Page 1
Hygienesystem Gerätedokument Einbau, Betrieb und Wartung Hygiene system Equipment Logbook Assembly, operation and maintenance Systeme d’hygiene Livret d’appareil Installation, fonctionnement et entretien Sistema d’igiene Verbale d’installazione Montaggio, funzionamento e manutenzione...
All inspection and service work must be entered into the device docu- mentation at the first pages of this manual When requested by an authorized engineer, METASYS agrees to make all documents available for the use of technically qualified service personnel.
450 mA Frequency 50/60 Hz max. power consumption 11 VA Volume of the H1 cartridge 125 ml (pouch: 500 ml) Suction system wet or dry vacuum pump Permissible vacuum 100 mbar - 250 mbar 128 x 132 x 95 mm Dimensions (H x W x D) (excl.
Installation requirements Practice personnel, technicians 7. Installation requirements A water filter (40 µm) and a pressure regulator (2 bar) must be attached to the mains water supply (included in the scope of delivery). The underpressure may not exceed 250 mbar at the cannula. ...
The valve block, in combination with the double membrane pump, is the heart of the Hygiene System H1. It determines the ratio of water to disinfectant, dispenses the cleaning agent to the respective handpiece and controls the elbow closure pieces.
Functional description Technicians 9. Functional description Function diagram, see illustration Suction flow Disinfectant GREEN&CLEAN H1 Water H1 cartridge or pouch DVGW container (water) Double membrane pump Pump chamber Suction flow closure Chemical closure Filter housing Handpiece Suction tube with inner tube...
Page 8
Functional description Technicians 9. Functional description See illustration Water is drawn out of the DVGW container, or GREEN&CLEAN H1 is drawn out of the cartridge or pouch by means of a water pressure-driven double membrane pump. See illustration Once drawn off, the disinfectant is pumped through the knee sealing membrane and through lines in the suction hoses to the handpiece in use.
If the yellow monitor light no. 2 (empty tank warning) on the control panel flashes and a short alarm is audible every time the suction tube is lifted, please exchange the H1 cartridge or renew the pouch as follows: Removing the H1 cartridge: ...
Hygiene system H1 - control panel Practice personnel, technicians 12. Hygiene system H1 - control panel 28.2 28.3 28.1 See illustration Control light 1 (red): Malfunction 28.1 flashing red and 5-times beeping in the event of a fault; additional short beep each time the suction hose is lifted, red flashing signal remains;...
Valves V1 - V5 (X5 - VENTILE), 6-pole 29.6 Control panel (X6 - BED- TEIL) 29.7 Vacuum switch 29.8 Transition board, see illustration 29.6 29.7 GREEN CLEAN H1 & Knee valve, right Knee valve, left Additional valve for third suction hose 30.6...
Page 13
BF, a cut-off relay that complies with IEC 601-1, section 20.2, insulation B-a must be used. This cut-off relay is switched through the output contact X2, whereupon the operating voltage used must be the same as of the H1 system. H1 mode setting 31.1 SIC1.
Inspections Cleaning and disinfection Practice personnel, technicians 14. Inspections 33.2 We recommend an annual service check of the Hygiene System H1 by an 33.3 authorised technician. 15. Cleaning and disinfection 33.1 The below mentioned parts can be cleaned and disinfected as follows:...
Attach the short end of the adapter to the refill pouch. Attach the long end of the adapter to the cartridge. Refill the cartridge Fill up the cartridge with GREEN&CLEAN H1 until the maximum level is reached. Remove adapter after use and rinse with water.
Page 16
METASYS ... makes the difference! Visit us at: METASYS Medizintechnik GmbH Florianistraße 3, 6063 Rum bei Innsbruck, Austria 1 +43 512 205420 | 5 +43 512 205420 7 www.metasys.com | info@metasys.com FRANCE ITALY GERMANY +49 8823 938 44 33 +33 4 37 90 22 15 +39 045 981 4477 info@metasys.com...