Download Print this page

Haier DW-86L579BP Operation Manual page 16

Ultra low temperature freezer

Advertisement

Control panel
•DW-86L579BP/729BP/829BP/959BP
•DW-86L579BPT/729BPT/829BPT/959BPT/959W/828W
14
• A fl ashing alarm cannot be cancelled unless the malfunction is eliminated. The buzzing
Caution
alarm can be temporarily silenced for 30 minutes by pressing the"Silence" key. However,
If the problem is not fi xed. the buzzer alarm will resume after 30 minutes.
• When using the freezer, the battery switch must be turned on to charge the battery.
• When there is a power outage, the battery sustains the temperature display. If the
battery voltage is insuffi cient. the temperature display will turn off .
• While the battery is still capable of providing power to the display, the temperature
display can be turned off by unplugging the power cord and turning off the battery
control switch.
• The freezer is also designed to auto-adjust the inner temperature set point in high
ambient temperatures When the ambient temperature is warmer than 35℃ and the
set point temperature is set to be lower than -82℃ , the set point temperature will
automatically default to -82℃ . If the ambient temperature is equal to or cooler than
30℃ ,the set point will resume at the intended inner set point. This feature extends the
life expectancy of the freezer.
DW-86L579BPT/729BPT/829BPT/959BPT/959W/828W
Alarm Type
Tricolor lamp panel
High-low temperature alarm
Power-off alarm
Door opening alarm
Main temperature sensor fault alarm
Heat exchanger fault alarm
Condenser dirty alarm
High ambient temperature alarm
Battery fault alarm
Backup system fault alarm
Automatic recovery of alarm
• Under alarm, the beeper alarm could be ceased by pressing any key, and the fl ashing alarm and
remote alarm will not be ceased(DW-86L579BPT/729BPT/829BPT/959BPT/959W/828W:Click "×"
on the upper right-hand corner in the prompt box to stop the buzzer alarm. If there is no prompt box,
click the alarm on the bottom of the homepage to view the alarm, and then stop the buzzer alarm.).
• If the alarm condition is still available, the beeper alarm will be recovered automatically 30 minutes
after ceasing.
Buzzer
A prompt box pops up.
Red
Y
Y
Red
Y
Y
Red
Y
Y
Yellow
Y
N
Yellow
Y
N
Yellow
Y
N
Yellow
Y
N
Yellow
Y
N
Yellow
Y
N
35