MRC Rova-M-C3 User's Operating Manual

Recirculating cooler

Advertisement

Quick Links

User's Operating Manual
Recirculating Cooler
Rova-M-C3
3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529
mrc@mrclab.com
MRC.4.21

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Rova-M-C3 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for MRC Rova-M-C3

  • Page 1 User’s Operating Manual Recirculating Cooler Rova-M-C3 3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 mrc@mrclab.com MRC.4.21...
  • Page 2 Congratulations! You have made an excellent choice! First of all, MRC thanks you for the trust you have placed. This operating manual has been designed to help you gain an understanding the principles of operating and possibilities of our bath liquid circulation unit. For optimum utilization of all functions, we recommend that you thoroughly study this manual prior to beginning operation.
  • Page 3: Table Of Contents

    Table of Contents Chapter I System Introduction................- 1 - Chapter II Installation Instruction................- 2 - Chapter III User’s Guide....................- 4 - 3.1 Preparations before Operation..............- 4 - 3.2 Controller Panel....................- 5 - 3.3 Operation Guide....................- 7 - 3.3.1 Switching on................... - 7 - 3.3.2 Setting the Setpoint Temperature.............
  • Page 4: Chapter I System Introduction

    Recirculating cooler Chapter I System Introduction The recirculating cooler, also called as chiller, draws heat from bath liquid via the cooling system for instruments or equipments that need to be cooled (Water is a common bath liquid, synthetic bath liquid, such as glycol water is used for lower temperature) to ensure instruments or equipments to work properly under the regular temperature.
  • Page 5: Chapter Ii Installation Instruction

    Recirculating cooler Chapter II Installation Instruction The recirculating cooler shall be equipped indoor with good environment. The principles of choosing the location as following: Location Principle Firm ground or experiment table in a dust-free and cool environment with good ventilation Far from droplets or steam Far from the location that influx, viscous flow, leak of fuel gas may happen...
  • Page 6 Recirculating cooler Unit installation: Unpack the unit and check the packing list to ensure accessories to be complete. Please contact with our sales services for solutions when there is a shortage. Move the unit from the wooden packing base to spacious position and check for damages incurred during transit.
  • Page 7: Chapter Iii User's Guide

    Recirculating cooler Chapter III User’s Guide 3.1 Preparations before operation 1. Please clean the water tank and workspace before operation. 2. Make sure the flange bolts of the external flexible hose tightened. 3. Power connection: connect the chiller’s power outlet with the single phase power of the workspace (please referring to duty plate of the unit).
  • Page 8: Controller Panel

    Recirculating cooler between 80℃ and 90℃.  Silicone oil is applied commonly when temperature range over 90℃. Warning Bath liquid over 35℃ should not be directly cooled via cryogenic circulator. That is, the compressor could not start when the temperature of the bath is over 35℃. 3.2 Controller Panel Above is the picture of the unit’s controller panel: There are two 4-digit LED displays.
  • Page 9 Recirculating cooler There are 8 keys on the controller panel, which are Pump key , Menu , Left-shift key , Add key , Reduce key , Esc key , Enter and Start/Stop key There four indicators function as following: Low water level alarm. No water flow alarm. Warning Red light flashes Protection delay alarm.
  • Page 10: Operation Guide

    Recirculating cooler 3.3 Operation guide 3.3.1 Switching on 1. Turn on the leakage protector. 2. Turn on the power switch. Then the unit begins to initializing interface. When initializing finished, the unit enters standby mode with a buzzer beep; 3. Press Start/Stop key. Check the fluctuation of the outlet pressure gauge and the vibration of the pump.
  • Page 11 Recirculating cooler directly or Menu key indirectly. 1. Single-point high precision mode: This mode is to control the water temperature of the bath in a constant value T00 via PID algorithm. The unit under this mode can also work under the Start/Stop economic mode by setting R_CS and R_CM parameters (please refer to Part 3.4 for details).
  • Page 12: Start/Stop Compressor

    Recirculating cooler than 25℃, the compressor starts to cooling, and the auxiliary display shows the objective cooling temperature as 15 ℃ ; when the water temperature is lower than 15 ℃ , the compressor stops to cooling, at the same time the auxiliary display shows the temperature upper limit as 25℃.
  • Page 13: Start/Stop Pump

