HP 85108A System Manual page 83

Pulsed-rf network analyzer system
Hide thumbs Also See for 85108A:
Table of Contents

Advertisement

Rise Time
Settling Time
Equivalent Measurement Bandwidth The e ective
Trigger Level
Trigger Width
Minimum Time Delay
Maximum Time Display
10% to 90% of the test and reference channel
detectors: 300 nanoseconds.
The time required for a ratioed,
high-leveled measurement to reach within 0.1 dB of its nal
value when the system is in pulse pro le mode.
post-detection bandwidth of the test and reference detectors in
the HP 8510.
TTL.
The minimum pulse width to be applied to
the rear panel input of the HP 8510 for the HP 8510 to make a
measurement. (External trigger mode: 100 nanoseconds).
The minimum time span of the x-axis
of the HP 8510 display in pulse pro le mode: 5 nanoseconds.
The maximum time span of the
x-axis of the HP 8510 display in pulse pro le mode: 40
milliseconds.

Advertisement

Table of Contents
loading

This manual is also suitable for:

85108

Table of Contents