CALIBRATION

This scale has been factory calibrated, and does not require calibration prior to
use. If required, the scale can be calibrated. Calibration of your scale is performed
using kilograms (KG) or pounds (LB), according to the units of measure used upon
entering into the calibration path. When calibrating in pounds, a total of 300 lbs is
required, applied in increments of 100 lb, 200 lb and 300 lb. If calibrating in kilograms, a
total of 120 kg is required, applied in increments of 40 kg, 80 kg and 120 kg.
THE 499KG CAN ONLY BE CALIBRATED IN KILOGRAMS.
Note: Only weights certified and traceable to national standards should be used
for calibration procedures.
With the scale off, press and hold the LB/KG button (499KL) or the
blank blue button (499KG) and ZERO buttons at the same time, then
press the ON/OFF button to turn on the scale. Release all three
buttons when "CAL" appears on the display. Numbers will appear
indicating that you can begin calibration.
When calibration numbers stabilize, press the ZERO button. "C-100"
(for LB mode) or "C-40" (for KG mode) will flash on the display,
followed by "0"or "1".
Place 100 lbs or 40 kg on the scale. Numbers will appear on the
display. When calibration numbers stabilize, press the ZERO button.
"C-200" (for LB mode) or "C-80" (for KG mode) will appear on the
display, followed by "0" or "1".
Place an additional 100 lbs or 40 kg on the scale, for a total of 200 lbs
or 80 kg. Numbers will appear on the display. When calibration
numbers stabilize, press the ZERO button. "C-300" (for LB mode) or
"C-120" (for KG mode) will appear on the display, followed by "0" or
"1".
Place an additional 100 lbs or 40 kg on the scale, for a total of 300 lbs
or 120 kg. Numbers will appear on the display. When calibration
numbers stabilize, press the ZERO button. "END" will appear on the
display, followed by "OFF". The scale will automatically shut off.
P/N UM499KL 499KG Rev 20190430_Web
ACTION
19
DISPLAY
SHOWS
"CAL" followed by
numbers
Numbers followed
by "C-100" (LB
mode) or
"C-40" (KG mode)
then "0" or "1".
Numbers followed
by "C-200" (LB
mode) or
"C-80" (KG
mode) then "0" or
"1".
Numbers followed
by "C-300" (LB
mode) or
"C-120" (KG
mode) then "0"
or "1".
Numbers followed
by "END" then
"OFF"