Maintaining the Validation Kit

Due to the nature of variable area flow meters, as used in the
validation kit and the product being manufactured to precise
standards, annual calibration is not required. Regular inspection is
needed.
It is recommended that the validation kit is returned to
Cantium Scientific Limited or an appointed re-seller for full factory
inspection every five years.
When not in use, the kit should always be stored in the padded
carry case and kept in a clean dry environment. The kit should be
cleaned with a soft dry cloth.
following checks:
1. Ensure the orange coloured seals in the head adaptor (2) are
clean, secure and level.
2. The calibrated flow tube (1) fits securely into the adaptor
head (2).
3. Looking into the bottom of the glass tube check the white
PTFE stopper is in-line with the direction of the tube and not
at an angle restricting flow. If not, this can be moved into
position using small pliers or tweezers.
4. The glass tube is parallel with the longer white plastic tube.
If not, this can be moved into position by hand, as the
silicone rubber fitting does allow a small degree of
movement.
5. The glass tube is firmly fitted into the silicone rubber fitting.
If required, the kit can be returned to Cantium Scientific Limited or
your local distributor for independent verification.
Inspection should include the