Product Approvals - ACR Electronics Nautilus Reference Manual

Table of Contents

Advertisement

Nautilus Data Loggers
6
Equation:
Nautilus 135
General
Battery:
Battery Life:
Internal Temperature Sensor
Type:
Range:
Accuracy:
Calibration:
Equation:
Specifications are subject to change without notice.
1.1.7

Product Approvals

Certified to CE standard EN61326: 1977 + A1: 1998 (European Emissions and Immunity)
covering ESD, RFI, EFT /B, Surge, Conducted Immunity, and Voltage Dips and Interruptions
Meets FCC standard 47 CFR Part 15, Subpart B: 1999, Class B, (US Radiated and Conducted
Emissions)
Use Equation [45] ET-016 Temperature in TrendReader
software. Other equations for this thermistor type may be
selected but will give incorrect results. You may also create
your own equations, please refer to the Equations section or
contact
Customer Service
3.6 Volt Lithium, 0.85 Amp-hour
3 years under normal use (factory replaceable)
Negative temperature Coefficient (NTC) Thermistor
10,000 ohms @ 25°C (77°F)
10°C to 135°C (50°F to 275°F)
± 0.5°C (± 0.9°F)
Factory calibration verification and NIST certificates are
available upon special request.
Use Equation [47] ET-081 Temperature in TrendReader
software. Other equations for this thermistor type may be
selected but will give incorrect results. You may also create
your own equations, please refer to the Equations section or
contact
Customer Service