Declaration – electromagnetic immunity
Guidance and manufacturer's declaration – electromagnetic immunity.
The device is intended for use in the electromagnetic environment specified below.
The customer or the user of the device should assure that it is used in such an environment.
IMMUNITY test
Electrostatic
discharge (ESD) IEC
61000-4-2
72
IEC 60601 test
level
± 6 kV contact
± 8 kV air
Compliance
level
± 6 kV contact
± 8 kV air
Electromagnetic
environment – guidance
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should be
at least 30 %.