  • Page 1 Manufacturer of Quality Medical Products Image Diagnostics, Inc. IDI 1000F Series Monitor Suspension System Model IDI 1000F-1 thru 1000F-6 With Adapter to CAS 8000 Rails Installation and Maintenance Guide D000-479 REV J MAY-2018 Image Diagnostics, Inc. Page...
  • Page 2: Table Of Contents

    Installation and Maintenance Guide for IDI 1000F Series Suspension System for use with CAS 8000 Rails The text of this manual was originally written, approved and published by the manufacturer in English. The information in this manual is subject to change without notice.
  • Page 3 Installation and Maintenance Guide for IDI 1000F Series Suspension System for use with CAS 8000 Rails A. INTRODUCTION The IDI 1000F Flat Panel Suspension System transports monitors used in medical imaging procedures. The system features a counterbalanced arm with vertical travel and internal cable routing. A dual-axis flat panel monitor mount permits independent panel mounting and positioning.
  • Page 4: Symbols

    Installation and Maintenance Guide for IDI 1000F Series Suspension System for use with CAS 8000 Rails B. SYMBOLS Attention, consult accompanying documents. Failure to follow these instructions can cause accidents resulting in serious injury to patient, user, and damage to equipment.
  • Page 5: General Safety

    Installation and Maintenance Guide for IDI 1000F Series Suspension System for use with CAS 8000 Rails C. GENERAL SAFETY Attention, consult accompanying documents. Failure to follow these instructions can cause accidents resulting in serious injury to patient, user, and damage to equipment.
  • Page 6: Safety Hazards

    Installation and Maintenance Guide for IDI 1000F Series Suspension System for use with CAS 8000 Rails D. SAFETY HAZARDS Installers and Operators using this equipment should understand the safety issues and operating instructions provided. Comments and questions regarding safety should be addressed to: Customer Support Image Diagnostics, Inc.
  • Page 7: Installation

    Installation and Maintenance Guide for IDI 1000F Series Suspension System for use with CAS 8000 Rails INSTALLATION UNPACK No special handling is required for unpacking this equipment at the site. Conventional shipping materials are used. Recycle or disposal of shipping material per local regulations...
  • Page 8 Installation and Maintenance Guide for IDI 1000F Series Suspension System for use with CAS 8000 Rails GUIDE (left view) and STANDARD RAIL ADAPTER SUBASSEMBLIES D000-479 REV J MAY-2018 Image Diagnostics, Inc. Page...
  • Page 9 Installation and Maintenance Guide for IDI 1000F Series Suspension System for use with CAS 8000 Rails Next, mount the two guide bars (item 27) to the guide rail adapter carriage subassembly using screws (item 33) and locknuts (item 16) as shown below.
  • Page 10 Installation and Maintenance Guide for IDI 1000F Series Suspension System for use with CAS 8000 Rails MOUNT PLATES & STRUTS (WIDE CARRIAGE) MOUNT PLATES & STRUTS Check bridge assembly carefully for both parallelism between components and perpendicularity with the longitudinal rails. Verify the load bearing support wheels on both sides of the adapter run true and as close as possible to the center of the longitudinal rails.
  • Page 11 Installation and Maintenance Guide for IDI 1000F Series Suspension System for use with CAS 8000 Rails 17 1/8 inches [43.5cm] Attach clips to middle plates (item 23) installed in step 5 with hardware (item 14, 26, 15, 16). Using LOCTITE #242 and tighten securely.
  • Page 12 Installation and Maintenance Guide for IDI 1000F Series Suspension System for use with CAS 8000 Rails Install the second end plate (item 22) at the near end of the rails, making sure rubber bumper faces in towards the carriage. Use LOCTITE #242 on all threads.
  • Page 13 Installation and Maintenance Guide for IDI 1000F Series Suspension System for use with CAS 8000 Rails Apply a very little amount of LOCTITE #242 to and insert the six setscrews (item 24) into the six threaded holes on second rotation cap (item 43). Thread the setscrews in until they contact the rotation cap (item 43).
  • Page 14 Installation and Maintenance Guide for IDI 1000F Series Suspension System for use with CAS 8000 Rails SET ROTATIONAL LIMITS (Refer to Monitor Mount Assembly Drawing) If desired, the stop pin (item 4) can be relocated to another of its four possible mounting locations.
  • Page 15 Installation and Maintenance Guide for IDI 1000F Series Suspension System for use with CAS 8000 Rails MAIN GROUND WIRE First attach main ground wire FROM SERVICE under split lockwasher and nut "A". Attach all other ground wires NUT "B" between nut "A", lockwashers, and nut "B".
  • Page 16 Installation and Maintenance Guide for IDI 1000F Series Suspension System for use with CAS 8000 Rails INSTALL MOBILE CABLE HANDLING ASSEMBLY (Refer to Cable Handling Kit Assembly Drawing) Slide nut plate of two hanger standoff assemblies (items 1&2) into rail track.
  • Page 17 Installation and Maintenance Guide for IDI 1000F Series Suspension System for use with CAS 8000 Rails Position shaft/cable handling subassemblies (items 1-5) to desired location on rail. Without disrupting alignment, remove shaft and shaft supports. Torque standoffs to ~35-50 inch pounds, 4.2-6.0(Nm). Use LOCTITE #242 on hardware.
  • Page 18 Installation and Maintenance Guide for IDI 1000F Series Suspension System for use with CAS 8000 Rails 15 3/4" 48" [1220mm] [400mm] 36" [914mm] STATIONARY CABLE 47" [1200mm] HANDLING ASSEMBLY MOBILE CABLE HANDLING ASSEMBLY OPTIMUM CABLE HOSE CONFIGURATION ACTUAL ROOM LAYOUT AND EQUIPMENT APPROX.
  • Page 19 Installation and Maintenance Guide for IDI 1000F Series Suspension System for use with CAS 8000 Rails Yoke Balance Adjustment WARNING! Yoke been leveled manufacturers. Make adjustment only if absolutely necessary to compensate for an unbalanced load. NOTE: Prior to yoke adjustment - payload must be secure and arm horizontal.
  • Page 20: Cleaning

