Sheree
February 24, 2025

I’m trying to reset the previous blood pressure readings the steps in the manual are not working or I’m doing something wrong

1 comments:
Mr. Anderson
February 24, 2025

To reset previous blood pressure readings on the MHC Medical Products SureLife Premium SL-BA202V:

1. Press the “PAGE” button to select the User Page you want to delete.
2. Press and hold the “MEMO” and “PAGE” buttons simultaneously for 5 seconds.
3. The symbol “EE EE” will be displayed, and all past readings stored in that User Page will be deleted.

This answer is automatically generated

Summary of Contents for MHC Medical Products SureLife Premium SL-BA202V

  • Page 3 Avoid smoking, eating, drinking ca einated drinks or exercising for 30 minutes before taking measurement. 1. Sit upright in a chair at a table with both feet at on the oor. 2. Remove tight- tting/ thick clothing from your upper arm. 3.
  • Page 5 Display Sticker Note: Peel o and remove the display sticker before use...
  • Page 6: Battery Installation

    2.1 Battery Installation 1. Press down and slide the lock to remove the battery cover. 2. Insert 4 AA batteries with the polarities in the correct direction (shown below). 3. Replace the battery cover. Press it lightly until the lock securely clicks into place. Note: Replace batteries when the battery indicator “...
  • Page 7: Date And Time Setting

    2.2 Date & Time Setting...
  • Page 8: Alarm Setting

    Note: • If you select Clock Function “ON”, you will continue to the Alarm Setting section. Otherwise setting will end. • When the clock function is enabled, the screen turn o after 3 minutes of inactivity. To access the alarm settings, you must rst follow steps 1–7.
  • Page 9 2.4 Language/Sound Setting...
  • Page 10 Note: • Avoid smoking, eating, drinking any ca einated drinks or exercising for 30 minutes before taking measurement. • Do not move or talk during measurement. 3.1 Taking Your Measurement 1. Relax 2. Sit upright in a chair at a table with both feet at on the oor.
  • Page 11: User Setting

    3.3 User Setting This unit allows up to 4 users to store measurement records in di erent User Pages. Each User Page can store up to 99 measurements. ” button to switch between User Page 1, 2, 3, Press “PAGE and 4.
  • Page 12: Irregular Heartbeat

    Irregular Heartbeat This unit has a unique feature that alerts user of irregular heartbeat detection during measurement. Note: An irregular heartbeat is de ned as a heartbeat rhythm that has a variation of more than 25% from the average rhythm detected during measurement. If such irregular rhythm occurs twice or more during the same measurement, the display screen will ash between the pulse value and “IH”...
  • Page 13 Regular but Below Normal Heartbeat If the monitor detects a pulse rate below 60 beats per minute during measurement, the monitor will ash between the pulse value and “LO” symbol. Example: (58/LO) Regular but Above Normal Heartbeat If the monitor detects a pulse rate above 100 beats per minute during measurement, the monitor will ash between the pulse value and “HI”...
  • Page 15: Delete Past Readings

    4.3 Delete Past Readings ” button to select the User Page 1. Press the “PAGE you want to delete. 2. Press and hold the “ MEMO” and “PAGE ” button simultaneously for 5 seconds. Symbol of “EE EE” will be displayed and all past readings stored in that User Page will be deleted.
  • Page 16 Problem Cause Solution No display when Have the batteries Replace new you press the run out? batteries. Power/Start Are the battery Re-align batteries button polarities wrong? in correct polarities. Check if the cu is “Err P” displayed Fail to in ate properly connected.
  • Page 17 • Keep the unit awa from direct sunlight, extreme temperatures, humidity or moisture. • Use a dry, soft cloth to clean the unit or, if desired, use a cloth lightly dampened with water. • Do not use alcohol, benzene, thinner or other volatile liquids to clean the unit.
  • Page 19 A Full 5-Year Warranty on Your Monitor The SureLife® Talking Arm Blood Pressure Monitor, excluding the arm cu , is warranted to be free from defects in materials and workmanship within 5 years from the date of purchase. A Full 1-Year Warranty on Your Arm Cu The arm cu is warranted to be free from defects and workmanship within 1 year from the date of purchase.
  • Page 20 MHC Medical Products, LLC 8695 Seward Road Fair eld, Ohio 45011 www.MHCmed.com Made in China. Distributed by MHC Medical Products, LLC. Package design © 2010 SureLife® is a trademark of MHC Medical Products, LLC. Manufactured in accordance with FDA and International Standards...