access Point The WiFi interface transfers the X-ray images to a SIDEXIS PC via a wireless practice network (WLAN). A wireless access point (or WLAN router) is required for the wireless connection between the WiFi interface and the PC.
  • Page 25: Pc System

    Sirona Dental Systems GmbH 3 System description Gebrauchsanweisung und Installation PC system 3.10 The digital X-ray exposures are transferred via a wireless network (WLAN) to a PC. In addition to SIDEXIS XG / SIDEXIS 4 (SIDEXIS XG version 2.56 and above) the SIDEXIS plug-in for XIOS XG must be installed. Information on the PC software can be found in the "SIDEXIS Plug-in for XIOS XG...
  • Page 26: Rfid Reader

    3 System description Sirona Dental Systems GmbH Gebrauchsanweisung und Installation RFID Reader 3.11 RFID = Radio Frequency Identification RFID technology can be used to pair the WiFi interface with the PC. An RFID reader which is inserted into a USB port on the PC is required for this.
  • Page 27: Technical Data

    Sirona Dental Systems GmbH 3 System description Gebrauchsanweisung und Installation Technical data 3.12 WiFi interface Electric class: Protection class II Type of protection against electric shock: Applied part type BF Degree of protection Usual device (without protection against against ingress of water:...
  • Page 28 3 System description Sirona Dental Systems GmbH Gebrauchsanweisung und Installation XIOS XG Select Sensors Technology: CMOS-APS (Active Pixel Sensor) Physical pixel size: 15 µm, image acquisition in 30 µm Line pairs: 16.7 Lp at 30 µm Measured resolution: 16 Lp/mm Theoretical resolution: 16.7 Lp/mm...
  • Page 29: Certification, Registration And Standards

    Sirona Dental Systems GmbH 3 System description Gebrauchsanweisung und Installation Operating conditions: Temperature: 10 °C (50 °F) to 40 °C (104 °F) Relative humidity: 20 % to 85 % Air pressure: 700 to 1060 hPa Operating altitude: ≤ 2000 m 3.13...
  • Page 30: Position Of The Labels

    3 System description Sirona Dental Systems GmbH Gebrauchsanweisung und Installation Year of manufacture YYYY Denotes sterilizable accessories. 134°C Item is only approved for one-off use. This device is sensitive to electromagnetic discharges (ESD). The electronics may be destroyed by contact with plug contacts under certain conditions, e.g., low humidity.
  • Page 31 Sirona Dental Systems GmbH 3 System description Gebrauchsanweisung und Installation Charger Power supply unit Sensors Sensor cable RFID reader 64 08 137 D3610 D3610.201.02.09.02    11.2017...
  • Page 32: Installation

    PC. 1. Log onto your PC with an administrator account. 2. Insert the "Sirona XIOS XG Select/Supreme 1.2 - Installation" CD into your PC's CD/DVD drive. 3. The setup normally starts automatically. If the setup does not start, double-click on the "Autorun.exe"...
  • Page 33 Sirona Dental Systems GmbH 4 Installation Gebrauchsanweisung und Installation 4. Click on "Installing the device connection with WiFi support" in the setup window. 5. Follow the additional instructions. 6. Restart the PC if you are required to do so. Ä The SIDEXIS plug-in for the XIOS XG WiFi system is installed.
  • Page 34: Check Firewall Settings

    4 Installation Sirona Dental Systems GmbH Gebrauchsanweisung und Installation Check Firewall Settings Please check the Firewall settings. Please go to the next section if there is no active Firewall. The steps required for different operating systems are described below as examples. The same steps must be applied where separate Firewall programs are in use: 1.
  • Page 35 Sirona Dental Systems GmbH 4 Installation Gebrauchsanweisung und Installation 3. Look for the "MarconiService" entry in the list and activate the check box. 4. If the "MarconiService" entry is not available then it must be added to the list. Click on the "Program..." or "Allow another program…"...
  • Page 36: Connect The Charger

