KANE515 ® True RMS 1000V Digital Multimeter w/Temperature INSTRUCTION MANUAL ENGLISH ® RANGE MIN/MAX HOLD FAST °F°C Temp 600V True RMS CAT III 1000V CAT II 1.800.561.8187 information@itm.com www. .com...
IMPORTANT SAFETY WARNINGS WARNING Read entire Safety Notes section regarding potential hazard and proper instructions before using this meter. In this manual the word “WARNING” is used to indicate conditions or actions that may pose physical hazards to the user. The word “CAUTION”...
OVERVIEW ® RANGE MIN/MAX HOLD FAST °F°C Temp True RMS A. Apo: Auto power off after 30 minutes of use. Press and hold the HOLD button while turning the meter on F. Function Dial: Turns on meter and is used to select to disable Apo.
SYMBOLS Negative AC/DC Voltage or Current Auto-ranging Overload: Range Exceeded Auto power off Active Low Battery Maximum measured Minimum measured Hold/Capture Value value displayed value displayed Voltage Amps µF Microfarad Microamps Milliamps Diode Nano Farads Ohms/Resistance Continuity Degrees Fahrenheit Degrees Celsius Fuse Warning or Caution Ground...
Resistance: <40M ® RANGE MIN/MAX HOLD FAST °F°C °F°C Features: RANGE MIN/MAX HOLD Temp WARNING True RMS • Do not measure resistance on a live circuit. Ranges Accuracy Resolution Overload Protection 600 to 6M ±(0.8% +5 dgts) 0.1 to 0.001M 600V ±(1.5% +5 dgts) 0.01M...
Diode GOOD DIODE BAD DIODE Open Diode Displays "OL" ® Both directions Forward Bias Displays approx. RANGE MIN/MAX HOLD FAST voltage drop °F°C °F°C '0' Both directions (shorted) • Forward voltage drop if forward biased. Reverse Bias Temp • “OL” if reverse biased. Displays "OL"...
AC/DC Milliamps: 400mA ® ® µ A RANGE MIN/MAX HOLD RANGE MIN/MAX HOLD FAST FAST Keep hands behind test lead guards when measuring current levels. Features: RANGE MIN/MAX HOLD ACµA Ranges Accuracy Resolution Overload Protection ±(1.2% +5 dgts) 600mA /1000V Fast Fuse DCµA Ranges Accuracy...
AC/DC Amps: <10A ® RANGE MIN/MAX HOLD FAST WARNING Keep hands behind test lead guards when measuring current levels. • Do not attempt to measure more than 10A AC. Features: HOLD RANGE MIN/MAX AC AMPS Ranges Accuracy Resolution Overload Protection 6A to 10A ±(1.5% +5 dgts) 0.001A to 0.01A...
Test Lead Notes CATIII 1000V Measurement Locations • Ensure the test lead shield is pressed firmly in place. Failure to use the CATIV shield increases arc-flash risk. CATII 1000V Measurement Locations 18mm 18mm • CAT IV shields may be removed for CAT II locations. This will allow testing on recessed conductors such as standard wall outlets.
Kane shall not be liable for loss of use of the instrument or other incidental or consequential damages, expenses, or economic loss, or for any claim or claims for such damage, expenses or economic loss.