    Installation and Maintenance Guide for IDI 1000F Series Suspension System for use with CAS 8000 Rails CLEANING THE EQUIPMENT No part of this unit is designed to be sterilized in an autoclave. Do not allow water or other liquids to enter the equipment as this may cause short circuits or corrosion.
  • Page 21 Installation and Maintenance Guide for IDI 1000F Series Suspension System for use with CAS 8000 Rails • Screws or some parts are not loosened or removed. • The pad mounted to the bottom of the monitor mount assembly is not damaged or missing.
  • Page 22 Installation and Maintenance Guide for IDI 1000F Series Suspension System for use with CAS 8000 Rails 3. Operational check Move all the movable sections and confirm the following:  Operation does not lack smoothness compared to when the monitor suspension was installed.
  • Page 23 Installation and Maintenance Guide for IDI 1000F Series Suspension System for use with CAS 8000 Rails Check that the hardware securing the end plates are tight. Perform a complete functional inspection: -Rotate monitor mount assembly through complete range. -Raise and lower monitor mount assembly.
  • Page 24 Installation and Maintenance Guide for IDI 1000F Series Suspension System for use with CAS 8000 Rails Check the pad attached to the underside of the monitor mount. If there is any damage to the pad or if the pad is not properly attached to the unit, it must be replaced.
  • Page 25: Customer Support

    Installation and Maintenance Guide for IDI 1000F Series Suspension System for use with CAS 8000 Rails H. CUSTOMER SUPPORT For technical assistance, be sure you have the complete model and serial number before contacting the local or national service office.
  • Page 26 Installation and Maintenance Guide for IDI 1000F Series Suspension System for use with CAS 8000 Rails August 2008 ECO 294 April 2012 ECO 324 May 2018 ECO E571 Added warnings for pad. D000-479 REV J MAY-2018 Image Diagnostics, Inc. Page...

This manual is also suitable for:

1000f-11000f-21000f-31000f-41000f-51000f-6

Table of Contents