    4 Installation Sirona Dental Systems GmbH Gebrauchsanweisung und Installation Connect the charger The WiFi interface storage battery is charged using the charger. The charger can also be used to connect the WiFi interface to a PC via USB. This is only necessary on initial installation and for WiFi interface configuration and updates.
  • Page 37: Setting Channels At The Wireless Access Point

    Sirona Dental Systems GmbH 4 Installation Gebrauchsanweisung und Installation Setting channels at the Wireless Access Point The WiFi interface can be operated only in the channels 1 to 11. Limit your router / access point to the channels 1 to 11 in the case of changing channels or set it to a fixed channel.
  • Page 38 4 Installation Sirona Dental Systems GmbH Gebrauchsanweisung und Installation ➢ Plug the WiFi interface into the charger. Ä After about 15 seconds, the LED at the WiFi interface starts to flash. Ä If the WiFi interface was turned off during plugging in, it automatically turns on and the LED turns violet.
  • Page 39 Sirona Dental Systems GmbH 4 Installation Gebrauchsanweisung und Installation 4. Enter the password if you are creating a connection to a protected network. 5. Select an attribution method from the "IP Address Configuration" menu. 6. Click on "Connect" . Ä A progress indicator appears on the lower edge of the window while the connection is established.
  • Page 40 4 Installation Sirona Dental Systems GmbH Gebrauchsanweisung und Installation successful. The WiFi interface network data (name, IP address, IP assignment and MAC address) is displayed to the left of this. The status "Ready" is displayed in the status bar on the lower edge of the window.
  • Page 41 Sirona Dental Systems GmbH 4 Installation Gebrauchsanweisung und Installation Ä The "Connect network" window opens. 2. Enter the network name (the SSID), password, security standard and operating type for the hidden network. Select the appropriate attribution method. Click on "Connect" .
  • Page 42 4 Installation Sirona Dental Systems GmbH Gebrauchsanweisung und Installation Disconnect the USB connection ➢ Remove the USB cable from the charger. It is no longer required after a successful configuration. Configure WiFi interface for several networks You can also configure WiFi interface for several networks. While doing this, observe the following: ●...
  • Page 43: Find An Error In The System Configuration

    Sirona Dental Systems GmbH 4 Installation Gebrauchsanweisung und Installation Find an error in the system configuration 4.7.1 Execute a diagnostics test A diagnostic test can be started to support troubleshooting. Start diagnostics test 1. Start the "WiFi Configuration Utility" configuration program. Start the "SIDEXIS Manager"...
  • Page 44 4 Installation Sirona Dental Systems GmbH Gebrauchsanweisung und Installation Troubleshooting The error messages provided by the diagnostic test are explained below with tips provided for corrective action: Diagnostic test Error message User action "WiFi Charger Con- "Unable to communicate with Plug the WiFi interface into the charger and nection"...
  • Page 45: Diagram For Fault Rectification

    Sirona Dental Systems GmbH 4 Installation Gebrauchsanweisung und Installation 4.7.2 Diagram for fault rectification 64 08 137 D3610 D3610.201.02.09.02    11.2017...
  • Page 46 4 Installation Sirona Dental Systems GmbH Gebrauchsanweisung und Installation 64 08 137 D3610 D3610.201.02.09.02    11.2017...
  • Page 47: Enabling The Wifi Interface In Sixabcon

    Sirona Dental Systems GmbH 4 Installation Gebrauchsanweisung und Installation Enabling the WiFi interface in SiXABCon Only when using the SIDEXIS plug-in for XIOS XG without RFID technology: X-ray images of WiFi interfaces can be transmitted only to PCs that are enabled for the relevant SIDEXIS XG / SIDEXIS 4 workstation as X-ray components.
  • Page 48: Pair Wifi Interface With The Pc

    4 Installation Sirona Dental Systems GmbH Gebrauchsanweisung und Installation Ä The WiFi interfaces are enabled for the PC. 4. Enable WiFi interfaces at additional SIDEXIS XG / SIDEXIS 4 workstations where necessary. Ä The WiFi interfaces are registered in SIDEXIS XG / SIDEXIS 4.
  • Page 49: Pairing Via A Selection List Using The Serial Number

