85670 SAINT PAUL MONT PENIT TEL : +33 (0)2 51 98 55 64 FAX : +33 (0)2 51 98 59 07 Email : info@medicatlantic.fr Site Internet : https://www.winncare.fr USER MANUAL MEDICAL BED KALIN (PO3P+OPTION) 580049 Anglais Creation 2014_01_01 Update 2021_07_08...
1. TRANSPORT AND STORAGE 2. BED ENVIRONMENT CONDITIONS 3. GENERAL USE 3.1. P REREQUISITES BEFORE USE AND INSTRUCTIONS FOR USE 3.1.1. T RAINING AND QUALIFICATIONS OF THE PERSON USING THE DEVICE 3.1.2. I NDICATIONS 3.1.3. C ONTRAINDICATIONS 3.2. C LINICAL BENEFIT DEVICE PERFORMANCE ACTION MECHANISM DEPENDING ON THE FUNCTIONS OPTIONS OR ACCESSORIES USED...
Page 3
8.2. W OODEN BARRIERS 8.2.1. I NSTALLATION OF THE BARRIER 8.2.2. U SE OF THE BARRIER 8.2.3. W A551-00 ALL STOP 8.3. A NGLED LIFTING POLE AND STAND 8.3.1. P OSITIONING OF THE STEM AND THE POLE ON THE BED BASE 8.3.2.
Dear Sir/Madam, You have acquired a WINNCARE medical bed equipped with its accessories, and we thank you for your custom. This class 1 medical device complies with Regulation (EU) 2017/745 relating to medical devices They are tested in conformity with standard IEC 60601-2-52 (2010) and IEC 50637 (2017) in their commercial configurations, including the boards and accessories that we manufacture, so as to ensure you maximum safety and performance.
The user and staff must be trained and aware of the risks associated with using the bed. He must not allow it by children and be vigilant when used by confused or disoriented people. 3.1.2. Indications Suitable for patients with temporary or permanent decreased independence due to their illness or disability. 3.1.3.
Page 6
In the case of the use of infrared remote control(s), WINNCARE allows the establishment of a single bed in the same room (or in a close environment) or a second bed only if the infrared options of 2 beds concerned are different (I and I1).
Brake the wheels. The mains socket should remain accessible to enable the bed to be disconnected quickly. Plug in the power lead, checking that the mains comply with the standards in force and that it is suitable for the supply box voltage. Also ensure that the power lead, the remote control lead as well as the cables of possible other devices are positioned correctly to prevent any risks of getting caught between the moving parts of the bed.
3.4.2. Electromagnetic compatibility The bed is an electromedical device requiring special precautions with regard to electromagnetic compatibility. The device must be installed and put into service according to the EMC information provided in this manual. The use of accessories, transducers, and cables other than those specified or supplied by the manufacturer may cause increased electromagnetic emissions or a decrease in the immunity of the device and may cause improper operation.
Kit for side loading bed S0200 Stainless steel intravenous stand on base Y0300 Lifting pole on U shaped base, fixed height 23.3 Only accessories and boards supplied by WINNCARE guarantee you a compatibility with our products Incompatible accessories and boards can pose RISKS...
6. USE 6.1. Purpose of medical device ➢ The device is intended to compensate for a handicap and as a preventive measure. ➢ These beds are intended : -Application environments 3 (Senior care centres if adult person with atypical anatomy), 4 and 5 (Home care and in-facility when equipped with the emergency CPR back-rest release option).
Foot board Head board Slatted folding leg lift Base and cross pieces Gallows Wheels housings (corner) Intravenous holder rod housing 6.4. Technical characteristics 6.4.1. Dimensional 42° Ø10 cm 12 cm WINNCARE recommend the use of the XS 150 patient lift.
626 mm 240 mm 470 mm 637 mm 637 mm 6.4.2. Bed weight TOTAL Bed with electric knee break leg lift 70 kg (without boards) 6.4.3. Noise The measurement of the maximum audible sound power in accordance with ISO 3746 is 45.2 dB(A). 6.4.4.
6.4.5. Protection level against dust and liquid penetration number (decade) number (unit) Index Protection against solids Protection against water intrusion No protection. No protection. Protected against solid bodies greater than Protected against vertical drops of water drops. 50 mm. Protected against solid bodies greater than Protected against falling drops of water up to 15°...
6.6. Remote control The remote control and its cable are protected in a lockable box when the functions of the bed are not modified. This arrangement avoids any risk of involuntary activation and suffocation. LOCATION OF THE REMOTE CONTROL ON THE BED...
