Hardware Manual
ORIGINAL INSTRUCTIONS 1.00 – MAY 2021

Advertisement

Table of Contents
loading

Summary of Contents for cetoni Nemesys S

  • Page 1 Nemesys S Hardware Manual ORIGINAL INSTRUCTIONS 1.00 – MAY 2021...
  • Page 2 CETONI GmbH Wiesenring 6 07554 Korbussen Germany +49 (0) 36602 338-0 +49 (0) 36602 338-11 info@cetoni.de www.cetoni.de CETONI Nemesys S Hardware Manual...
  • Page 3 If you control the products with a software from CETONI GmbH, you agree to the applicable license agreement, which can be read in the corresponding software manual. This and all other current product manuals can be found at https://www.cetoni.de/en/downloads/manuals.
  • Page 4 CETONI Nemesys S Hardware Manual...
  • Page 5: Table Of Contents

    Safety measures 2.4.5 Measures for Safe Operation Warranty and Liability Scope of Delivery Technical Data Environment Mechanical Data Electrical Data Interfaces Dimensional Drawing Dosing Performance Materials in Contact with Media 4.7.1 Syringes 4.7.2 Contiflow Valve CETONI Nemesys S Hardware Manual...
  • Page 6 OEM RS232 Cable Set Overload shut-off 5.10 Changing the syringe holder 5.11 Changing the Piston Holder 5.12 Universal Syringe Holder Nemesys S 5.13 Glass Syringes - Instructions for Use 5.13.1 Returning glass syringes 5.13.2 Warranty glass syringes Transport and Storage Maintenance and Care...
  • Page 7: Revision History

    Please let us know your device type and email address and we will send you the appropriate manual as a pdf file. The type of your product can be found on the label behind "Type:", according to the marked number in the following example: CETONI Nemesys S Hardware Manual...
  • Page 8: Introduction

    Introduction Preface Thank you for purchasing a product from CETONI. With this user manual we would like to support you as well as possible when handling the device. If you have any questions or suggestions, please do not hesitate to contact us.
  • Page 9: Norms And Directives

    General Description of the Device The CETONI Nemesys S is a syringe pump. It allows emptying and filling syringes by the relative linear movement of a syringe- and a piston holder. The use of two CETONI Nemesys S equipped with Contiflow Valves enables uninterrupted fluid dispensing.
  • Page 10: Safety Measures

    2.4.4 Safety measures The safety of the user and a failure-free operation of the CETONI Nemesys S are assured only if original parts are used. Only original accessories may be used. Warranty claims will not be accepted for damage due to the use of alien accessories or expendables.
  • Page 11: Measures For Safe Operation

    2.4.5 Measures for Safe Operation 2.4.5.1 ELECTROMAGNETIC EMISSIONS The CETONI Nemesys S is intended for use in any type of facility, connected directly to the public power supply network that supplies buildings used for domestic purposes. 2.4.5.2 ELECTROSTATIC DISCHARGE Floors should be made of wood, concrete, or ceramic tiles. If the flooring is made of a synthetic material;...
  • Page 12: Warranty And Liability

    CETONI GmbH considers itself responsible for the devices with regard to safety, reliability and function only if assembly, new settings, changes, extensions and repairs are carried out by CETONI GmbH or an authorised centre, and if the devices have been used in accordance with the instruction manual.
  • Page 13: Scope Of Delivery

    The following items should be included: CETONI NEMESYS S • With Contiflow Valve in the ordered version SCREW-IN SYRINGE HOLDER Other accessories such as syringes, tubing material, pressure sensors etc. have to be purchased separately. CETONI Nemesys S Hardware Manual...
  • Page 14: Technical Data

    DIMENSIONS (L x W x H) 310 x 56 x 108 mm WEIGHT 2,25 kg Electrical Data SUPPLY VOLTAGE (POWER SUPPLY UNIT) 24 V DC POWER CONSUMPTION 24 W Interfaces 1 Mbit/s RS-232 Section 5.8 ACCESSORY PORT Section 5.7 CETONI Nemesys S Hardware Manual...
  • Page 15: Dimensional Drawing

    Dimensional Drawing CETONI Nemesys S Hardware Manual...
  • Page 16: Dosing Performance

    Dosing Performance Only special SETonic glass syringes (see section 5.12) may be used with the CETONI Nemesys S. Only with the Universal Syringe Holder Nemesys S (see section 5.12) the use of other syringes is permitted, whereby the reliable and safe function in this case must be ensured by the operator and cannot be guaranteed by CETONI GmbH.
  • Page 17: Materials In Contact With Media

    The material (PPS or PEEK) is also engraved on the base body. PEEK (Polyetheretherketone), PPS (Polyphenylene sulphide), FFKM (Perfluoroelastomer), FKM (Fluoroelastomer), Al2O3 (Aluminum oxide ceramics) Al2O3 (Aluminum oxide ceramics) CETONI Nemesys S Hardware Manual...
  • Page 18: Operating The Hardware

    System Manual. However, there are the following deviations when configuring the CETONI Nemesys S and when separating: The CETONI Nemesys S is equipped with a rotary switch. This allows the module to be switched off so that it can remain connected to the system while other modules are being configured. The module...
  • Page 19: Configuration Contiflow Valve

    Configuration Contiflow Valve The CETONI Nemesys S is supplied with a Contiflow Valve. The installation on the device is described in section 5.10, the electrical connection in section 5.7. If you have created a configuration with your CETONI Nemesys S in QmixElements, you must assign the Contiflow Valve you have in hand to it.
  • Page 20 Nevertheless, be careful and take appropriate protective measures if pressure peaks are expected from your application. 10 bar 5 bar ATTENTION. Do not exceed the permissible pressure range to avoid damaging the valves. Therefore, use the pressure monitoring function of QmixElements. CETONI Nemesys S Hardware Manual...
  • Page 21: Dead Volume Free Syringe Holder