    Sirona Dental Systems GmbH 4 Installation Gebrauchsanweisung und Installation 4.9.1 Pairing via a selection list using the serial number In the standard version of pairing, each exposure readiness is queried by the manufacturers in SIDEXIS XG / SIDEXIS 4 as to which WiFi interface is used for the exposure.
  • Page 50: Pairing Using Rfid Technology

    As a result the user should be able to see the monitor during the pairing. The RFID readers for the XIOS XG WiFi system are from Sirona. ➢ Plug the RFID reader into an easily accessible USB port, e.g. on the monitor.
  • Page 51 Sirona Dental Systems GmbH 4 Installation Gebrauchsanweisung und Installation 4. Place a check mark for "WiFi device selection via RFID" in the "Exposure sequence" optional group and click "Apply" . Ä The WiFi interfaces are now paired with the PC via RFID technology.
  • Page 52: Remove Pairing

    4 Installation Sirona Dental Systems GmbH Gebrauchsanweisung und Installation 2. Hold the wireless symbol on the ON/OFF button of the WiFi interface directly at the RFID reader. Ä The text "Tag read completed" appears briefly over the info area of the Windows taskbar (Systray). The WiFi interface is paired with the PC.
  • Page 53: Connecting The Sensor

    Gebrauchsanweisung und Installation Connecting the sensor 4.10 A calibration file is saved on the XIOS XG Select and XIOS XG Supreme sensors which is transferred to the PC the first time that the sensor is used. ü The WiFi interface is configured for the practice network.
  • Page 54: Web Server And Upgrade Utility Page

    4 Installation Sirona Dental Systems GmbH Gebrauchsanweisung und Installation Web Server and Upgrade Utility page 4.12 Functional scope overview The Web Server and Upgrade Utility page is split up into multiple tabs which cover the following functions: ● "Status" Status display for the WiFi interface ●...
  • Page 55: Flash Upgrade

    Sirona Dental Systems GmbH 4 Installation Gebrauchsanweisung und Installation 4.12.1 Flash Upgrade IMPORTANT Plug the WiFi interface into the charger for the following steps. 1. Click on the "Wireless Monitor" symbol in the info area of the Windows taskbar (Systray) and select the "WiFi Interface Management"...
  • Page 56 4 Installation Sirona Dental Systems GmbH Gebrauchsanweisung und Installation 5. In the "Flash" group click on "Browse" and open the "uImage" file. It is provided on a CD or via an Internet download. 6. Click the "Submit" button (only once).
  • Page 57: Fpga Upgrade

    Sirona Dental Systems GmbH 4 Installation Gebrauchsanweisung und Installation 4.12.2 FPGA Upgrade FPGA = Field-programmable gate array IMPORTANT Plug the WiFi interface into the charger for the following steps. 1. Click on the "Wireless Monitor" symbol in the info area of the Windows taskbar (Systray) and select the "WiFi Interface...
  • Page 58 4 Installation Sirona Dental Systems GmbH Gebrauchsanweisung und Installation 5. In the "FPGA" group click on "Browse" and open the ".dat" file. It is provided on a CD or via an Internet download. 6. Click on the "Submit" button (once only).
  • Page 59: Sensor Upgrade

    Sirona Dental Systems GmbH 4 Installation Gebrauchsanweisung und Installation 4.12.3 Sensor Upgrade IMPORTANT The WiFi interface may not be in the charger tray for the following steps. 1. Switch the WiFi interface on and wait 10 seconds until the LED flashes.
  • Page 60 4 Installation Sirona Dental Systems GmbH Gebrauchsanweisung und Installation 7. Click the "Browse" button in the "Sensor" group and open the ".hex" file. It is provided on a CD or via an Internet download. 8. Click the "Submit" button (only once).
  • Page 61 Sirona Dental Systems GmbH 4 Installation Gebrauchsanweisung und Installation 14. Complete various test exposures to check the image quality. 64 08 137 D3610 D3610.201.02.09.02    11.2017...
  • Page 62: Set The Wifi Interface

