Output Rating: 24 V DC, 150 W Intended Use The MicroAire REF 7500-700 Battery Charger is intended to charge all MicroAire 14 . 4 volt NiMH battery packs . Introduction This manual was written to help describe the procedures required to keep the MicroAire Battery Charger system operating properly .
Use of the REF 7500-700 Battery Charger adjacent to or stacked with other equipment should be avoided because it could result in improper operation . If such use is necessary, the REF 7500-700 Battery Charger and the adjacent equipment should be observed to verify that they are operating normally .
. The user might be required to take mitigation measures, such as relocating or re-orienting the equipment . NOTE: The Battery Charger is designed for use only with MicroAire Batteries REF 6640-710, REF 7505-710 and REF 7500-620 Aseptic Battery using a REF 7500-625 charger adapter . NOTE: All personnel should become familiar with the power equipment before it is set-up for use in any procedure .
Page 5
. Manufacturer 3082 ISO 15223-1:2012 • When using MicroAire as the manufacturer, use the MicroAire LLC symbol . Indicates the European Conformity Mark with Notified Body Council Directive CE Mark with NB Number .
OPERATION 1 . Verify that the “Standby” LED is on . 2 . Insert a MicroAire REF 6640-710 or REF 7505-710 Battery into any slot . The battery can only be inserted into the charger in one position . 3 . If charging the REF 7500-620 Aseptic Battery, first place the REF 7500-625 Charger Adapter into the charger . Then place the REF 7500-620 into the REF 7500-625 .
Humidity Temperature SHIPPING CONDITIONS The REF 7500-700 may be shipped by any standard commercial method without special handling conditions . TROUBLESHOOTING If the unit fails to operate properly, check the following: 1 . The unit is connected to a functional power outlet .
. Extended Warranty An extended warranty is available on all MicroAire instruments . If the instrument is out of warranty, it must first be restored, if necessary, to full serviceable condition before being eligible for the extended warranty .
0 .15 8 .2 ELECTROMAGNETIC COMPATIBILITY WARNING: The REF 7500-700 Battery Charger is suitable for use in hospitals and surgery centers . The Battery Charger should not be used near High Frequency Surgical Equipment or Magnetic Resonance Imaging equipment . NOTE: The emissions characteristics of this product make it suitable for use in industrial areas and hospitals (CISPR 11 Class A) .
Page 10
IEC 61000-4-11 Cycles at 60 Hz) *NOTE: The 7500-700 Battery Charger has no Essential Performance as defined by ANSI/AAMI/ES60601-1 . Compliance with Immunity Testing is defined as: at least one charger output will remain functional after the test . IM-7500-700 Rev K 2021-05...
Need help?
Do you have a question about the 7500-700 and is the answer not in the manual?
Questions and answers