Input High Level Voltage; Notes For Selecting Sensor Cable; Notes For Dynamic Measurement - dacell AM-1000 User Manual

Dynamic strain amplifier
Table of Contents

Advertisement

4-3. High Level Voltage input
This device can measure the signal that inputted in DC (DCDT, High level,
Thermocouple etc).
For the measurement, following conditions should be satisfied.
On Common mode, the amp input should be as ± 10V. Please be careful since
1
when over ± 10V is inputted, the signal source will be cut off.
When Thermocouple temperature sensor is used, an error in measurement can
2
be occurred since the temperature-electromotive force might be generated on
the commissure between the connector pin of amplifier and thermocouple lead
wire. Therefore, you should do the cold junction compensation or you should
carry out the measurement in the place where without such a big temperature
change.
The input stage of amplifier are "+SIG", "-SIG". Since their minimum input
3
current is originated by this input impedance, it should be over 50mA to carry out
the measurement.

4-4. Notes for selecting sensor cable

The each space for selecting sensor cable should be:
electrically isolated.
1
Cable resistance should be small.
2
The cable resistance of the internal lines should be regular.
3
It is better to use the shield cable.
4

4-5. Notes for Dynamic measurement

Strain gauge or Transducer (located at the end of the measurement lead cable) is
exposed to various electric devices and equipment and also receive the analogue
frequency (50kHz, 60kHz) noise from the line power. Therefore, please carry out
the following procedures to minimize this kind of happening.
Please prepare for the shield cable. (If the shield cable cannot be ready,
1
please make the lead cable as shortest as possible.)
13

Advertisement

Table of Contents
loading

Table of Contents