    Sirona Dental Systems GmbH Gebrauchsanweisung und Installation 4.12.4 Set the WiFi Interface Changing these settings is not recommended by Sirona. IMPORTANT Ensure that the WiFi interface is marked as "Available" in the "WiFi Interface Management" window when adjusting the settings.
  • Page 63: Install Holders And A Patient Clip

    Sirona Dental Systems GmbH 4 Installation Gebrauchsanweisung und Installation If you have decided to operate a certain WiFi interface exclusively with a certain PC then this assignment cannot be overwritten by the RFID identification for as long as automatic pairing is switched on.
  • Page 64: Wall Holder For The Charger

    4 Installation Sirona Dental Systems GmbH Gebrauchsanweisung und Installation 4.13.2 Wall holder for the charger The wall holder for the charger also ensures accessibility and protection. The holder can be fastened on a wall using double-side adhesive tape or flat head screws and dowels (supplied).
  • Page 65: Operation

    Sirona Dental Systems GmbH 5 Operation Gebrauchsanweisung und Installation Operation Switch WiFi interface on/off ü The WiFi interface is set up according to the instructions in the "Installation" chapter. 1. Remove the WiFi interface from the charger. 2. Plug the sensor cable into the WiFi interface.
  • Page 66: Determine Unit Status

    5 Operation Sirona Dental Systems GmbH Gebrauchsanweisung und Installation Determine unit status 5.2.1 LED displays on the WiFi interface The WiFi interface is fitted with a multi-color LED. The charge state for the storage battery and the unit status can be determined using the light color and the flashing signals.
  • Page 67 Sirona Dental Systems GmbH 5 Operation Gebrauchsanweisung und Installation Startup sequence purple Firmware update Blue Flashing signals The connection status for the WiFi interface can be read using the flashing signals. Unit status LED display No sensor connected Sensor connected...
  • Page 68: Led Displays On The Charger

    5 Operation Sirona Dental Systems GmbH Gebrauchsanweisung und Installation 5.2.2 LED displays on the charger The charger is fitted with three LED displays. Ready LED The power supply unit is connected to the charger, see "Connect charger" [→ 36]. Connection LED The WiFi interface is plugged into the charger, see "Charge the WiFi interface storage battery"...
  • Page 69: Charge The Wifi Interface Storage Battery

    Sirona Dental Systems GmbH 5 Operation Gebrauchsanweisung und Installation Charge the WiFi interface storage battery Please remove the holder or the patient clip when the WiFi system is not being used and place the WiFi interface in the charger so that it can charge and be ready for the next examination.
  • Page 70: Enabling Exposure Readiness

    5 Operation Sirona Dental Systems GmbH Gebrauchsanweisung und Installation Enabling exposure readiness The WiFi-Interface must be paired with the PC for creating exposure readiness. For details on registration in SIDEXIS XG / SIDEXIS 4 and pairing, please refer to the sections "Register WiFi interface in SIDEXIS"...
  • Page 71 Ä The "Exposure readiness " window is displayed. "SIDEXIS XG / SIDEXIS 4" is ready for exposures when the green display is flashing here. "Exposure readiness " window when using a "XIOS XG Select" sensor 64 08 137 D3610 D3610.201.02.09.02    11.2017...
  • Page 72 5 Operation Sirona Dental Systems GmbH Gebrauchsanweisung und Installation "Exposure readiness " window when using a "XIOS XG Supreme" sensor 6. When using a XIOS XG Supreme sensor: Select the desired filter to be applied to the x-ray image produced depending on the indication.
  • Page 73: Sliding The Hygienic Protective Sleeve Over The Sensor

    Sliding the hygienic protective sleeve over the sensor Hygienic protective sleeves are available depending on the size of the sensor (0, 1 or 2). They fit both XIOS XG Select and XIOS XG Supreme sensors. To reorder the hygienic protective sleeves, see"Consumables and spare parts".
  • Page 74: Position The Sensor

