Device output parameters
The waveform below demonstrates a typical output waveform form the Kegel Toner
device:
Max Pulse Energy: 23 C
Use special precautions regarding EMC according to
the information provided below. Other portable and
mobile RF communications equipment can affect
performance. Do not use when adjacent or stacked
with other electrical equipment.
17. EMC
Wireless communications equipment such as wireless home network devices, mobile
phones, cordless telephones and their base stations, walkie-talkies can affect this
equipment and should be kept at least a distance d = 3,3 m away from the equipment.
(Note. As indicated in Table 6 of IEC 60601-1-2:2007 for ME EQUIPMENT, a typical
cell phone with a maximum output power of 2 W yields d = 3,3 m at an IMMUNITY
LEVEL of 3 V/m).
Use special precautions regarding EMC according to the information provided below.
• Other portable and mobile RF communications equipment can affect performance.
• Do not use when adjacent to or stacked with other electrical equipment.
• Use of leads or probes other than those listed in section 13 may affect EMC
performance.
Note: For hospital use, full EMC advice tables are available on request.