Table 6: Test points and trimpots allocation
Reference Points
Signal
Test Points
WARNING
ALARM
Ground
WARNING
Trimpots
ALARM
Span
FIGURE 8: Gas Monitor 1ACO Test Points and Trimpots.
The AMC-S2e supplies a linear 1.0 to 3.3V DC signal to the monitor. This translates to a 0.4 to
1.32 volt DC signal at TP4.
transmitters full scale calibration. The WARNING threshold is set using trimpot RV5. The
ALARM threshold is set using trimpot RV6.

5.6.1 Equipment Required

Digital multimeter with a minimum display range of 0 to 10.0 VDC.
Miniature screwdriver trimmer adjustment tool.
Zero & Span gases (contact AMC for information)

5.6.2 Calibration

The AMC-S2e is delivered already calibrated from the factory. When the AMC-S2e-CO is
calibrated, it will have a 3.3V at TB1 (or 1.32V at TP4) output in the presence of 100 PPM CO
gas. The zero gas output of the AMC-S2e varies from unit to unit. This calibration procedure
Gas Monitor 1ACO One-Channel Monitor
Sensor
TP4
TP5
TP6
TP7
RV5
RV6
R1
The alarms are set to correspond to some fraction of the
R1
TP7
1
JB8
JB10
TB1
16
TP6
RV6
RV5
TP5
TP4