    5 Operation Sirona Dental Systems GmbH Gebrauchsanweisung und Installation Position the sensor Since the positioning of the sensors in the beam path of the cone strongly influences the image quality, using the parallel technique with the XIOS XG sensor holder system or with the #AimRight# sensor holder system is recommended for optimal positioning of the sensors.
  • Page 75: Position Sensor With Single-Use Sensor Holder System

    Sirona Dental Systems GmbH 5 Operation Gebrauchsanweisung und Installation 5.6.1 Position sensor with single-use sensor holder system 5.6.1.1 Notes on the sensor holder tabs IMPORTANT The sensor holder tabs are single use devices and cannot be reused. IMPORTANT Adhesive properties may deteriorate over time based on age and storage conditions.
  • Page 76 5 Operation Sirona Dental Systems GmbH Gebrauchsanweisung und Installation 5. Position the sensor in the patient's mouth. 6. Bring the X-ray tube assembly into the correct position and take an X-ray exposure. 7. Remove the sensor from the hygienic protective sleeve. For this follow the instructions in section "Remove the hygienic protective...
  • Page 77: Posterior Tooth Exposures

    Sirona Dental Systems GmbH 5 Operation Gebrauchsanweisung und Installation 5.6.1.3 Posterior tooth exposures For posterior tooth exposures use the yellow sensor holder. 1. Place the yellow localizer ring (B) onto the double-angled guide rod (C). 2. Place the yellow sensor holder tab (A) onto the guide rod (C).
  • Page 78: Bite Wing Exposures

    5 Operation Sirona Dental Systems GmbH Gebrauchsanweisung und Installation 5.6.1.4 Bite wing exposures For bite wing exposures use the red sensor holder. 1. Place the red localizer ring (B) onto the straight guide rod (C). 2. Place the red sensor holder tab (A) onto the guide rod (C).
  • Page 79: Endodontics Exposures With The Half-Angle Technique

    Sirona Dental Systems GmbH 5 Operation Gebrauchsanweisung und Installation 5.6.1.5 Endodontics exposures with the half-angle technique For endodontics exposures with the half-angle technique use the green universal sensor holder tab. 1. Slide the sensor into the hygienic protective sleeve, see section "Slide the hygienic protective sleeve over the sensor".
  • Page 80: Measurement Exposure For Endodontics

    5 Operation Sirona Dental Systems GmbH Gebrauchsanweisung und Installation 5.6.1.6 Measurement exposure for endodontics Endodontic needles and files can remain in the root canal for the measurement exposure. For endodontics exposures use the grey sensor holder. 1. Place the grey localizer ring (B) onto the plastic guide rod (C).
  • Page 81: Position Sensor With Aimright Reusable Sensor Holder System

    Sirona Dental Systems GmbH 5 Operation Gebrauchsanweisung und Installation 5.6.2 Position sensor with Aimright reusable sensor holder system 5.6.2.1 Anterior tooth exposure Preparing the sensor holder For anterior tooth exposures use the blue sensor holder. 1. Fasten the guide rod for the sensor holder (C) in the perforation (B) of the localizer ring (A).
  • Page 82 5 Operation Sirona Dental Systems GmbH Gebrauchsanweisung und Installation Positioning the sensor for exposures of the lower jaw 1. Look through the localizer ring to check the alignment of the sensor. The sensor must be located centrally in front of the opening in the localizer ring.
  • Page 83 Sirona Dental Systems GmbH 5 Operation Gebrauchsanweisung und Installation Positioning the sensor for exposures of the upper jaw 1. Look through the localizer ring to check the alignment of the sensor. The sensor must be located centrally in front of the opening in the localizer ring.
  • Page 84 5 Operation Sirona Dental Systems GmbH Gebrauchsanweisung und Installation 5. Align the cone of the X-ray tube assembly parallel to the sensor directly on the localizer ring. 6. Release an X-ray exposure. Refer to sections Select the exposure parameters for the X-ray tube assembly and Releasing the exposure.
  • Page 85: Posterior Tooth Exposures