Page 16
Carry out a test cycle when the bed is empty to familiarise yourself with the bed functions. Lifting and lowering of back rest Lifting and lowering of variable height Lifting and lowering of electric folding Simultaneous lifting and lowering of back rest and folding...
KEY TO LOCK A KEY IN VERTICAL TURN THE KEY SLIGHTLY TO FUNCTION POSITION: THE THE RIGHT TO LOCK THE FUNCTION IS FUNCTION UNLOCKED 6.7. Braking 6.7.1. Separate braking Check that the wheels are locked by trying to move the bed. If this is not done, the patient or another person who leans on the bed may fall.
6.7.2. Centralized braking Braked wheels: press pedal (red side) with the foot. Freewheels: Press pedal (A or C) with the foot to get the middle position. C. Directional Wheel: Press pedal (green side) with the foot 7. SLEEPING PLAN 7.1. Back rest 7.1.1.
Version with handle on the back rest Version without handle on the back rest 7.1.2. If the bed is fitted with an emergency release for the back rest (Cardio Pulmonary Resuscitation) 1) Grasp the back rest with one hand. 2) With the other hand, activate one of the two handles on the back rest while lowering.
8.2. Wooden barriers INCOMPATIBLE SIDE BARRIERS CAN CAUSE HAZARDS. The wooden barriers are designed to be used on WINNCARE beds with a length of 1700 mm, with panels provided with the necessary reservations for their installation. The barrier was tested in accordance with the test method of the park bed standard NF EN 50637 :2017.
Page 22
The element composed of the wood and metal bar is always placed in the lower part. The arrow down and to the inside of the bed...
8.2.2. Use of the barrier ➢ To lower the side rail : Take the upper bar and lift it up to its locking at the top of the panel. Check if it is properly engaged.
Page 25
➢ To lower the side rail: 1) l Lift the upper bar 2) Press the locking buttons of each panel.
1) Slide the barrier gently down the bed (make sure that no objects or parts of the patient's body or nursing staff are in the operating area) until the stop is stopped . The upper bar comes to rest on the lower metal / wood bar assembly.
8.3. Angled lifting pole and IV stand 8.3.1. Positioning of the stem and the IV pole on the bed base 8.3.2. Specific features of the jib crane and lifting handle The A681-00 Stem has been sized for junior beds and complies with STANDARD EN 50637...
9. MAINTENANCE No maintenance should be done in the presence of a patient on the bed. 9.1. Identification Warning Manufacturer's address and Medical device) contact details Complies with regulations (EU) Product reference and name Read the instruction manual Product serial number YY : Year of manufacture PP : Product empty weight MM : Month of manufacture...
9.3. Maintenance Quality control of medical beds and its accessories will be made by technical staff or trained biomedical and taking into account the normal conditions of use specified in the user guide. The bed must be available to perform all quality control at least once a year, but also on special request and corrective maintenance on the performance that could be affected by the intervention.
9.4. Quality inspection of medical beds ANNUAL MAINTENANCE OF MEDICAL BEDS IDENTIFICATION ESTABLISHMENT Type model / Trademark : Date of manufacture : Serial or park number : Application Environment Congruent Improper VISUAL CONTROL applicable Identification - label General cleanliness Condition and attachment of head and footboards Lifting pole (fastening, strap, winder, …) Sleeping surface (bedspring) Verification of tightenings, diverse nuts and bolts, pins, pivot...
9.5. Cleaning and disinfection High pressure cleaning, machine cleaning, washing tunnel or jet cleaning is prohibited. Unplug the mains lead. Check that all the electrical parts are connected together. All the sockets of the supply box must be used or filled, otherwise its watertightness is not guaranteed.
Do not rinse or wipe. Check the operation of all the bed functions Repair if necessary Pack in thermoplastic film Warning In the event of additional precautions (Contact precautions, Droplets or Air), apply the measures recommended by the hospital hygiene department The use of a Javel water solution of more than 5000ppm (0.5% of active chlorine) should be justified by a microbiological risk and only applied for the required time (Risk of ageing of some materials, especially their colour).
9.8. Troubleshooting Guide DEFAULT POSSIBLE CASE HELP The whole bed is not working The bed is not electrically powered Check the electrical connection (Green power LED on controlbox) The remote control is fully locked Unlock the remote control The infrared remote control is not powered or is not Remove the rear cover connected to the receiver.
Environment Department of Winncare at n ° +33 (0)2.51.98.55.64 Winncare also encourages you to limit the possible effects of waste on the environment and public health in all other countries, by complying with local selective sorting laws, separating waste from Electrical and Electronic...