Before using, read this manual carefully, accessory directions for use, all precautionary information in boldface type, and specifications. Warning The HHP-201 is a prescription device and is to be operated by qualified personnel only. Explosion hazard. Do not use the HHP-201 handheld pulse oximeter in the presence of flammable anesthetic.
Safety Caution The HHP-201 is intended only as an adjunct in patient assessment. It must be used in conjunction with clinical signs and symptoms. Ambient light, movement, electromagnetic interference, artifacts, dysfunctional hemoglobin, and certain dyes, etc. may be interfering in the pulse oximeter's functions.
The Pulse Oximeter contains: mainframe and probe. Caution The HHP-201 is intended only as an adjunct in patient assessment. It must be used in conjunction with clinical signs and symptoms. Warning Do not use the HHP-201 during magnetic resonance imaging (MRI) scanning.
Note Pulse Indicator is a visual indication of the patient’s pulse straight, not that display is proportional to the pulse volume. The display update period of the HHP-201 in various operating conditions is 1/60 second. Description of audible indicators The following are audible indicators for which there are no accompanying symbols, keys or visual indicators....
3 Unpackingg and Installation 3.1 Unpacking Carefully remove the HHP-201 and its accessories form the shipping carton. Save the packing materials in case the HHP-201 must be shipped or stored. Inspect the equipment for mechanical damage. If yes, please contact your provider promptly.
Do not mix alkaline “AA” batteries with rechargeable batteries. When enplacing batteries, replace with four fresh (new) batteries. Do not mix used and new batteries. If don’t use HHP-201 for a period time, please take all batteries away. When charging the battery with power adapter, press for 1s, the charge indicator starts blinking.
Cautions The HHP-201 is intended only as an adjunct in patient assessment. It must be used in conjunction with clinical signs and symptoms. If the battery were insufficient to operate, the power indicator would turn red and blink for three times.
Pulse Oximeter, contact your provider or manufacturer During the Power-on, the HHP-201 automatically conducts a power-on self-test .If the HHP-201 detects an internal problem, an error tone sounds and the monitor displays an Error Code and corresponding number (see Troubleshooting section for error codes).
4.1.4 Turn Off The Pulse Oximeter turns itself off automatically after pressing keys for 2 seconds. The HHP-201 turns itself off automatically when the batteries are weak. 4.1.5 Charge Specification There is a USB socket on the pulse oximeter top left corner, that’s for charger plug.
Do not have on carrying pouch during charging 4.1.6 Storage Remove the batteries from the HHP-201 before long-term storage, or if the device won't be used for 6 months or more. This protects the device from damage due to batteries leaking acid.
The cover should be removed only by qualified service personnel. There are no user-serviceable parts inside. If you experience a problem while using the HHP-201 and are unable to correct it, contact qualified service personnel or representative. The HHP-201 service manual, which is for use by qualified service personnel, provides additional troubleshooting information.
Returned Goods Authorization number. Remove the batteries for shipping, and unplug the probe. Pack the HHP-201 in its original shipping carton .If the original carton is not available; use a suitable carton with appropriate packing material to protect it during shipping.
Technology,In 8 Guarantee The company warrants HHP-201 handheld Pulse oximeter at the time of its original purchase and for the subsequent time period of twelve months for the original purchaser. The company warrants SpO2 Probe free of defects at the time of its original purchase and for the subsequent time period of three months.
For all EQUIPMENT and SYSTEMS Guidance and manufacturer’s declaration-electromagnetic emission The HHP-201 Handheld Pulse Oximeter is intended for use in the electromagnetic environment specified below. The customer of the user of the HHP-201 Handheld Pulse Oximeter should assure that it is used in such environment.
Page 27
For EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING Guidance and manufacturer’s declaration-electromagnetic immunity The HHP-201 Handheld Pulse Oximeter is intended for use in the electromagnetic environment specified below. The customer or the user of HHP-201 Handheld Pulse Oximeter should assure that it is used in such an environment.
Page 29
RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the HHP-201 Handheld Pulse Oximeter is used exceeds the applicable RF compliance level above, the HHP-201 Handheld Pulse Oximeter should be observed to verify normal operation.
Page 30
The HHP-201 generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with these instructions, may cause harmful interference with other devices in the vicinity.