    Sirona Dental Systems GmbH 5 Operation Gebrauchsanweisung und Installation 5.6.2.2 Posterior tooth exposures Preparing the sensor holder For posterior tooth exposures use the yellow sensor holders. 1. For the right upper jaw and left lower jaw: Fasten the guide rod for the sensor holder (C) in the perforation (B) of the localizer ring (A).
  • Page 86 5 Operation Sirona Dental Systems GmbH Gebrauchsanweisung und Installation Position the sensor 1. Look through the localizer ring to check the alignment of the sensor. The sensor must be located centrally in front of the opening in the localizer ring.
  • Page 87 Sirona Dental Systems GmbH 5 Operation Gebrauchsanweisung und Installation Ä Example: Positioning on the upper jaw 3. Align the cone of the X-ray tube assembly parallel to the sensor directly on the localizer ring. 4. Release an X-ray exposure. Refer to sections Select the exposure parameters for the X-ray tube assembly and Releasing the exposure.
  • Page 88: Horizontal Bite Wing Exposures

    5 Operation Sirona Dental Systems GmbH Gebrauchsanweisung und Installation 5.6.2.3 Horizontal bite wing exposures Preparing the sensor holder For bite wing exposures use the red sensor holder. 1. Fasten the guide rod for the sensor (C) in the perforation (B) of the localizer ring (A).
  • Page 89 Sirona Dental Systems GmbH 5 Operation Gebrauchsanweisung und Installation 2. Position the sensor directly on the dental arch. Ä The bite block of the sensor holder must be located between the upper molars, parallel to the occlusal plane. Ä In order to avoid superimpositions, the sensor must be positioned parallel to the dental arch line.
  • Page 90 5 Operation Sirona Dental Systems GmbH Gebrauchsanweisung und Installation 64 08 137 D3610 D3610.201.02.09.02    11.2017...
  • Page 91: Select The Exposure Parameters For The X-Ray Tube Assembly

    Sirona Dental Systems GmbH 5 Operation Gebrauchsanweisung und Installation Select the exposure parameters for the X-ray tube assembly 5.7.1 X-ray doses and image quality Factors for determining the X-ray doses The dose to be set for X-ray exposure depends primarily on the following: ●...
  • Page 92: Heliodent Plus Exposure Times

    5 Operation Sirona Dental Systems GmbH Gebrauchsanweisung und Installation 5.7.3 HELIODENT Plus exposure times 5.7.3.1 Pre-programmed exposure times for XIOS XG sensors with 200 mm (8") FHA cone Upper jaw Lower jaw Upper jaw Lower jaw Exposure time in seconds...
  • Page 93: Pre-Programmed Exposure Times For Xios Xg Sensors With 300 Mm (12") Fha Cone (Round Or Square Cone)

    Sirona Dental Systems GmbH 5 Operation Gebrauchsanweisung und Installation 5.7.3.2 Pre-programmed exposure times for XIOS XG sensors with 300 mm (12") FHA cone (round or square cone) Upper jaw Lower jaw Upper jaw Lower jaw Exposure time in seconds with: 60kV 0.12...
  • Page 94: Releasing The Exposure

    5 Operation Sirona Dental Systems GmbH Gebrauchsanweisung und Installation Releasing the exposure ü The exposure parameters for the X-ray tube assembly are set, see "Select the exposure parameters for the X-ray tube assembly" [→ 91]. ü The sensor is positioned in the patient's mouth with the corresponding auxiliary exposure equipment, see "Position the...
  • Page 95: Removing The Hygienic Protective Sleeve From The Sensor

