Figure 31

Cleaning

To reduce the risk of electrical shock:
◼ Do not disassemble Sofia 2. Sofia 2 contains no user-serviceable components.
◼ Possible electrical shock: Turn off and unplug Sofia 2 before cleaning. Do not clean
the ports on the back of the instrument.
◼ Do not immerse or spray Sofia 2 in liquid.
◼ Do not clean the underside of the instrument.
Turn off and unplug Sofia 2 before cleaning. Use a soft cloth with 70% alcohol or 0.6%
bleach solution to clean the exterior of Sofia 2. Do not use soap or other solutions except
those noted below. Wipe the external surfaces and cassette drawer only.
Note: The following disinfectants identified in EPA's list of disinfectants for use against
Ebola virus have been found by Quidel to be compatible for use with Sofia 2:
• Clorox Healthcare; Bleach Germicidal Cleaner (EPA Reg #: 56392-7)
• Clorox Commercial Solutions; Clean-Up Disinfectant Cleaner with Bleach 1 (EPA
Reg #: 67619-17)
• Clorox Healthcare; Bleach Germicidal Wipes (EPA Reg #: 67619-12)
Removing the Drawer for Cleaning
Ensure that there is no cassette in the drawer and unplug Sofia 2. Gently pull the drawer out
until it stops, then lift up Sofia 2 and press the button as shown (Figure 32). The drawer can
then be removed.
Note: If resistance is met, do not force. Apply slightly more pressure to release the catch.
Page 42 of 48