4 Principles of operation

Command bar

The Command bar gives you a quick and direct access
to frequently used functions, regardless of the specific
type of task.
Freeze
You can freeze the A-scan, for example for measure-
ments on hot test objects, for measurements in difficult
coupling conditions, or for spot weld testing.
When the A-scan is frozen, the icon is colored and a cor-
responding status indicator is displayed above the A-
scan (see page 6 at the beginning of the operating man-
ual).
You can define the Freeze Mode (see page 111).
66
AutoXX
You can automatically set the first echo to a certain
screen height. This function is useful, for example, when
recording DAC curve points.
You can define the screen height (default = 80%) for the
echo amplitude (Auto XX Amplitude, see page 108).
Calibrate
During calibration this function is used to record the
backwall echoes from the calibration block.