BioShift Single Door UV-C Germicidal
Chamber - Instruction Manual
This manual provides user guidelines and additional operation
instructions for the Small BioShift Single Door UV-C Germicidal Chamber.
(Model 24-0256)
For service or parts call your
sales representative, visit our website:
www.once.lighting/uv-c
Or contact us at:
Once Inc.
United States
Germany
Poland
15255 23rd Ave N
Am Südfeld 7
Wloclawska 167
Plymouth, MN 55447
49377 Vechta
87-100 Tourn
+1 (763) 381-5621
+49 (0) 4441 8898-10
+48 (56) 699-5330
info@onceteam.com
info.germany@onceteam.com
info.poland@onceteam.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the BioShift Single Door UV-C Germicidal Chamber and is the answer not in the manual?

Questions and answers

Summary of Contents for Once BioShift Single Door UV-C Germicidal Chamber

  • Page 1 BioShift Single Door UV-C Germicidal Chamber - Instruction Manual This manual provides user guidelines and additional operation instructions for the Small BioShift Single Door UV-C Germicidal Chamber. (Model 24-0256) For service or parts call your sales representative, visit our website: www.once.lighting/uv-c...
  • Page 2: Table Of Contents

    TABLE OF CONTENTS Introduction ..............................2 Specifications .............................. 3 Safety ................................. 4-5 Installation ..............................6 Touchscreen Display ..........................7-9 Home Screen Components ....................7 Maintenance Screen Components .................. 8 Settings Screen Components .................... 9 Programming ............................10 Cycle Time ..........................10 Sleep Timer ..........................
  • Page 3: Introduction

    UV-C radiation. Thus, the BioShift Single Door UV-C Germicidal Chamber is most effectively used in addition and as a complement to disinfection products and procedures, and NOT as a replacement.
  • Page 4: Specifications

    SPECIFICATIONS Dimensions Outside 29.5” wide x 23” deep x 23.6” high (750 x 584 x 600 mm) Weight 110 lbs (50 kg) Functional Operating Power / Current 110V: 110W / 1 A, 240V: 115W / 550 mA Standby Power / Current 110V: 6W / 130 mA, 240V: 9W / 100mA Input Voltage 110-240VAC, 50/60 Hz (voltage fluctuation ±10%)
  • Page 5: Safety

    The lamps are sensitive to the oils in human skin, always handle with gloves. Disposal and Recycling All components must be disposed of in accordance with state and local laws. Refer to lamprecycle.org for more information. Visit www.once.lighting/bioshift-warnings or scan the QR code for additional warning information.
  • Page 6 SAFETY Features Emergency Stop The safety switch can be used at any time to stop the operation. Pressing EMERGENCY STOP disconnects all power. Open Door BioShift chamber will not operate when the door is open. The door cannot be opened during cycle time unless EMERGENCY STOP is pressed.
  • Page 7: Installation

    2000m. The input power switch is customizable to the front or rear side of the chamber. If this is desired, contact Once for further instructions. Use two people to carefully lift the BioShift chamber into place.
  • Page 8: Touchscreen Display

    TOUCHSCREEN DISPLAY The touchscreen display is located on the front of the unit. Home Screen Components (1) Cycle Time Display (2) Chamber Start / Stop (3) Chamber Status (4) Settings (5) Maintenance...
  • Page 9: Maintenance Screen Components

    TOUCHSCREEN DISPLAY Located under the Maintenance button. Maintenance Screen Components (1) View Lamp Life Remaining (2) Reset Lamp Life (3) View Lamp Replacement Info (4) Run 100mJ/cm Dose Test (5) Run 250mJ/cm Dose Test (6) Run 1000mJ/cm Dose Test...
  • Page 10: Settings Screen Components

    TOUCHSCREEN DISPLAY Located under the Home screen Settings button. Settings Screen Components (1) View Set Lamp Cycle (2) Set Lamp Cycle (3) Set Sleep Time (4) Expired Lamp Life (5) Contact ONCE ®...
  • Page 11: Programming

    Choose the Set Cycle button. Note: The time is set by arrowing up or down, programmed in MM:SS. • Once desired lamp life cycle has been chosen, select the check mark to set. PROGRAMMING ITEMS TO NOTE: • Controller cannot be programmed if EMERGENCY STOP button is pressed in.
  • Page 12: Maintenance

    Replacement lamps have been installed and Lamp Life needs to be reset. Simply follow the red exclamation point to the source of the issue. Note: Lamp life remaining will never go negative. Once it reaches zero, it will remain until Lamp Life is reset.
  • Page 13: Check Lamp Life

    Maintenance menu (see Maintenance Screen Components). Order Replacement Lamps and Ballasts The information required for ordering new lamps include the lamp part number and ONCE contact information. This information can be found from the Maintenance screen ® under Lamp Replacement Info.
  • Page 14: Cleaning

    It is important to regularly verify the UV-C dose output of the BioShift chamber to ensure proper disinfection. Every 50 cycles or once per month, whichever comes first, the operator shall verify that the UV-C dose exceeds a minimum of 100 mJ/cm in a short, programmed test cycle.
  • Page 15: Operating

    OPERATING Ensure power switch is turned on and Emergency Stop button is released. Place objects in chamber following guidelines below. • Do not stack items / supplies in the chamber. • Leave 6” between each product in the chamber with no more than 3 across. •...
  • Page 16: Contact Information

    Manual valid for all 24-0256 models Model number can be found on product nameplate For more information and references visit us online at www.once.lighting/uv-c For warranty information visit www.once.lighting/warranties For service or parts call your sales representative, visit our website: www.once.lighting/uv-c...