56
EN
6.3
Disinfection, Sterilization
If necessary, for example after an infectious
illness or if the device has become unusually
contaminated, you can also disinfect the
housing
, power supply cable
1
system
, the bacteria filter housing
12
the humidifier
. Please observe the operat-
32
ing instructions for the disinfectant used.
We recommend that you wear suitable
gloves when disinfecting the equipment
(e.g. household or disposable gloves).
Unit
•
The housing and the power supply cable
of SOMNOvent S are cleaned by being
simply wiped over with disinfectant. We
recommend TERRALIN
Hose system
•
When disinfecting use the same
procedure as for cleaning.
•
After disinfecting the equipment all
parts have to be thoroughly rinsed with
distilled water and dried.
•
Let the hose system drip. Dry the hose
system using SOMNOvent S as
described on page 50.
•
Steam sterilization of the creased hose
WM 24120 (grey) using equipment
complying with EN 285 is also possible.
Temperature: 134 °C, minimum treat-
ment time 3 minutes.
•
The creased hose WM 24130 (clear)
can also be cleaned using water up to
70 °C. Sterilisation is not permitted.
, the hose
3
and
2
®
for this purpose.