The GDS 10+ Oxymed oxygen enrichment monitor has been specifically designed to monitor oxygen levels
in the vicinity of assisted breathing apparatus requiring oxygen cylinders or dedicated oxygen supply lines.
The compact style offers functionality for the monitor to be used as a fixed or a transportable device,
boasting further benefits for quick installation via a standard wall socket outlet.

Advertisement

loading

Summary of Contents for GDS 10+ OXYMED

  • Page 1 OXYGEN MONITOR PRODUCT DESCRIPTION: The GDS 10+ Oxymed oxygen enrichment monitor has been specifically designed to monitor oxygen levels in the vicinity of assisted breathing apparatus requiring oxygen cylinders or dedicated oxygen supply lines. The compact style offers functionality for the monitor to be used as a fixed or a transportable device,...
  • Page 2 1. Connect a digital voltmeter (miIIivolt range) to the X and Y test terminals, the CV is preset in the current source mode. Relay Field Terminals to GDS Alarm Unit 2. Remove the sensor terminal connector from the PCB J4 or yellow JP1 Jumpers AC Relay 20mA wire and adjust the 4mA potentiometer for 4mA (4mV).
  • Page 3 Within easy access and audible distance of local staff The sensor cell mounted in the base of the instrument will require replacement at 18–month intervals, contact your • Within the target area at 1 to 1.5m height equipment supplier for details. www.gds–technologies.co.uk...
  • Page 4 Mains lead supplied as standard. For additional information, please see C1798 GDS Technologies Ltd operates under an integrated quality, health and safety and environmental management system certified to ISO 9001, ISO 45001 and ISO 14001 by NQA Certification Ltd. This document is not contractual and the equipment specification may be modified at any time without prior notice.