Page 1
Model 378B13 ICP MICROPHONE SYSTEM Installation and Operating Manual For assistance with the operation of this product, contact PCB Piezotronics, Inc. Toll-free: 800-828-8840 24-hour SensorLine: 716-684-0001 Fax: 716-684-0987 E-mail: info@pcb.com Web: www.pcb.com...
Page 2
Piezotronics, user servicing or repair is critical test. not recommended and, if attempted, may void the factory warranty. Routine PCB Piezotronics maintains an ISO- maintenance, such as the cleaning of 9001 certified metrology laboratory and electrical connectors, housings, and offers calibration services, which are...
Page 3
50% of the general contact numbers are: replacement cost returned item(s). PCB will provide a price PCB Piezotronics, Inc. quotation replacement 3425 Walden Ave. recommendation for any item whose Depew, NY14043 USA repair costs would exceed 50% of...
PRODUCT GUIDE 377 SERIES MICROPHONES 426 SERIES PREAMPLIFIERS Contents: Introduction ................1 Standards ..................1 Setting up a Microphone Measurement System ......1 Taking Measurements ..............2 Accessories ................. 3 Maintenance ................3 Calibration .................. 4 Warranty and Service ..............4 1.0 Introduction Thank you for your purchase of a PCB high quality microphone.
Externally Polarized Microphone System Prepolarized Microphone System 4.0 Taking Measurements The following formula is used to calculate the pressure measured by the microphone from the output volt- age signal. The sensitivity of a microphone is typically measured at 250 Hz. Pressure (Pa) = Voltage (mV) / Sensitivity (mV/Pa).
Temperature and pressure will affect the sensitivity of a microphone but these effects can be accounted for by adjusting the recorded microphone values using the coefficients specified for each microphone model. Simply calculate the difference in temperature and ambient pressure from those recorded during calibration and multiply this value by the proper coefficient to determine the sensitivity offset.
Class1 and IEC 60942-2003, Class 1, Sound Calibrators. 8.0 Warranty and Service All equipment and repair services provided by PCB Piezotronics, Inc. are covered by a warranty against defective material and workmanship under a Total Customer Satisfaction policy. If, at any time, for any reason, you are not completely satisfied with any PCB product, PCB will repair, replace, or exchange it at no charge.
Page 10
Model Number Revision: C ICP MICROPHONE SYSTEM 378B13 ECN #: 45005 Performance ENGLISH OPTIONAL VERSIONS Nominal Microphone Diameter 1/2" 1/2" Optional versions have identical specifications and accessories as listed for the standard model except where noted below. More than one option may be used.
Page 11
PCB Piezotronics Inc. claims proprietary rights in REVISIONS the information disclosed hereon. Neither it nor any reproduction thereof will be disclosed to others DESCRIPTION without the written consent of PCB Piezotronics Inc. RELEASED TO DRAFTING 44117 .52 [13.2] MICROPHONE CARTRIDGE .50 [12.7]...