OM-CP-RFCURRENT2000A
Additional Features and Operation
Alarm Settings/Rules
In the Omega Software, alarms can be created on two levels; in
the software itself, and programmed specifically to the data logger.
Creating alarm rules in the software allows the user to utilize alarms
while real time recording, and have the added benefit of features
such as e-mail and SMS notification. An alarm that is programmed on
the data logger will only alert the user from the device itself, whether
that is by a flashing LED, audible alarm, or local LCD notification.
Creating a Real time Alarm Rule
1. Within the Device tab in the Omega software, in the Alarms
group, click Manage Rules, then select New.
2. Enter a name in the Rule name box, and select whether
notifications will occur when All conditions are met or Any condition is met.
3. Select the desired conditions to have met before notifications occur, and select which notifications will
occur from the drop down menus.
4. If selecting to receive notifications by email or text message, click Email settings and make sure the correct
info has been entered. Email and text message notifications will not be received if settings are incorrect.
5. Click OK. The OK button will be disabled if the Rule name box is empty.
Changing the Alarm Settings of a Device
1. In the Connected devices panel, select the OM-CP-RFCURRENT2000A device.
2. Click the Properties button under the device tab at the top of the software, and select the Alarm tab in
the Properties window.
3. The High and Low settings indicate reading thresholds at which the alarm becomes active.
4. The Delay setting indicates how long to wait before the alarm becomes active after passing an alarm
threshold.
5. The Use cumulative alarm delay checkbox indicates whether the alarm delay should reset when reading
values fall back within the alarm threshold.
6. Once the alarm is configured appropriately, click Apply to save the settings.
**Please refer to the Help file in the Omega software for
more information regarding alarm settings**
Mounting Instructions
The base provided with the OM-CP-RFCURRENT2000A
can be used in two ways.
The bottom of the logger snaps in place to use and
mount the unit on a table top or horizontal surface.
The base can also securely snap to the backside of
the logger for wall mounting if desired (as shown to
the right).
There are two holes in the base to allow for screws
for mounting.