5. Press the ENTER
button. The display
shows FFFF to indicate factory settings have been restored
6. Press the ENTER button to return to the main screen

Installation Note

For best results, we recommend installing this equipment in an indoor area free from wind. A
cold wind from outside might influence the temperature readings. We also recommend that the
temperature measurement area is protected from bright sunlight to avoid false high results.

Temperature Detection Operational Notes

1. Normal ambient temperature is 32°F - 86°F [0°C - 30°C]. When using it at 14°F [-10°C],
you may have to increase the environmental temperature compensation value
a. Increase the value 5°C at a time, until the scanned result difference between it
and the forehead thermometer is within plus or minus 0.3°C
b. You can also use this method to calibrate the sensor if the scanned results of
many persons are low
2. Measurement method: Measure the forehead or the wrist surface
3. The standard measurement distance is 0.4-4.0 IN [1cm-10cm]
4. The measurement time is 0.5 seconds. After the measurement, advise people to leave
quickly. Otherwise the measurement may be repeated
PORTABLE THERMAL DETECTOR