Rating of laboratory
thermostats and
laboratory baths
18
Minichiller®
▪ Only certified companies may maintain, repair or decommission refrigeration systems. The re-
sponsible bodies must verify that these companies are certified.
▪ Regular leak tests of stationary refrigeration systems by certified personnel (such as Huber service
engineers). The required test interval is based on the refrigerant filling capacity and the type of
refrigerant, converted to CO
▪ Responsibility of responsible bodies operating a plant to recover F-gases by certified personnel.
▪ Obligatory documentation requirement in the refrigeration system's operation manual, specifying
type and volume of refrigerant used or recovered. The responsible body must keep the records
for at least 5 years after their creation and present it to the responsible authority upon request.
▪ Temperature control units with natural refrigerants (NR) are exempt from this Directive.
▪ Please refer to the data sheet or rating plate of your temperature control unit for the quantity
and type of refrigerant.
▪ Additional information about the definition of the inspection interval can be found on our web-
site.

Requirements for operators

Work on the temperature control unit is reserved for appropriately qualified specialists, who have
been assigned and trained by the responsible body to do so. Operators must be at least 18 years old.
Under 18-year olds may operate the temperature control unit only under the supervision of a quali-
fied specialist. The operator is responsible vis-a-vis third-parties in the work area.

Obligations of the operators

Carefully read the operation manual before operating the temperature control unit. Please observe
the safety instructions. When operating the temperature control unit, wear appropriate personal
protective equipment (e.g. safety goggles, protective gloves, non-slip shoes).

General information

Description of workstation

The workstation is located at the control panel in front of the temperature control unit. The work-
station is determined by the customer's connected peripheries. Accordingly, it must be designed
safe by the responsible body. The workstation design also depends on the applicable requirements
of the German occupational health and safety regulations [BetrSichV] and the risk analysis for the
workstation.

Safety devices to DIN 12876

The rating of your temperature control unit is stated on the data sheet in the appendix.
Temperature
Classification
control medium
I
Non-combustible
II
Combustible
III
Combustible
a)
Usually water; other fluids only if non-combustible even within the temperature range of an individual fault.
b)
The temperature control media must have a fire point of ≥ 65 °C.
c)
The overheat protection can, for instance, be realized using a suitable fill level sensor or a suitable temperature
limiter.
Optional at the choice of the manufacturer.
d)
▪ Temperature control units with heating correspond to class number III/FL. These temperature
control units are characterized by an "H" in the device name.
▪ Temperature control units without heating correspond to class number I/NFL.
2 equivalent.
2
Technical requirements
a)
Overheat protection
b)
Adjustable overheat protection
Adjustable overtemperature protection
b)
and additional low-level protection
Introduction
Chapter 1
Identification
c)
NFL
FL
FL