OPERATION MANUAL-ECHO-FLEX™ 5002
OPERATION MANUAL
Revision: January 2021
ECHO-FLEX™ 5002
IBIOM Instruments Ltd. 670 Joseph-Louis-Mathieu Street, Sherbrooke, Quebec, J1R 0X3
Tel.: 1-877-563-5777 ● Fax: 450-445-9837 ● info@ibiom.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ECHO-FLEX 5002 and is the answer not in the manual?

Questions and answers

Summary of Contents for IBIOM Instruments ECHO-FLEX 5002

  • Page 1 OPERATION MANUAL-ECHO-FLEX™ 5002 OPERATION MANUAL Revision: January 2021 ECHO-FLEX™ 5002 IBIOM Instruments Ltd. 670 Joseph-Louis-Mathieu Street, Sherbrooke, Quebec, J1R 0X3 Tel.: 1-877-563-5777 ● Fax: 450-445-9837 ● info@ibiom.com...
  • Page 2: Table Of Contents

    SPECIFICATIONS ......................23 Product specifications ....................23 Electrical specifications ....................24 Environmental conditions ....................24 Certification .........................25 LIMITED LIFETIME WARRANTY ..................26 WASTE DISPOSAL ......................27 MAINTENANCE ......................28 TROUBLESHOOTING ....................29 SPARE PARTS LIST ......................30 © Copyright IBIOM Instruments ltd 2021 Page 2 of 31 Revision: January 2021...
  • Page 3: Contact Us

    This information is available on label located on base of the table. Phone 1-877-563-5777 450-445-9837 Email info@ibiom.com www.ibiom.com Address IBIOM Instruments Ltd 670 Joseph-Louis-Mathieu Street Sherbrooke, Quebec J1R 0X3 Canada © Copyright IBIOM Instruments ltd 2021 Page 3 of 31 Revision: January 2021...
  • Page 4: Intended Use

    In accordance with directive on waste electrical and electronic equipment (WEEE), Product’s electrical system and accessories should not be disposed as unsorted municipal waste. Consult your instructional policies and local regulations regarding disposal. © Copyright IBIOM Instruments ltd 2021 Page 4 of 31 Revision: January 2021...
  • Page 5: Warnings

    WARNING. Keep all cables out of reach of children. WARNING. Keep all cables away from moving parts. WARNING. Do not move table by pulling hand control, power cable or any other cable. © Copyright IBIOM Instruments ltd 2021 Page 5 of 31 Revision: January 2021...
  • Page 6 WARNING. Defective or damaged batteries may leak acid and adequate precautions must be taken during handling and transportation. WARNING. Do not open the battery case as damage to the cell or circuitry may develop excessive heat. © Copyright IBIOM Instruments ltd 2021 Page 6 of 31 Revision: January 2021...
  • Page 7: Label

    Double electrical insulation Obligation to read the operation manual Required voltage (120V), alternating current Frequency (60Hz), max amperage (3.9A) Maximum activation time followed by minimum deactivation time © Copyright IBIOM Instruments ltd 2021 Page 7 of 31 Revision: January 2021...
  • Page 8: Operation Instructions

    4. If table is equipped with a battery (option), leave table plug to the wall for 12 hours to recharge battery. BATTERY 5. Table is ready to use. © Copyright IBIOM Instruments ltd 2021 Page 8 of 31 Revision: January 2021...
  • Page 9: Installation

    ALL POSITIONS. Dimensions show maximum space needed. Dimensions may be less than the one indicated below depending on options. © Copyright IBIOM Instruments ltd 2021 Page 9 of 31 Revision: January 2021...
  • Page 10 OPERATION MANUAL-ECHO-FLEX™ 5002 © Copyright IBIOM Instruments ltd 2021 Page 10 of 31 Revision: January 2021...
  • Page 11 4. Remove any remaining packaging from the product. Example: Protective plastic on the hand control cable. 5. Connect the power cord. 6. Perform an operation test to ensure that everything is working properly. © Copyright IBIOM Instruments ltd 2021 Page 11 of 31 Revision: January 2021...
  • Page 12: Central And Direction Lock

    Red: all 4 casters are locked; • Neutral: all casters unlocked and free to roll; • Green: 1 caster at foot end locks in direction and 3 others remain free. © Copyright IBIOM Instruments ltd 2021 Page 12 of 31 Revision: January 2021...
  • Page 13 Tilt down back rest section Move down table Move up table Tilt up back rest section Move down table Tilt down back rest section Trendelenburg Trendelenburg Reverse Trendelenburg Reverse Trendelenburg © Copyright IBIOM Instruments ltd 2021 Page 13 of 31 Revision: January 2021...
  • Page 14 Tilt down back rest section Triple foot control Button Function Move up table Move down table Tilt up back rest section Tilt down back rest section Trendelenburg Reverse Trendelenburg © Copyright IBIOM Instruments ltd 2021 Page 14 of 31 Revision: January 2021...
  • Page 15: Security Rail And Cardiac Cutout

