HP 85108A System Manual page 128

Pulsed-rf network analyzer system
Hide thumbs Also See for 85108A:
Table of Contents

Advertisement

Hardware States
System Phaselock
Mult. Source
Leveling Source #1
Leveling Source #2
Sweep Mode
1
Instrument States
De ne:
Source 1 Mult. Numer. = 1 Mult. Denom. = 1 O set Frequency = 0.000000000 GHz
Source 2 Mult. Numer. = 1 Mult. Denom. = 1 O set Frequency = 0.020000017 GHz
Receiver Constant Frequency = 0.020000000 GHz
System Parameter
During pulsed-RF
STEP or FRQ LIST
Set source #2 power leveling to Ext. if an HP 8349B ampli er is
used to amplify the LO signal.
System Parameter
Power Source #1
Power Slope src 1
Power Source #2
Power Slope src 2
Pulse con g detector wide BW
Set source #1 power level to the highest level possible
1
without IF overload. Power must be set in the 'normal
BW' mode then changed to the 'wide BW' mode, if
desired.
Set source #2 power to -10 dBm if an HP 8349B
2
ampli er is used to amplify the LO signal
Set to normal BW for nonpulsed-RF operation
3
During operation
operation
of 2nd test set
None
On/Save
Int.
Int.
1
Pulsed-RF
CW
+8 dBm
10 dBm
1
OFF
OFF
10 dBm
N/A
2
2.5 dB/GHz
N/A
normal BW
3
Settings:
Int.
O
Int.
N/A
Any

Advertisement

Table of Contents
loading

This manual is also suitable for:

85108

Table of Contents