    Recirculating cooler Press Start/Stop key under working interface. The pump operation indicator is on. The auxiliary display shows the compressor protection countdown. After countdown finished, the compressor begin to run, and the auxiliary display shows the target temperature., and the cooling indicator will indicate cooling state according to the computing result of the PID modifier.
  • Page 14: Switching Off

    Recirculating cooler adjusting temperature. Ordinarily it is about 15 to 20 minutes. 3.3.6 Switching off 1. Press Start/Stop key to stop the compressor, the system enters into standby state, then press Pump key to turn off then pump. 2. Switch off the power supply of the control panel; 3.
  • Page 15 Recirculating cooler I for details). 1) Input code 0022 to enter into operation model base. The first interface: the auxiliary display shows R_AM, which is the code for automatic start with power on; the main display shows setting parameter, 0 stands for automatic start, 1 for manual start, 2 for remote control switch, 3 for remote communication control.
  • Page 16: Common Alarms

    Recirculating cooler the keyboard locks; long press Esc to quit lock state; Choose NO to disable the function. 7) The seventh interface: the auxiliary display shows R_RS, the main display shows NO or YES, this parameter can restore factory default setting; the set method is ditto.
  • Page 17 Recirculating cooler shows E_LW. Remark: To protect cooling unit, please switch off it before check. Low water flow alarm: E_LF ■ In working state, when the water flow amount is lower than the valve (not all the units available of flow valve, it is optional for customers), the water flow alarm functions with buzzer beeps and the auxiliary display shows E-LF.
  • Page 18: Hardware Port (Optional)

    Recirculating cooler Remark 1: To protect cooling unit, please switch off it before check. Remark 2: The compressor can not be started when the temperature of the circulation liquid is higher than the temperature of compressor protection. Under temperature limit alarm: E_L1、E_L2 ■...
  • Page 19: Chapter Iv Trouble Shooting Guide

    Recirculating cooler The definition of the 12 pins of the aviation plug: Pin 1 & 2: water flow signal output, which function with normal water flow and disconnect with abnormal water flow. Pin 3 & 4: water level signal output, which function with normal water level and disconnect with abnormal water level.
  • Page 20: Appendix I Parameter Database

    Recirculating cooler screwdriver, then the pump can be restarted properly. Attention: Different unit may apply for different pump type, the method above may be not always applicable.  Phenomenon: water level alarm indicator flashes and buzzer beeps. Possible reason and solution: a) Reason: water shortage in the water tank, low water level.
  • Page 21 Recirculating cooler Initial Item Parameter Meaning Option or Range data 0 - automatic start 1-manual start Automatic start with 2-remote switch Control mode R_AM power on control remote - communication control Temperature control 1 - Start/Stop Temperature mode option type R_CM control mode 2-PID...
  • Page 22 Recirculating cooler Option Initial Item Parameter Meaning Range Data C_PL+1 High temperature The first point of the user defined ~ E_H1 C_PH alarm 1 high temperature alarm C_PH E_H1+1 High temperature The second point of the user ~ E_H2 C_PH alarm 2 defined high temperature alarm C_PH...
  • Page 23 Recirculating cooler Option Initial Item Parameter Meaning Range Data 7 - E_ S3 alarm 8 - E_CE alarm 9 - E_ PL alarm 10 - E_PH alarm 11 - E_CH alarm 0-No alarm E_H1 - alarm E_H2 - alarm E_L1 -...
  • Page 24: Appendix Ii Warning Information

    Recirculating cooler Option Initial Item Parameter Meaning Range Data 11 - E_CH alarm _ _ _ _END Parameter set end signal _ _ _ _ Appendix II Warning Information Alarm lamp Auxiliary Main Item Solution and buzzer display display quit when High temperature Current...
  • Page 25 Recirculating cooler Alarm lamp Auxiliary Main Item Solution and buzzer display display normal water level and back to original work mode Stop heating/cooling, quit at Sensor 1 short 1S flash E_S1 -999 normal water level and back to original work mode Stop heating/cooling, quit at Sensor 2 break 1S flash...