Olympus AFU-100 Instructions For Use Manual

Olympus AFU-100 Instructions For Use Manual

Peristaltic pump unit

Advertisement

INSTRUCTIONS FOR USE
PERISTALTIC PUMP UNIT
AFU-100
REF:
WB950167 (100...240 V~)
WB135745-W10

Advertisement

Table of Contents
loading

Summary of Contents for Olympus AFU-100

  • Page 1 INSTRUCTIONS FOR USE PERISTALTIC PUMP UNIT AFU-100 REF: WB950167 (100…240 V~) WB135745-W10...
  • Page 3: Table Of Contents

    Olympus AFU-100 Instructions for Use Contents Important information Intended use Instructions for use User qualification Peristaltic pump unit compatibility Repair and modification Signal words Danger, warnings and cautions General description Inspection prior to use Symbols, definitions and functions Symbols and descriptions 4.1.1 Front panel...
  • Page 4 Instructions for Use Olympus AFU-100 Operation Power supply Acoustic signals and tone volume adjustments Pump tube insertion and connection Flow selection Store and recall flow levels Activation / deactivation of the flow Procedure after use Cleaning, storage and disposal Cleaning...
  • Page 5: Important Information

    Olympus AFU-100 Instructions for Use Important information Intended use The AFU-100 is a peristaltic pump unit intended for irrigation of instruments or irrigation / flushing / cleansing of tissue surfaces and wounds supporting endoscopic diagnosis or therapy. Contraindications: The peristaltic pump unit must not be applied for intravenous fluid administration.
  • Page 6: Signal Words

    Instructions for Use Olympus AFU-100 Signal words The following signal words are used throughout these instructions for use: DANGER Indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury. WARNING Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
  • Page 7 To prevent user shock, malfunction and damage of the peristaltic pump unit, keep liquids away from all electrical equipment. If liquid gets on or into the peristaltic pump unit, terminate operation immediately and contact Olympus or your distributor. CAUTION Injury during servicing When the housing is opened, there is a danger of electric shock.
  • Page 8 Instructions for Use Olympus AFU-100 Fire / Explosion DANGER Explosion This peristaltic pump unit is not explosion-proof. Do not use the unit within an explosion zone. WARNING Risk of fire Disconnect the power plug before changing the fuses! Replace fuses as marked. The fuses must only be replaced by authorised technicians.
  • Page 9: General Description

    Instructions for Use General description The AFU-100 peristaltic pump unit is a roller pump. The rollers of the pump rotors squeeze the inserted pump tube so that the liquid in the pump tube is transported. This enables liquid media to be transported without coming into contact with the peristaltic pump unit.
  • Page 10: Symbols, Definitions And Functions

    Instructions for Use Olympus AFU-100 Symbols, definitions and functions Symbols and descriptions 4.1.1 Front panel Power on / off Caution Defibrillation proof type CF applied part (cardiac application) Up (increase flow level) Down (decrease flow level) Flow on / off...
  • Page 11: Rear Panel

    Olympus AFU-100 Instructions for Use 4.1.2 Rear panel Manufacturer Date of manufacture Serial number Reference number Indication of the direction of higher volume Loudspeaker Footswitch connection Interface for electrosurgical unit connection Operating instructions Waste electrical and electronic equipment Fuse Product carries the CE mark...
  • Page 12: Packaging

    Instructions for Use Olympus AFU-100 4.1.3 Packaging Keep away from rain Fragile; handle with care This way up Temperature limit during transport storage Stacking limit by number (including lowest box) Atmospheric pressure limitation during transport and storage Relative humidity limitation during transport...
  • Page 13: Front Panel

    Olympus AFU-100 Instructions for Use Front panel Level display (14) Up / down button (5 / 6) Error indicator (7) Power switch (0) Memory 3 button (3) Memory 1 indicator (8) Flow activation / deactivation Electrosurgical unit assignment indicator (12)
  • Page 14: Buttons / Switches

    Instructions for Use Olympus AFU-100 4.2.1 Buttons / Switches Power switch This switch turns on and off the peristaltic pump unit. 1 / 2 / 3 Memory 1 / 2 / 3 button The memory buttons are used to store and recall previously stored flow levels.
  • Page 15: Rear Panel

    Olympus AFU-100 Instructions for Use Rear panel Volume control (20) Type plate (19) Equipotential bonding connector (17) Communication port electrosurgical unit (16) Footswitch connector (15) Mains connector with integrated fuses (18) Footswitch connector This terminal connects the footswitch with the peristaltic pump unit, so that the flow can be activated / deactivated by the footswitch pedal (21).
  • Page 16: Accessories

    7.4, Flow selection). Installation and commissioning The unit must be properly installed and commissioned by Olympus or your distributor or by a person or firm commissioned and authorised by the manufacturer. Prepare the peristaltic pump unit and other equipment to be used with this peristaltic pump unit before each use.
  • Page 17: Installation Of Equipment

    Olympus AFU-100 Instructions for Use Installation of equipment CAUTION Install the unit on a stable, level surface. Otherwise, the unit could fall, causing equipment damage and / or injury to the user or patient. If the peristaltic pump unit is placed on a cart, the cart must be of adequate strength and size to hold the unit securely.
  • Page 18: Connection To An Ac Mains Power Supply

    Instructions for Use Olympus AFU-100 Connection to an AC mains power supply Confirm that the power of the peristaltic pump unit is off. Connect the power cord to the mains connector (18) of the peristaltic pump unit (see fig. 6.2).
  • Page 19: Equipotential Bonding

    If the voltage of the facility is different from the voltage indicated on the type plate of the peristaltic pump unit, contact Olympus or your distributor. Equipotential bonding To increase electrical safety, the unit can be connected to the equipotential bonding system of the treatment room.
  • Page 20: Connection To An Electrosurgical Unit

