Tanks 2 and 3 are both ultrasonic rinse tanks. Fresh rinse water can be introduced into the system with the excess cascading to drain. The E782 unit consists of two components: the main cleaning unit containing the 316 Stainless Steel tanks housed in a 304 stainless cabinet and a stainless power module cabinet containing the bulk of the electronics.
Page 4
cooling by the fan can be restricted because of tight enclosures. Clearance of 1” is necessary both at the air intake and exhaust. PLUMBING A number of valves and connectors have been supplied with unit. With the PVC and plastic fittings and valves, add Teflon tape to the threads and HAND tighten.
Page 5
1. The initial start up works best if the pump filter system is on the floor or lower than the tank. 2. Add liquid to the tank and allow liquid to flow into filter chamber. Make sure ball valve on the pump filter is open. 3.
Page 6
been adjusted to control the bath temperature at 160degrees F when tank covers are on. If temperature needs to be changed, there is a port at the rear of each tank module where a screwdriver can be used to adjust the control knob of thermostat. The thermostat is the same as used in the drying chamber mentioned below.
Page 7
Modular Circuit Board Unit E782 is equipped with three modular circuit boards that are easily replaced if a problem occurs. Three small indicator lights on the power module indicate that circuit boards are working. If the indicator light goes out, a replacement circuit board can be shipped immediately.