INSTRUCTION MANUAL
PERSONAL ULTRA FREEZER
Revision
: 01
Date
: 2012.11.15
Cat. No.
: 9.900.900.752
Snowbird

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ScanCool Snowbird and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for LaboGene ScanCool Snowbird

  • Page 1 INSTRUCTION MANUAL Snowbird PERSONAL ULTRA FREEZER Revision : 01 Date : 2012.11.15 Cat. No. : 9.900.900.752...
  • Page 2 Symbols used in this manual Note   Used to direct attention to a special item. Warning Used in case of danger of a serious accident or lethal injury. LaboGene ApS Industrivej 6-8 Vassingerød DK-3540 Lynge Denmark Phone No. +45 39 40 25 66 Fax No.
  • Page 3: Table Of Contents

    Snowbird Personal Ultra Freezer Table of contents Introduction..........................4 Safety precautions ........................4 Installation..........................4 Description ..........................5 Start-up ............................5 Switch power on .........................5 Start-up the freezer ........................6 Cooling commences ........................6 Operation ...........................6 Programming and default settings ..................6 Maintenance ..........................7 Trouble shooting ........................8 Technical specifications......................9 Safety and supervisory accessories ..................9 External alarm installation ......................9 Accessories for Snowbird freezer ..................9...
  • Page 4: Introduction

    Snowbird Personal Ultra Freezer Introduction This manual describes installation, use and maintenance of the Snowbird personal ultra freezer. The Snowbird is a –80 °C freezer primarily designed for direct installation in laboratories, research departments etc. where easy and quick freezing facilities are required several times a day. The design allows for installation under standard work tables or laboratory benches.
  • Page 5: Description

    Snowbird Personal Ultra Freezer Description User interface and display Front panel Air grille with filter Door handle with key lock Door Wheels at the rear Adjustable feet at the front Mains connection CTS connection Battery backup Rechargable battery Start-up Switch power on Insert the battery into battery holder.
  • Page 6: Start-Up The Freezer

    Snowbird Personal Ultra Freezer Note   The backup battery will only slowly discharge if the main power is disconnected. Therefore it is recommended that the battery is removed if the freezer is not operational for long periods. In order to avoid switching off the freezer unintentionally we recommend to not install a switch between the mains supply and the freezer.
  • Page 7: Maintenance

    Snowbird Personal Ultra Freezer Via the four buttons on the user interface keypad it is possible to change the following conditions: Temperature set point, default -80 °C Max. temperature, alarm for too high temperature, default -60 °C Min. temperature, alarm for too low temperature, default -90 °C Delay for alarm of open door, default 1 minute Change between C and...
  • Page 8: Trouble Shooting

    Snowbird Personal Ultra Freezer Remove the top front panel. This panel is positioned on the freezer housing with guide pins at either side, with locating holes on the main body of the cabinet. Stand in front of the freezer facing it. Take a firm grip of both sides of the front plate and pull it forwards (towards you) using minimum force.
  • Page 9: Technical Specifications

    Snowbird Personal Ultra Freezer Technical specifications Description Specification Temperature range -90 °C to –65 °C ± 2,5 °C Storage volume 87 L DxWxH (internal) 510 x 440 x 390 mm DxWxH (External) 716 x 650 x 862,5 mm Shelf Dimensions 475 x 440 mm Required power supply one phase, 230 V, 50 Hz + earth...
  • Page 10 - Low Voltage Directive, as amended 2004/108/EC - EMC Directive, as amended 97/23/EC - Pressure equipment, as amended 2011/65/EU - RoHS Directive 2012/19/EU - WEEE Directive Lynge, January 2013 Rasmus Sørensen QA Manager LaboGene ApS, Industrivej 6-8, Vassingerød, 3540 Lynge, Denmark 08012013QA...