Blink
code
4
Power time
exceeded
5
Brake failure
No neutral position
6
when switching
Scooter on.
36
Fault
Consequence
for the Scooter
Stops driving
Stops driving
Stops driving
Comments
The maximum current was exceeded
over too long a period, probably because
the motor was overloaded or has been
working against an immovable
resistance. Switch the scooter off, wait a
few minutes and then switch on again.
The electronic system has determined a
motor short-circuit. Check the wiring
harness for short-circuit and check the
motor.
Contact your Invacare® dealer.
Ensure that the disconnection lever is
pressed in.
There is a defect in the braking coil or in
the cabling. Check the magnetic brake
and cabling for open or short-circuited
circuitry. Contact your Invacare® dealer.
Drive lever is not in neutral when the
keyswitch was turned. Put the drive lever
in neutral, turn the power off and then
turn on again.
It may be necessary to recalibrate the
drive lever. Contact your Invacare®
dealer.