    Sirona Dental Systems GmbH 5 Operation Gebrauchsanweisung und Installation Removing the hygienic protective sleeve from the sensor 5.9.1 With single-use sensor holder system Have the sensor holder tab and the sensor holder glued onto the hygienic protective sleeve. This makes it easier to remove the hygienic protective sleeve.
  • Page 96 5 Operation Sirona Dental Systems GmbH Gebrauchsanweisung und Installation 3. Slide the sensor further out of the hygienic protective sleeve with your thumb. 4. Hold the sensor cable firmly to prevent the sensor from falling out of the hygienic protective sleeve.
  • Page 97: With Aimright Reusable Sensor Holder System

    Sirona Dental Systems GmbH 5 Operation Gebrauchsanweisung und Installation 5.9.2 With Aimright reusable sensor holder system NOTICE The sensor cable is sensitive to mechanical influences. The cable may become damaged or may wear out prematurely. ➢ Do not pull on the sensor cable when pulling the sensor out of the hygienic protective sleeve.
  • Page 98: Alignment Of The X-Ray Image

    5 Operation Sirona Dental Systems GmbH Gebrauchsanweisung und Installation Alignment of the X-ray image 5.10 The sensor alignment is shown on the X-ray image by a small rectangle of inverted pixels. Positioning the sensor Sensor alignment Inverted pixels (encircled) and im-...
  • Page 99: Resend Non-Transmitted Wifi Exposures To Sidexis Xg / Sidexis 4

    Sirona Dental Systems GmbH 5 Operation Gebrauchsanweisung und Installation Resend non-transmitted WiFi exposures to 5.11 SIDEXIS XG / SIDEXIS 4 Auto recovery The WiFi interface tries to recover the network connection over an interval of 2 minutes if the WiFi signal strength is not sufficient for the data transmission.
  • Page 100 5 Operation Sirona Dental Systems GmbH Gebrauchsanweisung und Installation Select one of the following options. ● Click on the "Yes" button to transfer the image to "SIDEXIS XG / SIDEXIS 4" . The image is transferred. Exposure readiness is established after this.
  • Page 101: Maintenance And Inspection

    Care, cleaning agents, and disinfectants NOTICE Approved care, cleaning, and disinfecting agents Use only care, cleaning and disinfecting agents approved by Sirona! For a continuously updated list of approved agents, please visit "www.sirona.com" . To access the online portal for technical documentation, follow the "SERVICE"...
  • Page 102: Sensor Holders

    6 Maintenance and inspection Sirona Dental Systems GmbH Gebrauchsanweisung und Installation 2. The sensor and the sensor cable must be wiped clean with disinfectant completely and thoroughly at least twice. After this, wipe the WiFi interface and the charger clean with disinfectant.
  • Page 103: Regular Inspections

    Sirona Dental Systems GmbH 6 Maintenance and inspection Gebrauchsanweisung und Installation Regular inspections Inspections must be performed at scheduled intervals to protect the health and safety of patients, users and other persons. Before and during operation The system owner or an appointed person must ensure that all components such as cables, sensors and housing parts are in sound condition.
  • Page 104: Replace The Sensor Cable

    6 Maintenance and inspection Sirona Dental Systems GmbH Gebrauchsanweisung und Installation Replace the sensor cable The sensor cable can be replaced with a new one in the event of damage. A "sensor cable" replacement set is required for this, see Consumables and spare parts [→ 108]"Consumables and spare parts".
  • Page 105 3. Using the screwdriver supplied with the set remove the 2 screws which fasten the cable to the sensor. Replace the Elastomer strip 1. For XIOS XG Select sensors: XIOS XG Select Remove the existing elastomer strip from the sensor with tweezers.
  • Page 106: Replace The Wifi Interface Storage Battery

    6 Maintenance and inspection Sirona Dental Systems GmbH Gebrauchsanweisung und Installation 2. Screw the sensor cable to the sensor using the screws supplied. Initially only tighten the screws until a light resistance can be felt. After this tighten both screws carefully.
  • Page 107 Sirona Dental Systems GmbH 6 Maintenance and inspection Gebrauchsanweisung und Installation 1. There is a screw that secures the cover of the storage battery compartment at the opposite side of the LED. Unscrew this using the supplied screwdriver. 2. Remove the battery compartment cover and retain it.
  • Page 108: Consumables And Spare Parts

