Chapter 4: Using The Instrument; Calibrating The Instrument - Barbieri Spectro LFP S3 Operating Manual

Automatic reflection and transmission spectrophotometer
Table of Contents

Advertisement

Spectro LFP RT - Operating Manual

Chapter 4: Using the instrument

4.1 Calibrating the instrument

The instrument is capable of self-calibration.
Reflection
Reflection calibration is done by means of its internal white calibration
standard placed under the measuring head.
Note: The validity of this reference white is 24 months from the pro-
duction date. See the serial no.on the reference white . The first four
characters indicate the year and the month of production. For example a
calibration strip with a serial number A601021indicates: Year: A6= 2006,
Month: 01= January.This reference white is valid until January, 2008.
The instrument automatically calibrates itself before starting measure-
ments.