Psg Dac Cables; Capno Rtc Cable; Figure 2. Replacing The Moisture Trap/Filter - Nonin RespSense II LS1R Operator's Manual

Capnography monitor
Table of Contents

Advertisement

1. Place the filter in the moisture trap so the silicone gasket faces up and hydrophobic
material fits inside the moisture trap opening (figure 2-A).
2. Using the guide marks on the back of the monitor, align the groove at the top of the
moisture trap with the ridge at the top of the trap housing on the monitor (figure 2-B).
3. The trap is tilted in top first, then bottom. Press the moisture trap into position using the tab
(figure 2-C). There should be a tactile click and the moisture trap should be flush with the
side and back of the monitor.
4. To remove the moisture trap and replace the filter, use the tab to pull the moisture trap
away from the monitor. Remove the filter from the moisture trap. If the filter is not in the
moisture trap, check to see if it is still attached to the monitor.
A

PSG DAC Cables

Nonin's PSG DAC (Polysomnography Digital-to-Analog Converter) cables connect the monitor to
a polysomnograph to record data (EtCO
Refer to the individual PSG DAC cable instructions for use for more information.

Capno RTC Cable

Nonin's Capno RTC digital USB cable transmits real-time data from the monitor to another device
(e.g., computer).
Refer to the Capno RTC instructions for use for more information.
14

Figure 2. Replacing the Moisture Trap/Filter

, respiratory rate, SpO
2
System Components
B
C
and CO
).
2
2