VERDER CARBOLITE GERO G Series Installation, Operation And Maintenance Instructions page 7

1200c tube furnace controller
Table of Contents

Advertisement

Ensure that the product is placed in such a way that it can be quickly switched off or
disconnected from the electrical supply.
This horizontal tube product is supplied with the controls in an integral base (or, to
order, with the base control panel blanked off and the controls in a separate box). A
horizontal tube support kit.
All G-range tube furnaces have "modular" heating elements in which resistance wire is
partially embedded in vacuum formed light-weight refractory fibre. The modules are
150 mm long. In all standard models the two end elements (one at each end) are rated
at 1170 W maximum, while the other (centre) elements are rated at 780 W.
All models are supplied with a pair of "tube adaptors", which can be made to suit any
work tube size up to 170 mm outside diameter. Spare or replacement adaptors can be
ordered.
The horizontal models generally require no special setting up. Observe the above
paragraph on siting. See section 2.3 for work tube and end adaptor considerations.
Under no circumstances should any objects be placed on top of the product.
Always ensure that any vents on the top of the product are clear of any
obstruction. Always ensure all cooling vents and cooling fans (if fitted) are
clear of any obstruction.
2.0   Installation
7