FCS2 (Focht Live-Cell Chamber System) Instructions
A Message from the Developer:
The FCS2 utilizes the most advanced technology for optimum functionality and versatility for the observation of living cells on a
light microscope. As part of our dedication to respond to our customers needs we would be very appreciative of any comments or
suggestions you may have relative to our products continued improvement.
Before You Begin:
Check to make sure there is no visible damage in shipment and save all packing materials.
The Controller box should contain; controller, power cord, 2.5mm screwdriver, 1/8" jack, and 3 pin mini DIN
connector.
Check the packing list and call your supplier if there are any discrepancies.
Numeric Display should be illuminated when the controller is turned on. If it does not light, check the main power source then the
fuse block located in the power input module where the AC power cord plugs in.
Display Selector is the black button adjacent to the display. It selects the source of the information displayed. When the controller
is turned on, the default condition is to display the setpoint value indicated by the red light adjacent to the word Setpoint. The button
selects one of a sequence of display options. This button can be pressed at any time and has no effect on the temperature regulation
functions of the controller.
Display Lights indicate the source of the value displayed on the numeric display.
Setpoint is adjusted by a potentiometer accessible with the 2.5mm screwdriver provided. To adjust the setpoint set the numeric
display to Setpoint, then adjust the potentiometer to the desired value. If you need to change the setpoint value during your
experiment on a routine basis there is a remote setpoint jack on the back of the controller. See Remote Setpoint.
Slide is the location where the calibration of the circuit reading the Microaqueduct Slide can be adjusted.
Chamber is the location where the calibration of the circuit reading the Chamber can be adjusted.
Reset (Red Button) resets the controller if the alarm sounds. Note: If the alarm went off there must have been a reason. Check the
chamber before resetting the controller.
Calibrate substitutes precision resistors into the circuit that reads the thermistors so that calibration adjustments can be made to a
known value.
Chamber is a six pin mini DIN connector where the FCS2 or FCS3 is plugged in.
3560 Beck Road • Butler, PA 16002 • Voice: 724-282-7145 • Fax: 724-282-0745 • E-mail: info@bioptechs.com • Web: www.bioptechs.com
Front Panel of FCS2 & FCS3 Controller
Figure 1

Control Interface Descriptions:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the FCS2 and is the answer not in the manual?

Questions and answers

Summary of Contents for Biotech FCS2

  • Page 1 A Message from the Developer: The FCS2 utilizes the most advanced technology for optimum functionality and versatility for the observation of living cells on a light microscope. As part of our dedication to respond to our customers needs we would be very appreciative of any comments or suggestions you may have relative to our products continued improvement.
  • Page 2 Back Panel of the FCS2 & FCS3 Controller Figure 2 AC Power line input, power switch, voltage selector, and fuses are all incorporated in the AC Power Input Module. The controller can operate between 90 - 260V 50-60Hz, or a 24V battery can be used as an alternate power supply. The fuse carrier contains two 0.5A slow blow fuses.
  • Page 3 • the feedback loop that controls the microaqueduct temperature is reliant on this thermal contact point. Plug the FCS2 Chamber into the controller (6 Pin mini DIN). • The display can be switched between Set, Slide and Chamber at any time to monitor temperature without interfering •...
  • Page 4 2. Coverslips and gaskets should be hand cleaned and short cycle autoclaved before use. Do not use alcohol on gaskets. 3. The white plastic (Delrin) portion of the FCS2 will be in fluid communication with the specimen and can be autoclaved at 120°C for 15 minutes.

This manual is also suitable for:

Fcs3

Table of Contents