Multi-Parameter Water Quality Monitor
INSTRUCTION
SHEET
Shop online at: omega.com e-mail: info@omega.com
For latest product manuals: omegamanual.info
MADE IN TAIWAN

Specifications

Range
EC: 0.00 to 19.99 mS
CF: 0.0 to 199.9
TDS: 10 to 19990 ppm
°C: -50 to 70°C
Resolution
EC: 0.01 mS
CF: 0.1
TDS: 10 ppm
Accuracy
pH: ±0.1
°C: ±1.0
mV: ±0.1% F. S ±1 digit
EC, CF, TDS: ±2% F S
Operating Temperature Compensation: 0 to 50°C
Power: Rechargable battery (included) with 120Vac power adaptor (included)
Dimensions: 155 x 86 x 22 mm
Weight: 230 g
Front Panel Description
1 Display
2 Power adaptor receptacle
3 pH action function switch
4 ORP action function switch
5 pH 4.0 calibration inset
6 pH 7.0 calibration inset
M4930/0511
pH: 0.00 to 14.00
mV: 0 to ±1999
°F: -58 to 158°F
pH: 0.01
mV: 1
°C: 0.1 °F: 0.2°
7 °C/°F action function switch
8 EC/CF/ppm action function switch
9 EC/CF/TDS calibration inset
pH or ORP probe inset
10
EC/CF/TDS/Temperature probe inset
11
ON/OFF switch
12
SM
Extended Warranty
Program

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the PHH-127 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Omega Engineering PHH-127

  • Page 1 Extended Warranty Program PHH-127 Multi-Parameter Water Quality Monitor M4930/0511 INSTRUCTION SHEET Shop online at: omega.com e-mail: info@omega.com For latest product manuals: omegamanual.info MADE IN TAIWAN Specifications Range EC: 0.00 to 19.99 mS pH: 0.00 to 14.00 CF: 0.0 to 199.9 mV: 0 to ±1999...
  • Page 2 Operating Information 1. Connect the power adaptor to the power supply outlet. 2. Connect the pH or ORP electrode to the BNC on the instrument. Remove the protective cap. 3. Turn on the meter. 4. Immerse the electrode in the solution to be tested. Stir gently and wait for the reading to stabilize.
  • Page 3 It is the policy of OMEGA Engineering, Inc. to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification.

Table of Contents