    Dead volume free syringe holder The dead volume free syringe holder is included in the scope of delivery of the CETONI Nemesys S and can be mounted instead of the Contiflow Valve (see 5.10). It is particularly suitable for use with small syringes when a continuous fluid flow is not required.
  • Page 22 Screw the syringe into the syringe holder. If there is no tight connection when screwing it in directly, insert a PTFE sealing washer (AD:5,2 mm x ID:1,9 mm x T:0,9 mm) into the bore of the syringe holder. You can obtain this from CETONI. PTFE Sealing Washer (optional)
  • Page 23 If the piston holder is set for a particular syringe, you can remove this syringe and, for example, insert it again when it is filled. IMPORTANT. Syringes, and particularly the seals, are wear parts. Check them on a regular basis and replace them if necessary. CETONI Nemesys S Hardware Manual...
  • Page 24: Fluidics

    Tube to the reservoir Tube to the application LEDs For information on how to configure the Continuous Flow in CETONI Elements, please refer to the software manual. ATTENTION. Only use fittings and hoses that are approved for the expected pressure and check the tightness of all fluidic connections after connection and at regular intervals.
  • Page 25: Upright Operation

    For easier bleeding of the syringes, it is often helpful to operate the syringe pump upright. To increase the standing safety in this case, the CETONI Nemesys S is equipped with a stand foot. This is magnetically located in a recess on the side. Take it out, turn the module upright and put the stand foot back in place rotated by 90°.
  • Page 26: Accessory Port

    The accessory port allows the integration of accessories such as valves and pressure sensors. To do this, plug the cable connector of the accessory component into the socket of the CETONI Nemesys S until it engages. Make sure that the plug can only be inserted when the coding nose is pointing upwards.
  • Page 27: Rs-232 Connection

    Now, you can reactivate your system and send or receive data through RS232. Since every module contains a gateway from RS232 to the system’s internal CAN bus, you can now address each module of your system with only one RS232 cable. CETONI Nemesys S Hardware Manual...
  • Page 28 Nemesys interface and the 9-pin D-Sub: Nemesys Interface (socket) 9-pin D-Sub socket Cable RS232 RX Orange TXD Transmit Data RS232 TX Brown RXD Receive Data Signal GND Black GND Signal GND D-Sub housing = shielding Shielding CETONI Nemesys S Hardware Manual...
  • Page 29: Overload Shut-Off

    Overload shut-off The CETONI Nemesys S has a nominal thrust force of 400 N, which enables pressure to be built up in the syringe as well as the associated application. To protect the device and the application from damage due to overload, it is equipped with a force measuring device.
  • Page 30: Changing The Piston Holder

    After you have also unscrewed the lock nut and unscrewed the rest of the adjusting screw, the two screws with which the piston mount is fastened are accessible. You can unscrew these with a 3 mm and a 4 mm Allen key and then change the piston holder. CETONI Nemesys S Hardware Manual...
  • Page 31 Handle of Adjusting Screw Piston Holder Lock Nut Adjusting Screw Grub Screws Screws Schrau CETONI Nemesys S Hardware Manual...
  • Page 32: Universal Syringe Holder Nemesys S

    5.12 Universal Syringe Holder Nemesys S Before mounting a syringe on the CETONI Nemesys S, it must be configured and selected in the software. The procedure is described in the software manual. For this you need the Volume (Scale Volume), the nominal stroke (Scale Length) and the maximum stroke (Piston Stroke).
  • Page 33 In the event of an unexpected voltage drop (e.g. power failure), take appropriate measures to prevent push-back (e.g. check valve). IMPORTANT. CETONI does not give any guarantee that any syringe will fit into the Nemesys S universal syringe holder. The accuracy of the dispensing depends on the correct configuration of the syringe.
  • Page 34: Glass Syringes - Instructions For Use

    Only special SETonic glass syringes may be used with the CETONI Nemesys S. These syringes are matched to the CETONI Nemesys S with regard to their dimensions. The use of other syringes can lead to collisions and thus to damage to the syringe or the device.
  • Page 35 At the end of the cleaning process, it is best to use a solvent with a low boiling point (e.g., alcohol). o Dry the cylinder and the piston in an air stream. ATTENTION. Do not immerse the complete glass body into solvents! The glued parts may detach. CETONI Nemesys S Hardware Manual...
  • Page 36: Returning Glass Syringes

    Therefore, SETonic assumes no liability for damage and consequential damage caused by the use of the products. All products provided from SETonic are wear and tear parts. SETonic does not give a warranty about the lifetime. CETONI Nemesys S Hardware Manual...
  • Page 37: Transport And Storage

    Please do not lift and transport the devices in the plugged together state, unless you use the original packaging. The CETONI Nemesys S contains a lithium metal battery. A consignment may consist of a maximum of two packages, each containing a maximum of four devices. No further markings are required for shipping.
  • Page 38: Maintenance And Care

    CETONI GmbH to coordinate further actions. The device may only be opened by CETONI GmbH or thereby authorized service staff. Otherwise, the warranty claims are void.
  • Page 39: Disposal

    Disposal Please send your old devices back to CETONI GmbH. We will take care of proper disposal. The device contains a battery and must not be disposed of with domestic waste. Please follow the instructions in chapter 6 for the return shipment.