    To lower security rail, pull handle with one hand and hold security rail with other hand. • To move up security rail, lift up rail and it will engage automatically. Always make sure it is securely locked. © Copyright IBIOM Instruments ltd 2021 Page 15 of 31 Revision: January 2021...
  • Page 16 OPERATION MANUAL-ECHO-FLEX™ 5002 Cardiac cutout opened. Security rail lowered down. Cardiac cutout opened. Security rail lowered down. © Copyright IBIOM Instruments ltd 2021 Page 16 of 31 Revision: January 2021...
  • Page 17: Positioning Cushion

    To lower cushion, pull handle and lower down cushion to horizontal position and release handle. It will lock automatically. Make sure it is well lock in place before using with patient. © Copyright IBIOM Instruments ltd 2021 Page 17 of 31 Revision: January 2021...
  • Page 18: Pole

    WARNING. Never exceed safe working load of 25lb or 11kg on IV pole. • To remove IV pole, simply remove from hanger. • IV pole hanger. © Copyright IBIOM Instruments ltd 2021 Page 18 of 31 Revision: January 2021...
  • Page 19: Rechargeable Battery

    It is possible to use battery while it is plugged to wall outlet without any risk of damage. • When battery level is low and needs to be charged, it will emit audible signal. © Copyright IBIOM Instruments ltd 2021 Page 19 of 31 Revision: January 2021...
  • Page 20: Cardiac Cutout In Backrest

    To open cardiac cutout, pull handle located under mattress and pivot cutout. • To close cardiac cutout, pull handle located under mattress, pivot cutout to closed position. It will lock automatically. Make sure it is locked properly. © Copyright IBIOM Instruments ltd 2021 Page 20 of 31 Revision: January 2021...
  • Page 21: Cleaning Instructions

    • Always follow manufacturer’s cleaning products instructions. • Clean spill as soon as possible to prevent penetration and becoming a permanent stain. © Copyright IBIOM Instruments ltd 2021 Page 21 of 31 Revision: January 2021...
  • Page 22: Cleaning Procedure

    Please contact IBIOM for further details. 3. Clean all parts that may be contaminated and/or that are in contact with the patient, such as security rails, IV pole, handles, table base, etc. © Copyright IBIOM Instruments ltd 2021 Page 22 of 31 Revision: January 2021...
  • Page 23: Specifications

    • NFPA 260 • UFAC Class 1 IBIOM Instruments Ltd. reserves the right to change specifications without notice. * May vary depending on options. ** If cleaning instructions are followed. © Copyright IBIOM Instruments ltd 2021 Page 23 of 31...
  • Page 24: Electrical Specifications

    Triple foot switch: IP20 Button controller: IPX0 Battery: IPX6 Duty cycle Intermittent 2/18 Max 2 minutes on 18 minutes off IBIOM Instruments Ltd. reserves the right to change specifications without notice. 8.3 Environmental conditions Conditions Temperature Relative humidity Pressure Normal condition...
  • Page 25: Certification

    8.4 Certification All IBIOM's product are certified according to CSA standard for electrical medical device except contrary indication. Label is located on the base of the product. © Copyright IBIOM Instruments ltd 2021 Page 25 of 31 Revision: January 2021...
  • Page 26: Limited Lifetime Warranty

    OPERATION MANUAL-ECHO-FLEX™ 5002 9. LIMITED LIFETIME WARRANTY IBIOM instruments ltd has a lifetime warranty for all their products against all manufacturing defects under normal operating conditions as described in manual. These following components are excluded from this lifetime warranty: mattresses, retractable rod mechanisms, brake mechanisms, armrest mechanisms, wheels and all mobile parts.
  • Page 27: Waste Disposal

    MATERIAL GROUP Actuators, lifting column, control box, controls, cables and other Electronic electronic components Metal structure Metal Battery pack Batteries Plastic parts Plastic Mattress Foam and textiles © Copyright IBIOM Instruments ltd 2021 Page 27 of 31 Revision: January 2021...
  • Page 28: Maintenance

    OPERATION MANUAL-ECHO-FLEX™ 5002 11. MAINTENANCE See service manual. © Copyright IBIOM Instruments ltd 2021 Page 28 of 31 Revision: January 2021...
  • Page 29: Troubleshooting

    Make sure there is power in control box. the electrical outlet. If the Wall outlet is defective. problem persists, the control box is defective. © Copyright IBIOM Instruments ltd 2021 Page 29 of 31 Revision: January 2021...
  • Page 30: Spare Parts List

    OPERATION MANUAL-ECHO-FLEX™ 5002 13. SPARE PARTS LIST See service manual. © Copyright IBIOM Instruments ltd 2021 Page 30 of 31 Revision: January 2021...
  • Page 31 OPERATION MANUAL-ECHO-FLEX™ 5002 IBIOM INSTRUMENTS LTD 670 JOSEPH-LOUIS-MATHIEU STREET SHERBROOKE, QUÉBEC J1R 0X3 CANADA www.ibiom.com © Copyright IBIOM Instruments ltd 2021 Page 31 of 31 Revision: January 2021...