Setting Up

Tube and head adaptor assembly

The first step is to insert the calibrated flow tube (1) into the head
adaptor (2), as shown in figure 2. This forms the validation rig.
Figure 2: Tube and head adaptor assembly

Preparing the air sampler

Fit the supplied Petri dish (4) or contact plate (3) into the MB1 or
MB2 bioaerosol sampler to be validated and fit the sampling head.
Then place the validation rig on top of the MB1 or MB2 sampling
head as shown in figure 3.
Figure 3: Fitting validation rig to the MB1/MB2
bioaerosol sampler