Correlation Flag
Correlation Type
Default
User Correlation
MGL Correlation (for
Ca)
Flame Correlation
Error Flag
Correlation Type
MV Range Error
Noise Error
Drift Error
Range Low
Range High
Too Low
Too High
Not Calibrated
Not Consistent
Cannot Measure
Li Cannot Measure
Reading 1 mV Sign
This character represents if value is positive or negative. "+" represents positive and "-" negative
numbers. During calibration, this represents STD-A. During sample analysis, it represents
Sample mV.
Reading 1 mV Value
Ion concentration value is sent out in 7 Digits. This number is 1000 times the concentration
determined in the analysis. For example, 01234560 represent value of 1234.56. During
calibration, this represents STD-A. During sample analysis, it represents Sample mV.
Reading 1 mV Try Count (Valid only for SMARTLYTE Calibration)
The number represents number of tries instrument ran the standard fluid before producing stable
results. During calibration, this represents STD-A.
Reading 2 mV Sign
This character represents if value is positive or negative. "+" represents positive and "-" negative
numbers. During calibration, this represents STD-B. During sample analysis, it represents
Standard Fluid mV.
Reading 2 mV Value
Ion concentration value is sent out in 7 Digits. This number is 1000 times the concentration
determined in the analysis. For example, 01234560 represent value of 1234.56. During
calibration, this represents STD-B. During sample analysis, it represents Standard Fluid mV.
Flag Code
D
U
M
F
Flag Code
V
N
D
L
H
<
>
X
!
@
M
ECO 9085 SOP05-5038F Rev.05 Effective Date: 12/06/18 Page 74 of 81