    Instructions for Use Olympus AFU-100 NOTE If the peristaltic pump unit is connected via the communication cable with an electrosurgical unit and communication is established, the footswitch of the peristaltic pump unit cannot be used. In this case, the peristaltic pump unit will be controlled by the electrosurgical unit’s footswitch.
  • Page 21: Power Supply

    9 (Troubleshooting). If the irregularity is still suspected after consulting chapter 9, contact Olympus or your distributor. Damage or irregularity may compromise patient or user safety and may result in more severe equipment damage. Power supply Switch on the peristaltic pump unit with the power switch (0). The unit executes a self test.
  • Page 22: Pump Tube Insertion And Connection

    Instructions for Use Olympus AFU-100 Pump tube insertion and connection The tubing sets are supplied in individual, sterile packaging. Check the sterile tubing packaging for damage and remove the tubing from the packaging. Lift the protective cover of the pump head.
  • Page 23: Store And Recall Flow Levels

    Olympus AFU-100 Instructions for Use For changing the flow level by e.g. 3 (from flow level 3 to 6), press the up / down button (5 / 6) for 3 times. For a significant change of the flow level continuously press the up / down button (5 / 6).
  • Page 24: Procedure After Use

    Instructions for Use Olympus AFU-100 DANGER If inadequate flow is observed or the flow does not stop, end the procedure immediately. Use the power switch (0) as emergency stop for malfunctions (e.g. in case the flow activation / deactivation button (4) or the footswitch does not react).
  • Page 25: Cleaning, Storage And Disposal

    Olympus AFU-100 Instructions for Use Cleaning, storage and disposal After each use, perform the following cleaning procedures immediately. If cleaning is delayed, debris encrustation may become a source of infection. Encrustation may also result in malfunction of the peristaltic pump unit. For maintenance and storage of other items than those described below, refer to the respective instructions for use.
  • Page 26: Disposal Of The Unit

    Troubleshooting If the peristaltic pump unit has visible damage, do not use the unit and contact Olympus or your distributor. If the unit is not functioning properly, use the information in this chapter to identify and correct the malfunction. If the problem cannot be resolved by the described remedial action, stop using the peristaltic pump unit and contact Olympus or your distributor for repair.
  • Page 27: Error Codes And Measures

    Periodic safety checks The peristaltic pump unit and the footswitch must undergo a safety check in yearly intervals in accordance with the national statutory regulations. For this purpose, Olympus or a firm authorised by Olympus must be commissioned by the user. WB135745-W10...
  • Page 28: Returning The Peristaltic Pump Unit For Repair

    With the peristaltic pump unit, include a description of the malfunction or damage and the name and the telephone number of a contact person. Repairs must only be carried out by Olympus or a firm authorised by Olympus. Service documents such as circuit diagrams, parts lists, equipment descriptions and setting instructions are available from Olympus for technicians who are authorised to carry out maintenance and repair.
  • Page 29: Specifications

    CAUTION The achievable flow rate also depends on the used instrument and its accessories. 11.2 Technical data Unit designation / description AFU-100 peristaltic pump unit REF: WB950167 (100…240 V~) Compatible electrosurgical units ESG-100 REF: WB991036 (220…240 V~) REF: WB991046 (100…120 V~)
  • Page 30 Instructions for Use Olympus AFU-100 General data Protection class according to IEC 60601-1 CF, Class I Classification according to MDD 93/42/EEC Voltage range 100…240 V~ Frequency 50 / 60 Hz Maximum input power 60 VA Terminal for potential equalization Power fuse...
  • Page 31 Olympus AFU-100 Instructions for Use Accessories Compatible pump tubes Ø 8.0 mm, Ø 4.8 mm REF: WB920138 outer inner Ø 6.4 mm, Ø 3.2 mm REF: WB920137 outer inner Environmental conditions (pump tube) Normal operation Temperature + 10…+ 40° C Relative humidity 15…80 %, non-condensing...
  • Page 32: Electromagnetic Compatibility (Emc)

    Instructions for Use Olympus AFU-100 12 Electromagnetic compatibility (EMC) 12.1 Guidance and manufacturer’s declaration – electromagnetic emissions The peristaltic pump unit is intended for use in the environment specified bellow. The user of the peristaltic pump unit should assure that it is used in such an environment.
  • Page 33: Guidance And Manufacturer's Declaration - Electromagnetic Immunity

    Olympus AFU-100 Instructions for Use 12.2 Guidance and manufacturer’s declaration – electromagnetic immunity The peristaltic pump unit is intended for use in the environment specified bellow. The user of the peristaltic pump unit should assure its use in such an environment.
  • Page 34 Instructions for Use Olympus AFU-100 Immunity test IEC 60601-1-2 Compliance Electromagnetic test level level environment – guidance Portable mobile high frequency communication equipment should be used not closer to any part of the peristaltic pump unit, including cables, then the recommended...
  • Page 35: Recommended Separation Distances Between Portable And Mobile High Frequency Communication Equipment And The Peristaltic Pump Unit

    Olympus AFU-100 Instructions for Use The field strength from fixed transmitters, such as base stations from radio (cellular / cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be theoretical predicted with exact accuracy.
  • Page 36 OLYMPUS AUSTRALIA PTY. LTD. 31 Gilby Road, Mount Waverly VIC 3149, Australia Fax: +61 3 9543 1350, Phone: +61 3 9265 5400 OLYMPUS (BEIJING) SALES & SERVICE, CO., LTD. OLYMPUS (BEIJING) SALES & SERVICE, CO., LTD. OLYMPUS (BEIJING) SALES & SERVICE, CO., LTD.