2.
When stable is displayed touch the CAL display key, "C 0 GM" will be
displayed, (scale units automatically defaults to grams mode during
calibration as calibration weights are measured in grams). Once "STABLE"
is displayed, touch the CAL display key a second time to confirm the 0 gram
load (Photo 8).
3. When "C 50 GM" and stable is displayed, place one of the 50 gram weights onto
the platen. When "STABLE" is displayed touch the CAL display key to confirm the
50 gram load (Photo 9).
4. When "C 100 GM" and stable is displayed, place the second 50 gram weight onto
the platen. When "STABLE" is displayed touch the CAL display key to confirm the
100 gram load (Photo 10).
RC444_98945_RangeMaster2000_UserManual_rev102318.indd 8
Photo 8: Calibration Step #2
Photo 9: Calibration Step #3
8
10/24/18 3:04 PM