Summary of Contents for Cosmo Bio My Run CBJ-IMR-201-EX

  • Page 1 Power supply Instruction Manual CBJ-IMR-201-EX TOYO EKIMAE BLDG. 2-20, TOYO 2-CHOME, KOTO-KU. TOKYO 135-0016, JAPAN TEL : (81)3-5632-9617 FAX : (81)3-5632-9618 e-mail : export@cosmobio.co.jp URL : www.cosmobio.com...
  • Page 2: Important User Information

    This product complies with the requirements terms and conditions of sale of the company of Low Voltage within the Cosmo Bio Co., Ltd., group that Directive (LVD) 73/23/EEC through the supplies them. A copy of these terms and harmonized standard EN 61010-1, 1993 + A1, conditions is available on request.
  • Page 3: Safety Information

    Any electrophoresis equipment connected to the power supply should meet the requirements of the IEC 1010-1:1993 electrical standard. 7. _Note that the output is connected to the chassis/reference earth. 8. _Please contact Cosmo Bio if a problem occurs. (Never replace the fuse yourself.) Never open the cover.
  • Page 4: Built-In Safety Features

    2.2 Built-in safety features The CBJ-IMR-201-EX has been tested and complies with the IEC 61010-1 (EN 61010-1) electrical safety standard. 1. _Functional earth leakage. Should the power supply be connected to an electrophoresis unit that has a leakage path to earth, the CBJ-IMR-201-EX will detect this fault and the high voltage will be turned off.
  • Page 5: Power Supply Operation

    Connect the leads from the electrophoresis unit (red to red and black to black) The red lead is the positive, and the black is the negative. Do not touch the red and black lead points. WARNING! Lead Points Use only undamaged electrical wire and equipment approved for the voltage you will use.
  • Page 6 B. Turn on the power on the back panel of the power supply. Power switch Power code If you want to run your gel at constant voltage: 1. _Select the voltage by using the key of the remote control. 2. _Adjust the voltage with the up and down key.
  • Page 7: Installation Condition

    You want to stop the electrophoresis: Select and hold the key for two seconds, OFF appears on the display. To start, select the key. If the timer is selected, it will automatically restart at the value before the electrophoresis was stopped.
  • Page 8: Maintenance

    Wipe the instrument regularly with a damp cloth. Let the instrument dry completely before use. Otherwise, no user maintenance is necessary. All servicing should be entrusted to qualified personnel only. Please contact Cosmo Bio Co., Ltd., for more service information. 8. Environmental Condition 1. Indoor use.