INSTRUCTIONS POUR L'EMPLOI

L'APPAREIL EST UN ASPIRATEUR PORTATIF POUR CHIRURGIE A L'EXCLUSION DU DRAINAGE THORACIQUE.
TEMPÉRATURES DE SERVICE: COMPRISES ENTRE 0 ET 50°
HUMIDITÉ DE SERVICE COMPRISE ENTRE 20 ET 90% D' H.R.
A PLUS DE 2500 M D'ALTITUDE, LA FORCE D'ASPIRATION PEUT DIMINUER.
A)
VISSER FORTEMENT LE COUVERCLE SUR LE BOCAL POUR ASSURER UNE PARFAITE ETANCHÉITÉ.
B)
RELIER LE TUYAU COURT AVEC FILTRE [1] AU RACCORD "ASPIRATION" [2]
C)
RELIER L'AUTRE EXTRÉMITÉ DU TUYAU AU RACCORD "VACUUM" [3] AVEC FLOTTEUR SUR LE COUVERCLE DU BOCAL
D)
RELIER LE TUYAU LONG [4] A L'AUTRE RACCORD [5] "PATIENT" SUR LE COUVERCLE DU BOCAL (RACCORD SANS FLOTTEUR)
E)
RELIER L'AUTRE EXTRÉMITÉ DU TUYAU LONG [4] AU RACCORD [6] POUR ENFILER LA SONDE ET, LE CAS ÉCHÉANT, LA SONDE EN
DOTATION [10]
F)
BRANCHER L'APPAREIL A LA PRISE DU SECTEUR, APRES AVOIR VERIFIE QUE LE VOLTAGE CORRESPOND BIEN A CELUI QUI EST
INDIQUE SUR LA PLAQUE POSEE A L'ARRIERE [11] .
G)
PRESSER L'INTERRUPTEUR ON/OFF [7]
H)
ON PEUT VARIER L'INTENSITÉ D'ASPIRATION AU MOYEN DU DISPOSITIF DE RÉGLAGE [8] SITUÉ SUR LE DEVANT DE L'APPAREIL
ET LA CONTRÔLER SUR LE VACUOMÈTRE [9] QUI SE TROUVE ÉGALEMENT SUR LE DEVANT DE L'APPAREIL
I)
IL EST CONSEILLÉ DE PROCÉDER À UN PREMIER CYCLE D'ASPIRATION AVEC AU MOINS 500 ML D'EAU PURE
J)
LE LIQUIDE ASPIRÉ COULERA DANS LE BOCAL. SI LA QUANTITÉ DE CE LIQUIDE EST SUPÉRIEURE À LA CAPACITÉ DU BOCAL, LA
VALVE DE TROP-PLEIN ARRÊTE L'ASPIRATION.
K)
ÉTEINDRE L'APPAREIL IMMÉDIATEMENT APRÈS USAGE EN PORTANT L'INTERRUPTEUR ON/OFF [7] SUR LA POSITION "OFF".
LE FABRICANT, L'ASSEMBLEUR, L'INSTALLATEUR OU L'IMPORTATEUR NE SONT CONSIDÉRÉS COMME RESPONSABLES DE LA
SÉCURITÉ, DU BON FONCTIONNEMENT ET DE LA FIABILITÉ DE L'APPAREIL QUE SI L'INSTALLATION ÉLECTRIQUE SUR LAQUELLE IL EST
BRANCHÉ EST CONSTRUITE SELON D.L. 46/90.

ACCESSOIRES EN DOTATION

CODE N°
05.00077 + 05.00063
05.0012
05.0016
03.2507
SI L'ON UTILISE DES ACCESSOIRES AUTRES QUE CEUX FOURNIS PAR FAZZINI S.R.L., IL EST INDISPENSABLE DE S'ASSURER QUE
LEURS CARACTERISTIQUES SONT IDENTIQUES A CELLES DES ACCESSOIRES D'ORIGINE.
NOTA BENE: LES BOCAUX FOURNIS N'ONT PAS UNE FONCTION DE MESURE: ILS NE FOURNISSENT QU'UNE INDICATION DE LA
QUANTITE QU'ILS CONTIENNENT.
CE QU'IL FAUT FAIRE APRÈS UTILISATION DE L'APPAREIL.
ATTENTION: TOUJOURS SE SERVIR DES MOYENS DE PROTECTION INDIVIDUELS (COMME LES GANTS DE CAOUTCHOUC PAR EX.)
CHAQUE FOIS QU'ON ENTRE EN CONTACT AVEC DES PARTIES POUVANT ÊTRE CONTAMINÉES.
DÉTACHER LES PARTIES JETABLES ET LES ÉLIMINER.
S'ASSURER QUE L'ASPIRATEUR EST EN PARFAIT ÉTAT.
PROCÉDER À SON NETTOYAGE COMME INDIQUÉ CI-APRÈS.
IL EST OPPORTUN DE STÉRILISER LES TUYAUX D'ASPIRATION APRÈS CHAQUE UTILISATION.
LE FILTRE EST JETABLE; IL FAUT DONC LE REMPLACER CHAQUE FOIS QU'ON S'EST SERVI DE L'APPAREIL ET S'ASSURER, AVANT DE
L'UTILISER À NOUVEAU, QUE CE REMPLACEMENT A ÉTÉ FAIT
NETTOYAGE
BOCAL : LE BOCAL PEUT ÊTRE MIS EN AUTOCLAVE À UNE TEMPÉRATURE N'EXCÉDANT PAS 121 ºC POUR UN MAXIMUM DE 3
MINUTES, AU-DELÀ DESQUELLES IL RISQUE D'ÊTRE ENDOMMAGÉ. IL PEUT AUSSI ÊTRE DÉSINFECTÉ AVEC DE L'ALCOOL
DÉNATURÉ OU UNE SOLUTION À BASE D'HYPOCHLORITE EN VENTE DANS TOUTES LES PHARMACIES.
LIMITE MAXIMUM DE RÉUTILISATION: 30 CYCLES.
ASPIRATEUR : UNE FOIS L'APPAREIL DÉBRANCHÉ, LE NETTOYER AVEC UN CHIFFON DOUX IMBIBÉ DE DÉTERGENT NON
ABRASIF.
F‐X50.60.90.100.XX_User Manual_ Rev.6_09.07.2010.doc 
F-X50 / F-60 / F-90 / F-100
DESCRIPTION
SÉRIE DE TUYAUX
SONDE ASPIRATION CH 16 (USAGE UNIQUE)
FILTRE ANTIBACTERIEN/HYDROFUGE (USAGE UNIQUE)
CABLE D'ALIMENTATION
ASPIRATORE
Istruzioni per l'uso
SUCTION PUMP
Instructions for use
ASPIRATEUR
Mode d'emploi
ASPIRADOR
Manual de instrucciones
ASPIRADOR
Instruções de uso