Thank you for purchasing the RCBS Digital Pocket Scale.
Please read all operating instructions carefully prior to
use. This electronic scale is a precision instrument, and will
provide years of reliable service. Never load the scale with
more than the maximum load capacity of 100 grams or 1500
grains. Overloading will permanently damage the scale.
Avoid any exposure to extreme heat or cold, your scale works
better when operated at normal room temperature. Always
allow the scale to acclimate to normal room temperature and
allow for sufficient warm up time prior to calibration. This is
done to ensure the internal components have stabilized and
reached thermal equilibrium. 30 minutes will be a sufficient
warm up period.
Keep your scale in a clean environment, free of dust, dirt and
moisture. Be aware of vibration and air currents as this can
also affect the scales performance. Maintain an adequate
distance from other electronic devices as some electronics
can cause an electrical disturbance which can also affect
the scales accuracy. Handle with care and apply all items to
be weighed directly onto the platform or in the provided
scale pan. This scale is built for its durability; however, try to
avoid unnecessary rough handling as this may permanently
damage the load cell.
Avoid shaking, dropping or otherwise shocking the scale. This
is a precision instrument and must be handled with extreme
care.
ONLY OPERATE THE SCALE ON A STABLE, VIBRATION FREE
SURFACE.
Prior to use, turn on scale by pressing the "ON/OFF" key.
Place the calibration weight onto the platform and allow
unit to warm up for 30 minutes prior to use.
CALIBRATION
1. Place scale on a flat surface.
2. Press "ON/OFF" key to turn the scale on.
3. When "0.0" or "0.000" is displayed, press and hold the
"MODE" key until "CAL" is displayed
4. Press the "MODE" key again, "50.00g" will flash.
5. Place the 50 gram Calibration weight onto the center of
the platform.
6. Leave the calibration weight on the platform until "PASS"
is displayed.
7. Remove calibration weight.
8. Scale will return to zero.
OPERATION
1. Use scale only after successful completion of calibration.
2. Use the "MODE" key to select the desired units of
measure; Grams (g) or Grains (gn).
3. Place the scale pan onto the platform and press the
"TARE" key. This will tare out the scale pan, so objects
can be placed inside of the scale pan for weighing.
4. If the scale pan is not being used, press the "TARE" with
nothing on the platform, this will allow for placing
objects directly onto the platen for weighing.
It is good practice to periodically check the scale by placing
the calibration weight onto the platen. If the displayed weight
is off by more than +/- .01 g or .1 gn. ; re-calibrate the scale.

Advertisement

loading

Summary of Contents for RCBS 1500 GN

  • Page 1 30 minutes prior to use. USER MANUAL CALIBRATION Thank you for purchasing the RCBS Digital Pocket Scale. 1. Place scale on a flat surface. Please read all operating instructions carefully prior to 2. Press “ON/OFF” key to turn the scale on.
  • Page 2 IMPORTANT: PLEASE READ ALL INFORMATION FIRST ammunition reloaded with this product. Like any scale, you control the accuracy of your RCBS® Digital Mini-Scale. Before using the scale, read this instruction manual carefully to fully learn how to safely operate POWDER the scale.
  • Page 3 Send a check in the amount of ten dollars ($10) U.S. payable to RCBS Operation for return shipping and handling, Re-check frequently during use with the calibration weight provided. If accuracy is found to be out of tolerance, the scale must be recalibrated.
  • Page 4 For you, this is free of charge. If the equipment contains exchangeable (rechargeable) batteries, these too RCBS • 605 Oro Dam Blvd. East • Oroville, CA 95965 must be removed before and, if necessary, in turn be disposed of according to the relevant regulations (see also the respective comments in this unit’s...

This manual is also suitable for:

98914