    7 Consumables and spare parts Sirona Dental Systems GmbH Gebrauchsanweisung und Installation Consumables and spare parts Hygienic protective sleeves for sensors Hygienic protective sleeves for sensor size 0 and 1, quantity 300 REF 64 09 960 Hygienic protective sleeve for sensor size 2, quantity 300...
  • Page 109 Sirona Dental Systems GmbH 7 Consumables and spare parts Gebrauchsanweisung und Installation Aimright reusable sensor holder system Localizer ring (red), quantity 1 REF 65 45 599 Sensor holder for bite wing exposures (red), for sensor size 1, quantity REF 65 45 557...
  • Page 110 Test phantom for sensor size 0, 1 and 2 for consistency test REF 64 00 449 Sensors with cable for WiFi interface XIOS XG Select WiFi size 0 - 45 cm REF 64 06 271 XIOS XG Select WiFi size 1 - 45 cm...
  • Page 111 Sirona Dental Systems GmbH 7 Consumables and spare parts Gebrauchsanweisung und Installation XIOS XG Supreme WiFi size 1 - 45 cm REF 64 06 297 XIOS XG Supreme WiFi size 2 - 45 cm REF 64 06 289 Sensor cable for WiFi interface...
  • Page 112 7 Consumables and spare parts Sirona Dental Systems GmbH Gebrauchsanweisung und Installation XIOS XG WiFi RFID reader, quantity 3 REF 64 04 334 XIOS XG WiFi RFID reader extension cable, 2m REF 64 04 581 64 08 137 D3610 D3610.201.02.09.02    11.2017...
  • Page 113: Electromagnetic Compatibility

    (18") The UNIT may only be operated with accessories and spare parts approved by Sirona. Unapproved accessories and spare parts may lead to an increased emission or to a reduced immunity to interference. The UNIT should not be operated in the immediate vicinity of other devices.
  • Page 114: Electromagnetic Emission

    8 Electromagnetic compatibility Sirona Dental Systems GmbH Gebrauchsanweisung und Installation Electromagnetic emission The UNIT is intended for operation in the electromagnetic environment specified below. The customer or user of the UNIT should make sure that it is used in such an environment.
  • Page 115: Interference Immunity

    Sirona Dental Systems GmbH 8 Electromagnetic compatibility Gebrauchsanweisung und Installation Interference immunity The UNIT is intended for operation in the electromagnetic environment specified below. The customer or user of the UNIT should make sure that it is used in such an environment.
  • Page 116: Working Clearances

    8 Electromagnetic compatibility Sirona Dental Systems GmbH Gebrauchsanweisung und Installation Interference immunity IEC 60601-1-2 test Compliance level Electromagnetic environment - tests level guidelines Conducted RF interfer- d= [1.2] √P ence 150 kHz to 80 MHz IEC 61000-4-6 Radiated RF interfer- 3 V/m d= [1.2] √P...
  • Page 117 Sirona Dental Systems GmbH 8 Electromagnetic compatibility Gebrauchsanweisung und Installation Power rating of the transmitter Working clearance according to transmission frequency [m] 150 kHz - 80 MHz 80 MHz - 800 MHz 800 MHz - 2.5 GHz d= [1.2] √P d= [1.2] √P...
  • Page 118: Disposal

    9 Disposal Sirona Dental Systems GmbH Gebrauchsanweisung und Installation Disposal In accordance with Directive 2012/19/EU and national disposal regulations regarding old electrical and electronic devices, please be advised that such items must be disposed of in a special way within the European Union (EU).
  • Page 120 We reserve the right to make any alterations which may be required due to technical improvements. © Sirona Dental Systems GmbH Sprache: englisch Printed in Germany D3610.201.02.09.02      11.2017 Ä.-Nr.: 124 617 Sirona Dental Systems GmbH 64 08 137 D3610 Order No    Fabrikstr. 31...

This manual is also suitable for: