TXR 325D Preinstallation Manual

Standard frequency series generators
Hide thumbs Also See for 325D:

Advertisement

Quick Links

MEDICAL SYSTEMS
Technical Publication
M-2008-7-12
Pre-Installation Manual
STANDARD FREQUENCY SERIES GENERATORS
Contents: Models
TXR 325D
TXR 325M
TXR 425
TXR 525SFQ

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 325D and is the answer not in the manual?

Questions and answers

Summary of Contents for TXR 325D

  • Page 1 MEDICAL SYSTEMS Technical Publication M-2008-7-12 Pre-Installation Manual STANDARD FREQUENCY SERIES GENERATORS Contents: Models TXR 325D TXR 325M TXR 425 TXR 525SFQ...
  • Page 3 January 22, 2004 Added Tables and Illustrations April 4, 2006 General Revision Added TXR 425 Model Designation July 11, 2008 This document is the English original version, edited and supplied by the manufacturer. All Copy including Advisory Symbols: Type Style Arial.
  • Page 4: Table Of Contents

    Standard Frequency Series Generators Pre-Installation TABLE OF CONTENTS Section Page 1 INTRODUCTION ....................1 1.1 Responsibility of Purchaser ...............1 2 PRE-INSTALLATION DATA ................2 2.1 Physical Characteristics ................2 2.2 Method of Mounting ...................2 2.3 Minimum Recommended Free Area for Service Access ......2 2.4 HV Transformer and Console —...
  • Page 5: Introduction

    Standard Frequency Series Generators Pre-Installation SECTION 1 INTRODUCTION This Pre-installation document provides the information and data needed to plan and qualify the customer site prior to equipment delivery and installation. This document considers only the generator and its associated components. Product information, environmental and electrical requirements are specified.
  • Page 6: Pre-Installation Data

    WEIGHT Length Width Height STANDARD CONTROL CONSOLES RAD Control Consoles 457 mm 305 mm 941 mm 61 kg TXR 325 D 18 in 12 in 37 in 135 lbs 457 mm 305 mm 941 mm 61 kg TXR 325 M...
  • Page 7: Hv Transformer And Console - Illustration

    2.4 HV Transformer and Console — Illustration Illustration 2-1 HV Transformer And Console STANDARD FREQUENCY CONTROL 18" 457mm 12" 305mm SIDE BACK VIEW VIEW POWER INLET 11" 279mm TXR 325-1, TXR 625-1 TRANSFORMER 22" 559mm TOP VIEW SIDE VIEW M-2008-7-12...
  • Page 8: Room Requirements

    Standard Frequency Series Generators Pre-Installation SECTION 3 ROOM REQUIREMENTS 3.1 ENVIRONMENTAL REQUIREMENTS LINE POWERED GENERATORS Temperature range of -40 to 70 degree C. Storage / Transport Environmental Relative Humidity range of 10 to 100% Conditions Atmospheric Pressure range of 600 hPa to 1060 hPa Temperature range of 10 to 70 degree C.
  • Page 9: Line Powered Generators - Power Line Requirements

    Standard Frequency Series Generators Pre-Installation 3.3 LINE POWERED GENERATORS – POWER LINE REQUIREMENTS Operation TXR 425 GENERATOR MODEL TXR 325D TXR 325 M 40 kW Maximum Power kW 40 kW 40 kW 400 mA Maximum mA 400 mA 400 mA...
  • Page 10: Maximum Line Impedance

    Standard Frequency Series Generators Pre-Installation 3.4 MAXIMUM POWER LINE IMPEDANCE • Maximum Power Line Impedance. The Impedance of the Power Line in the installation must be lower than the maximum value indicated below: SINGLE-PHASE GENERATORS POWER LINE VOLTAGE 40 Kw 40 Kw 50 Kw 208 VAC...
  • Page 11: Interconnection And Grounding Requirements

    Standard Frequency Series Generators Pre-Installation 3.6 INTERCONNECTION AND GROUNDING REQUIREMENTS Every installation must be provided with a main line disconnect device (thermomagnetic breaker) and the remote disconnect devices required at all Consoles that are not located next to the line safety switch. (For more information about interconnection and grounding refer to “Installation” Document).
  • Page 12: Safety Requirements

    Standard Frequency Series Generators Pre-Installation Illustration 3-3 Grounding Diagram (EARTHING DIAGRAM) 3.7 SAFETY REQUIREMENTS Devices such as Safety Switch / Emergency Switch, Warning Lights, and Door interlock Switch should be supplied and installed by the customer. (Refer to illustration 3-4, 3-5 and 3-6). THIS PART OF PAGE INTENTIONALLY LEFT BLANK M-2008-7-12...
  • Page 13: Safety Switch / Emergency Switch

    Standard Frequency Series Generators Pre-Installation 3.7.1 SAFETY SWITCH / EMERGENCY SWITCH The main Safety Switch should be installed in the Room Electrical Cabinet (Room Disconnect) (close to the Generator Cabinet), and provided with light indicators for “Power On / Off”. It should be used for main disconnection, and located in an accessible place where it can be seen and controlled during operation and service.
  • Page 14: Door Interlock Switch

    Standard Frequency Series Generators Pre-Installation 3.7.2 DOOR INTERLOCK SWITCH The Door Interlock Switch indicates to the operator when Doorways to the X-ray room are open. It inhibits the X-ray generation by preventing Prep and Expose Commands, according to the Local Standards and customer preferences.
  • Page 15: Remote Ready State (Warning Lights)

    Standard Frequency Series Generators Pre-Installation 3.7.3 REMOTE READY STATE (WARNING LIGHTS) The Warning Lights are signal lamps installed outside of the X-ray room (near to the main entrance) that indicates: The system is under voltage (Red lamp “ON”). X-Ray exposure is in process (Yellow lamp “ON”). For connections see illustration 3-6 below. The customer is responsible for ordering and installing the Warning Light Relay Coils and the Warning Lights. Wire to customer preference (one or both lamps).
  • Page 16: Certifications And Classification

    CONTINUOUS MODE NOMINAL ELECTRIC POWER OUTPUT 6.1 CONTINUOUS MODE NOMINAL ELECTRIC POWER OUTPUT (KW) • Continuous mode nominal power for 0.1s at 100 kV is as follows: TXR 325D,TXR 325M, and TXR 425 40kW. Model TXR 525 SFQ, 50kW. Reference IEC 60601-2-7 CL. 6.8.2 4) For Additional Information see the Installation and Operation Manuals.
  • Page 17: Sample Room Plans

    Standard Frequency Series Generators Pre-Installation SECTION 7 SAMPLE ROOM PLANS 7.0 Systems without X-Ray Tables NOTE: If MA/RA stand used focal center from column is 22.5". THIS PART OF PAGE INTENTIONALLY LEFT BLANK M-2008-7-12...
  • Page 18: Systems With Tables (Medical)

    Standard Frequency Series Generators Pre-Installation 7.1 Systems with X-Ray Tables (medical) NOTE: If CSM stand used top rail extends 12" from the wall. THIS PART OF PAGE INTENTIONALLY LEFT BLANK M-2008-7-12...
  • Page 20 5481 Skyland Boulevard East • Cottondale, Alabama 35453 USA Telephone toll free in the USA: 1(800) TXR-X-Ray Telephone (205) 556-3803 • Fax (205) 556-3824 • www.txr.com PUBLICATIONS AVAILABLE FOR STANDARD FREQUENCY SERIES UNITS A full set of manuals are provided with each x-ray control.